Systematic evaluation of 16 FDK filter configurations on a GE eXplore CT 120 scanner produces MTF10 values from 0.93 to 2.35 lp/mm, integrated NPS from 75,670 to 13,259 HU², and Rose-criterion detectable diameters from 0.93 to 7.74 mm depending on contrast.
Title resolution pending
1 Pith paper cite this work. Polarity classification is still indexing.
1
Pith paper citing it
fields
physics.med-ph 1years
2026 1verdicts
ACCEPT 1representative citing papers
citing papers explorer
-
Resolution-Noise Characteristics of Common FDK Filter Kernels: A Practical Reference for Preclinical Cone-Beam Micro-CT
Systematic evaluation of 16 FDK filter configurations on a GE eXplore CT 120 scanner produces MTF10 values from 0.93 to 2.35 lp/mm, integrated NPS from 75,670 to 13,259 HU², and Rose-criterion detectable diameters from 0.93 to 7.74 mm depending on contrast.