Current Sensing Noise Thermometry: A fast practical solution to low temperature measurement
pith:LBTG62KS Add to your LaTeX paper
What is a Pith Number?\usepackage{pith}
\pithnumber{LBTG62KS}
Prints a linked pith:LBTG62KS badge after your title and writes the identifier into PDF metadata. Compiles on arXiv with no extra files. Learn more
read the original abstract
We describe the design and performance of a series of fast, precise current sensing noise thermometers. The thermometers have been fabricated with a range of resistances from 1.290 $\Omega$ down to 0.2 m$\mathrm{\Omega}$. This results in either a thermometer that has been optimised for speed, taking advantage of the improvements in superconducting quantum interference device (SQUID) noise and bandwidth, or a thermometer optimised for ultra-low temperature measurement, minimising the system noise temperature. By using a single temperature calibration point, we show that noise thermometers can be used for accurate measurements over a wide range of temperatures below 4 K. Comparisons with a melting curve thermometer, a calibrated germanium thermometer and a pulsed platinum nuclear magnetic resonance thermometer are presented. For the 1.290 $\mathrm{\Omega}$ resistance we measure a 1 % precision in just 100 ms, and have shown this to be independent of temperature.
This paper has not been read by Pith yet.
discussion (0)
Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.