High-resolution gamma-ray spectroscopy with a microwave-multiplexed transition-edge sensor array

Authors on Pith no claims yet
classification ⚛️ physics.ins-det astro-ph.IMhep-ex
keywords readoutsensorarraychangesgamma-raylambdamicrowavemicrowave-multiplexed
0
0 comments X
read the original abstract

We demonstrate very high resolution photon spectroscopy with a microwave-multiplexed two-pixel transition-edge sensor (TES) array. We measured a $^{153}$Gd photon source and achieved an energy resolution of 63 eV full-width-at-half-maximum at 97 keV and an equivalent readout system noise of 86 pA/$\sqrt{\text{Hz}}$ at the TES. The readout circuit consists of superconducting microwave resonators coupled to radio-frequency superconducting-quantum-interference-devices (SQUID) and transduces changes in input current to changes in phase of a microwave signal. We use flux-ramp modulation to linearize the response and evade low-frequency noise. This demonstration establishes one path for the readout of cryogenic X-ray and gamma-ray sensor arrays with more than $10^3$ elements and spectral resolving powers $R=\lambda/\Delta\lambda > 10^3$.

This paper has not been read by Pith yet.

discussion (0)