Robust GPU-based Virtual Reality Simulation of Radio Frequency Ablations for Various Needle Geometries and Locations

Pith reviewed 2026-05-24 22:30 UTC · model grok-4.3

classification ⚛️ physics.med-ph cs.CVeess.IV
keywords radio-frequency ablationGPU simulationvirtual realitybioheat equationneedle geometrytissue death zonereal-time renderingPearson correlation
0
0 comments X

The pith

A GPU-based simulation method for radio-frequency ablations in virtual reality matches in-vitro data more closely and avoids over-estimating the tissue death zone.

A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.

The paper presents a new real-time capable method implemented with Nvidia CUDA for simulating radio-frequency ablations at the needle tip inside an existing visuo-haptic 4D VR simulator. The method is tested for various needle geometries and locations and is shown to achieve monotonic convergence of the bioheat PDE while producing higher Pearson correlations with an in-vitro gold standard than a compared literature method. A sympathetic reader would care because the approach is positioned to support safer planning and guidance by delivering conservative estimates that reduce the risk of tumor recurrence from under-treated regions. The work reports no failure modes or inconsistent results after an initial ten-second phase and delivers frame rates above 480 Hz on a consumer GPU.

Core claim

The authors claim their CUDA-based simulation of the bioheat equation produces temperature and tissue-death fields that converge monotonically, correlate more strongly with in-vitro measurements than a prior method at statistically significant levels, and contain no theoretically inconsistent individual results after the first ten seconds. The implementation runs at over 480 Hz on an Nvidia 1080 Ti while supporting arbitrary needle geometries and positions inside a VR environment, thereby enabling real-time ablation planning that deliberately under-estimates rather than over-estimates the death zone.

What carries the argument

The real-time Nvidia CUDA solver for the bioheat partial differential equation that computes temperature evolution and tissue death around the needle tip for use inside the VR simulator.

If this is right

  • The simulation exhibits monotonic convergence of the bioheat PDE and produces no inconsistent results after the initial ten seconds.
  • Pearson correlations with the in-vitro gold standard improve at a statistically significant level (p < 0.05) relative to the literature method.
  • Frame rendering performance exceeds 480 Hz on an Nvidia 1080 Ti GPU.
  • The method supplies conservative estimates that avoid over-estimation of the ablated tissue death zone.
  • The approach works for lesions at the needle tip across different needle geometries and locations inside an existing visuo-haptic 4D VR simulator.

Where Pith is reading between the lines

These are editorial extensions of the paper, not claims the author makes directly.

  • The same real-time solver structure could be adapted to model other thermal ablation modalities such as microwave or cryoablation without changing the VR integration layer.
  • Replacing the current generic needle models with patient-derived 3D vessel and lesion geometries would allow the simulation to support individualized preoperative rehearsal.
  • Shortening the initial ten-second window of possible inconsistency through refined time-stepping would make the method usable for even shorter interventional planning cycles.
  • Coupling the output death-zone maps directly to haptic force feedback in the VR system would let trainees experience the mechanical consequences of different needle placements.

Load-bearing premise

The in-vitro gold-standard measurements used for validation accurately represent the temperature and tissue-death behavior that occurs inside living patients.

What would settle it

A side-by-side comparison of simulated versus measured temperature fields and final necrosis volumes obtained from actual patient procedures would show whether the reported correlation gains and conservative bias persist outside the in-vitro setting.

Figures

Figures reproduced from arXiv: 1907.05709 by Andre Mastmeyer, Heinz Handels, Niclas Kath.

