pith. sign in

arxiv: 1805.11148 · v2 · pith:5425GDTRnew · submitted 2018-05-25 · ⚛️ physics.ins-det · cond-mat.mes-hall· cond-mat.str-el

Atomically resolved probe-type scanning tunneling microscope for use in harsh vibrational cryogen-free superconducting magnet

classification ⚛️ physics.ins-det cond-mat.mes-hallcond-mat.str-el
keywords magnetsuperconductingtemperaturetunnelingcryogen-freeharshscanningatomic
0
0 comments X
read the original abstract

We present a probe-type scanning tunneling microscope (STM) with atomic resolution that is designed to be directly inserted and work in a harsh vibrational cryogen-free superconducting magnet system. When a commercial variable temperature insert (VTI) is installed in the magnet and the STM is in turn housed in the VTI, a lowest temperature of 1.6 K can be achieved, where the STM still operates well. We have tested it in an 8 T superconducting magnet cooled with the pulse-tube cryocooler (PTC) and obtained atomically revolved graphite and NiSe2 images as well as the scanning tunneling spectrum (STS, i.e. dI/dV spectrum) data of the latter near its critical temperature, which show the formation process of the superconducting gap as a function of temperature. The drifting rates of the STM at 1.6 K in X-Y plane and Z direction are 1.15 and 1.71 pm/min respectively. Noise analysis for the tunneling current shows that the amplitudes of the dominant peaks (6.84 and 10.25Hz) are low. This is important as a cryogen-free magnet system has long been considered too harsh for any atomic resolution measurement.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.