Cryostat for Ultra-low-energy Threshold Germanium Spectrometers
classification
⚛️ physics.ins-det
nucl-ex
keywords
cryostatdetectiongermaniumlow-energyspectrometersthermalthresholdacquired
read the original abstract
This paper presents progress on the development of a cryostat intended to improve upon the low-energy threshold (below 0.5 keV) of p-type point contact germanium gamma-ray spectrometers. Ultra-low energy thresholds are important in the detection of low-energy nuclear recoils, an event class relevant to both dark matter direct detection and measurement of coherent neutrino-nucleus scattering. The cryostat design, including a thermal and electrical-field model, is given. A prototype cryostat has been assembled and data acquired to evaluate its vacuum and thermal performance.