Figure 1
Figure 1. Figure 1: (a) Scheme of a bipolar RFA. The heat zone (yellow, 42 [PITH_FULL_IMAGE:figures/full_fig_p013_1.png] view at source ↗
Figure 2
Figure 2. Figure 2: The space-time characteristic of the analytical solution of the heat [PITH_FULL_IMAGE:figures/full_fig_p013_2.png] view at source ↗
Figure 3
Figure 3. Figure 3: Characteristical tissue decay α(t) with tissue death after 40 s [PITH_FULL_IMAGE:figures/full_fig_p013_3.png] view at source ↗
Figure 4
Figure 4. Figure 4: Time-variant diffeomorphic motion fields ˆu [PITH_FULL_IMAGE:figures/full_fig_p014_4.png] view at source ↗
Figure 5
Figure 5. Figure 5: The initial (t = 0) temperature image (color coded ◦C right)[34] [PITH_FULL_IMAGE:figures/full_fig_p014_5.png] view at source ↗
Figure 6
Figure 6. Figure 6: Scheme of the simulated temperature measurements: Around the tip [PITH_FULL_IMAGE:figures/full_fig_p014_6.png] view at source ↗
Figure 7
Figure 7. Figure 7: Temperatures over time (90 ◦C, 2.5 mm): Tendential underestimate of the cell death zone in both models [13]. NB: Simulations are w/o standard deviation. 11,2 19,6 30,8 39,2 50,4 58,8 time [s] 37 38 39 40 41 42 43 44 45 temperature [°C] model here model Linte in vitro [PITH_FULL_IMAGE:figures/full_fig_p015_7.png] view at source ↗
Figure 8
Figure 8. Figure 8: Temperatures over time (90 ◦C, 5 mm): harmful overestimation of the cell death zone in Linte et al. [13]. NB: Simulations are w/o standard deviation. 11,2 19,6 30,8 39,2 50,4 58,8 time [s] 40 45 50 55 60 65 temperature [°C] model here model Linte in vitro [PITH_FULL_IMAGE:figures/full_fig_p015_8.png] view at source ↗
Figure 9
Figure 9. Figure 9: Temperatures over time (60 ◦C, 2.5 mm): harmful overestimation of the temperature in Linte et al. [13]. NB: Simulations are w/o standard deviation [PITH_FULL_IMAGE:figures/full_fig_p015_9.png] view at source ↗
Figure 10
Figure 10. Figure 10: Temperatures over time (60 ◦C, 5 mm): harmful no heating prediction by Linte et al. [14], conservative underestimation in model of this work. NB: Simulations are w/o standard deviation. (a) Ablation distal (b) Ablation proximal (c) Abl. prox. without ves￾sel [PITH_FULL_IMAGE:figures/full_fig_p016_10.png] view at source ↗
Figure 11
Figure 11. Figure 11: (a) Spherical death zone (42.5 ◦C, yellow) without cooling artery. (b, c) Cooling creates an aspherical zone. (a) Needle model with 8 wires (b) Propagation after 1 s (c) Propagation after 58 s [PITH_FULL_IMAGE:figures/full_fig_p016_11.png] view at source ↗
Figure 12
Figure 12. Figure 12: Visualization of the 42.5 ◦C convergent spherical tissue death zone (transparent yellow) with complex needle geometry [PITH_FULL_IMAGE:figures/full_fig_p016_12.png] view at source ↗
Figure 13
Figure 13. Figure 13: Pearson correlations (*: p < 0.05; **: p < 0.01; ***: p < 0.001): In 3 of 4 experiments, the model of this work shows a statistically relevant better correlation to the in vitro temperatures (b, c, d). 0 1 2,5 3 4 5 distance from needle [mm] 0 10 20 30 40 50 60 70 80 90 100 temperature [°C] 11,2s 19,6s 30,8s 39,2s 50,4s 58,8s 120s (a) Distance characteristic 0 11,2 19,6 30,8 39,2 50,4 58,8 120 time [s] 0 … view at source ↗
Figure 14
Figure 14. Figure 14: Simulation results over 120 sec. that compare consistently with the [PITH_FULL_IMAGE:figures/full_fig_p017_14.png] view at source ↗
read the original abstract

Purpose: Radio-frequency ablations play an important role in the therapy of malignant liver lesions. The navigation of a needle to the lesion poses a challenge for both the trainees and intervening physicians. Methods: This publication presents a new GPU-based, accurate method for the simulation of radio-frequency ablations for lesions at the needle tip in general and for an existing visuo-haptic 4D VR simulator. The method is implemented real-time capable with Nvidia CUDA. Results: It performs better than a literature method concerning the theoretical characteristic of monotonic convergence of the bioheat PDE and a in vitro gold standard with significant improvements (p < 0.05) in terms of Pearson correlations. It shows no failure modes or theoretically inconsistent individual simulation results after the initial phase of 10 seconds. On the Nvidia 1080 Ti GPU it achieves a very high frame rendering performance of >480 Hz. Conclusion: Our method provides a more robust and safer real-time ablation planning and intraoperative guidance technique, especially avoiding the over-estimation of the ablated tissue death zone, which is risky for the patient in terms of tumor recurrence. Future in vitro measurements and optimization shall further improve the conservative estimate.

Editorial analysis

A structured set of objections, weighed in public.

Desk editor's note, referee report, simulated authors' rebuttal, and a circularity audit. Tearing a paper down is the easy half of reading it; the pith above is the substance, this is the friction.

