Code-division multiplexing for x-ray microcalorimeters

Authors on Pith no claims yet
classification ⚛️ physics.ins-det astro-ph.IM
keywords multiplexingcode-divisionhighermicrocalorimetersmultiplexedreadoutresolutionapproach
0
0 comments X
read the original abstract

We demonstrate the code-division multiplexed (CDM) readout of eight transition-edge sensor microcalorimeters. The energy resolution is 3.0 eV (full width at half-maximum) or better at 5.9 keV, with a best resolution of 2.3 eV and a mean of 2.6 eV over the seven modulated detectors. The flux-summing CDM system is described and compared with similar time-division multiplexed (TDM) readout. We show that the sqrt(Npixels) multiplexing disadvantage associated with TDM is not present in CDM. This demonstration establishes CDM as both a simple route to higher performance in existing TDM microcalorimetric experiments and a long-term approach to reaching higher multiplexing factors.