Fabrication and Performance Tests of a Segmented P-Type HPGe Detector

classification ⚛️ nucl-ex
keywords detectorhpgep-typesegmentedbeencircumferentialcuttingdead
0
0 comments X
read the original abstract

A p-type HPGe detector has been segmented by cutting, with a diamond saw, and etching six circumferential grooves through the Li-diffused dead layer. The degree of segmentation was tested with the 88 keV gamma rays from a well-collimated source of 109Cd. The fraction of events, recognized as occurring in more than one segment, and rejected from the energy interval 2038 +/- 5 keV, was measured as 0.59.