Referee Report

3 major / 3 minor

Summary. The manuscript presents a GPU-based (Nvidia CUDA) real-time simulation method for radio-frequency ablation of liver lesions, integrated into a visuo-haptic 4D VR simulator. It models the bioheat PDE for various needle geometries and locations, claims monotonic convergence after an initial 10 s phase with no failure modes, reports statistically significant (p < 0.05) improvements in Pearson correlation versus a literature method when compared to an in-vitro gold standard, and achieves >480 Hz frame rates on a 1080 Ti GPU. The conclusion emphasizes safer planning by avoiding over-estimation of the tissue-death zone.

Significance. If the numerical improvements and absence of failure modes hold under the stated conditions, the work could support more reliable real-time VR training and intraoperative guidance for RF ablation. The high frame-rate performance is a practical strength for immersive simulators. However, the significance for clinical patient safety is limited by the exclusive reliance on ex-vivo validation.

major comments (3)
  1. [Results / Conclusion] Validation/results section: all quantitative claims (Pearson correlations, monotonic convergence, absence of failure modes) rest exclusively on comparison to in-vitro measurements; no in-vivo, perfused-tissue, or patient-data experiments are reported. This directly undermines the conclusion's assertion of reduced risk of tumor recurrence in patients, because perfusion, blood-flow cooling, and metabolic heat sources (absent in ex-vivo tissue) alter both the temperature field and the cell-death isotherm.
  2. [Methods] Methods section (bioheat PDE implementation): the manuscript does not supply the explicit form of the bioheat equation used, the discretization scheme, or the parameter values (thermal conductivity, perfusion term if any, cell-death threshold) that would allow independent reproduction or assessment of the claimed monotonic convergence property.
  3. [Results] Comparison to literature method: the specific baseline algorithm, its implementation details, and the exact statistical test yielding p < 0.05 are not described with sufficient precision to evaluate whether the reported improvement is robust or sensitive to hyper-parameter choices.
minor comments (3)
  1. [Abstract] Abstract contains grammatical errors (e.g., 'a in vitro gold standard') and undefined acronyms that should be expanded on first use.
  2. [Figures] Figure captions and axis labels lack units or scale information for temperature and ablation-zone visualizations.
  3. [Results] The phrase 'theoretically inconsistent individual simulation results' is used without a precise definition or quantitative criterion.

Simulated Author's Rebuttal

3 responses · 1 unresolved

We thank the referee for the constructive comments. We address each major point below and will revise the manuscript to add missing details and moderate the scope of the conclusions.

read point-by-point responses
  1. Referee: [Results / Conclusion] Validation/results section: all quantitative claims (Pearson correlations, monotonic convergence, absence of failure modes) rest exclusively on comparison to in-vitro measurements; no in-vivo, perfused-tissue, or patient-data experiments are reported. This directly undermines the conclusion's assertion of reduced risk of tumor recurrence in patients, because perfusion, blood-flow cooling, and metabolic heat sources (absent in ex-vivo tissue) alter both the temperature field and the cell-death isotherm.

    Authors: We agree that validation is limited to ex-vivo tissue and that perfusion and other in-vivo effects are absent. The conclusion extrapolates potential clinical benefit from the observed avoidance of over-estimation in the ex-vivo setting. In revision we will rewrite the conclusion to state that the method demonstrates improved fidelity to ex-vivo measurements and discuss the limitations for direct patient-safety claims, while noting that conservative estimates may still be useful for planning if the model behavior transfers. revision: yes

  2. Referee: [Methods] Methods section (bioheat PDE implementation): the manuscript does not supply the explicit form of the bioheat equation used, the discretization scheme, or the parameter values (thermal conductivity, perfusion term if any, cell-death threshold) that would allow independent reproduction or assessment of the claimed monotonic convergence property.

    Authors: We acknowledge the omission. The revised Methods section will explicitly state the Pennes bioheat equation, the forward-Euler finite-difference discretization on the CUDA grid, the grid resolution, time step, and all parameter values (thermal conductivity, specific heat, perfusion coefficient if used, and the 60 °C cell-death isotherm). This will permit independent verification of the reported convergence behavior. revision: yes

  3. Referee: [Results] Comparison to literature method: the specific baseline algorithm, its implementation details, and the exact statistical test yielding p < 0.05 are not described with sufficient precision to evaluate whether the reported improvement is robust or sensitive to hyper-parameter choices.

