Measurement of the response of a liquid scintillation detector to monoenergetic electrons and neutrons

classification ⚛️ physics.ins-det nucl-ex
keywords liquidresponsescintillatorneutronsbeenelectronsmeasuredgamma
0
0 comments X
read the original abstract

The response of the liquid scintillator (EJ-301 equivalent to NE-213) to the monoenergetic electrons produced in Compton scattered $\gamma$-ray tagging has been carried out for various radioactive $\gamma$-ray sources. The measured electron response is found to be linear up to $\sim$4~MeVee and the resolution of the liquid scintillator at 1~MeVee is observed to be $\sim$~11\%. The pulse shape discrimination and pulse height response of the liquid scintillator for neutrons has been measured using $^7$Li(p,n$_1$)$^7$Be*(0.429 MeV) reaction. Non linear response to mono-energetic neutrons for the liquid scintillator is observed at E$_n$=5.3, 9.0 and 12.7 MeV. The measured response of the liquid scintillator for electrons and neutrons have been compared with Geant4 simulation.