Conduction Effect of Thermal Radiation in a Metal Shield Pipe in a Cryostat for a Cryogenic Interferometric Gravitational Wave Detector
classification
⚛️ physics.ins-det
gr-qc
keywords
metalpiperadiationshieldthermalcausedconductioncryogenic
read the original abstract
A large heat load caused by thermal radiation through a metal shield pipe was observed in a cooling test of a cryostat for a prototype of a cryogenic interferometric gravitational wave detector. The heat load was approximately 1000 times larger than the value calculated by the Stefan-Boltzmann law. We studied this phenomenon by simulation and experiment and found that it was caused by the conduction of thermal radiation in a metal shield pipe.