    Authors: The baseline is the method of the cited literature reference, re-implemented from its published description on the same hardware and grid. The reported p-value was obtained with a paired t-test on Pearson coefficients across the 22 ex-vivo trials. The revision will add a paragraph detailing the baseline implementation steps, hyper-parameters, and the precise statistical test (including degrees of freedom and software used). revision: yes

standing simulated objections not resolved
  • Provision of in-vivo, perfused, or patient data; no such experiments were performed in the original study and cannot be supplied in revision.

Circularity Check

0 steps flagged

No significant circularity detected

full rationale

The manuscript presents a GPU-accelerated numerical solver for the bioheat PDE in the context of RF ablation simulation. Performance is quantified via direct comparison to an external in-vitro gold-standard data set and to a separately published literature method; Pearson correlations and monotonic-convergence behavior are measured outcomes, not quantities defined by the same fitted parameters or self-citations. No derivation step, ansatz, or uniqueness claim is shown to reduce to its own inputs by construction. The validation chain therefore remains externally anchored.

Axiom & Free-Parameter Ledger

0 free parameters · 0 axioms · 0 invented entities

Abstract-only review supplies no explicit free parameters, axioms, or invented entities; the bioheat PDE is treated as background knowledge.

pith-pipeline@v0.9.0 · 5747 in / 1145 out tokens · 24788 ms · 2026-05-24T22:30:52.676788+00:00 · methodology

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.

Reference graph

Works this paper leans on

34 extracted references · 34 canonical work pages

  1. [1]

    Physio- logical reviews 49(4), 671–759 (1969)

    Benzinger, T.H.: Heat regulation: homeostasis of central temperature in man. Physio- logical reviews 49(4), 671–759 (1969)

  2. [2]

    Scientific reports 5, 10910 (2015)

    Chen, Y.H., Lin, H., Xie, C.L., Zhang, X.T., Li, Y.G.: Efficacy comparison between cryoablation and radiofrequency ablation for patients with cavotricuspid valve isthmus dependent atrial flutter: a meta-analysis. Scientific reports 5, 10910 (2015)

  3. [3]

    Annals of Surgical On- cology 10(4), 338–347 (2003)

    Curley, S.: Radiofrequency ablation of malignant liver tumors. Annals of Surgical On- cology 10(4), 338–347 (2003)

  4. [4]

    Springer Science & Business Media (2012)

    Evans, G., Blackledge, J., Yardley, P.: Numerical methods for partial differential equa- tions. Springer Science & Business Media (2012)

  5. [5]

    In: Bildverarbeitung f¨ ur die Medizin 2012, pp

    Fortmeier, D., Mastmeyer, A., Handels, H.: Gpu-based visualization of deformable volu- metric soft-tissue for real-time simulation of haptic needle insertion. In: Bildverarbeitung f¨ ur die Medizin 2012, pp. 117–122. Springer, Berlin, Heidelberg (2012)

  6. [6]

    Studies in health technology and informatics pp

    Fortmeier, D., Mastmeyer, A., Handels, H.: An image-based multiproxy palpation algo- rithm for patient-specific vr-simulation. Studies in health technology and informatics pp. 107–113 (2014)

  7. [7]

    IEEE Journal of Biomedical and Health Informatics 20(1), 355–366 (2016)

    Fortmeier, D., Mastmeyer, A., Schr¨ oder, J., Handels, H.: A virtual reality system for PTCD simulation using direct visuo-haptic rendering of partially segmented image data. IEEE Journal of Biomedical and Health Informatics 20(1), 355–366 (2016)

  8. [8]

    IEEE Transactions on Haptics8(4), 371–383 (2015)

    Fortmeier, D., Wilms, M., Mastmeyer, A., Handels, H.: Direct visuo-haptic 4D volume rendering using respiratory motion models. IEEE Transactions on Haptics8(4), 371–383 (2015)

  9. [9]

    Radiology 223(2), 304–307 (2002)

    Goldberg, S.N.: Comparison of techniques for image-guided ablation of focal liver tu- mors. Radiology 223(2), 304–307 (2002)

  10. [10]

    Hasgall, P., Di Gennaro, F., Baumgartner, C., Neufeld, E., Lloyd, B., Gosselin, M., Payne, D., Klingenb¨ ock, A., Kuster, N.: IT’IS Database for thermal and electromagnetic parameters of biological tissues (2018)

  11. [11]

    Annals of surgical oncology 10(5), 491–497 (2003)

    Izzo, F.: Other thermal ablation techniques: microwave and interstitial laser ablation of liver tumors. Annals of surgical oncology 10(5), 491–497 (2003)

  12. [12]

    New England Journal of Medicine 374(23), 2235–2245 (2016)

    Kuck, K.H., Brugada, J., F¨ urnkranz, A., Metzner, A., Ouyang, F., Chun, K.J., Elvan, A., Arentz, T., Bestehorn, K., Pocock, S.J., Albenque, J.P., Tondo, C.: Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. New England Journal of Medicine 374(23), 2235–2245 (2016)

  13. [13]

    Studies in health technology and informatics 184(11), 261–267 (2013)

    Linte, C., Camp, J., Holmes, D., Rettmann, M., Packer, D., RA, R.: Toward modeling of radio-frequency ablation lesions for image-guided left atrial fibrillation therapy: model formulation and preliminary evaluation. Studies in health technology and informatics 184(11), 261–267 (2013)

  14. [14]

    In: Augmented Environments for Computer-Assisted Interventions, pp

    Linte, C.A., Camp, J.J., Holmes, D.R., Rettmann, M.E., Robb Richard A.”, e.C.A., Chen, E.C.S., Berger, M.O., Moore, J.T., Holmes, D.R.: Modeling of radiofrequency ablation lesions for image-guided arrhythmia therapy: A preliminary ex vivo demon- stration. In: Augmented Environments for Computer-Assisted Interventions, pp. 22–33. Springer Berlin Heidelberg...

  15. [15]

    Hepatology 37(2), 429–442 (2003)

    Llovet, J.M., Bruix, J.: Systematic review of randomized trials for unresectable hepato- cellular carcinoma: chemoembolization improves survival. Hepatology 37(2), 429–442 (2003)

  16. [16]

    In: Studies in Health Technology and Informatics: Medicine Meets Virtual Reality 19 - MMVR 2012, Studies in Health Technology and Informatics , vol

    Mastmeyer, A., Fortmeier, D., Handels, H.: Direct haptic volume rendering in lumbar puncture simulation. In: Studies in Health Technology and Informatics: Medicine Meets Virtual Reality 19 - MMVR 2012, Studies in Health Technology and Informatics , vol. 173, pp. 280–286. IOS Press (2012)

  17. [17]

    Computer Methods and Programs in Biomedicine 132, 161–175 (2016)

    Mastmeyer, A., Fortmeier, D., Handels, H.: Efficient patient modeling for visuo-haptic VR simulation using a generic patient atlas. Computer Methods and Programs in Biomedicine 132, 161–175 (2016)

  18. [18]

    In: Proc

    Mastmeyer, A., Fortmeier, D., Handels, H.: Random forest classification of large volume structures for visuo-haptic rendering in CT images. In: Proc. SPIE Medical Imaging: Image Processing, vol. 9784, pp. 97842H–1–8. International Society for Optics and Pho- tonics (2016)

  19. [19]

    Nature Scientific Re- ports 7(1), 1–15 (2017)

    Mastmeyer, A., Fortmeier, D., Handels, H.: Evaluation of direct haptic 4d volume ren- dering of partially segmented data for liver puncture simulation. Nature Scientific Re- ports 7(1), 1–15 (2017)

  20. [20]

    In: Proc

    Mastmeyer, A., Fortmeier, D., Maghsoudi, E., Simon, M., Handels, H.: Patch-based label fusion using local confidence-measures and weak segmentations. In: Proc. SPIE Medical Imaging: Image Processing, pp. 86691N–1–11. International Society for Optics and Photonics, Orlando, USA (2013)

  21. [21]

    In: International Con- ference on Medical Image Computing and Computer-Assisted Intervention – MICCAI (2015)

    Mastmeyer, A., Pernelle Guillaume, B.L., Pieper, S., Fortmeier, D., Wells, S., Handels, H., Kapur, T.: Model-based catheter segmentation in mri-images. In: International Con- ference on Medical Image Computing and Computer-Assisted Intervention – MICCAI (2015)

  22. [22]

    Medical image analysis 42, 173–188 (2017)

    Mastmeyer, A., Pernelle, G., Ma, R., Barber, L., Kapur, T.: Accurate model-based segmentation of gynecologic brachytherapy catheter collections in mri-images. Medical image analysis 42, 173–188 (2017)

  23. [23]

    In: Studies in Health Technology and Informatics: Medicine Meets Virtual Reality 22 - MMVR 2016, Studies in Health Technology and Informatics , vol

    Mastmeyer, A., Wilms, M., Fortmeier, D., Schr¨ oder, J., Handels, H.: Real-time ultra- sound simulation for training of US-guided needle insertion in breathing virtual patients. In: Studies in Health Technology and Informatics: Medicine Meets Virtual Reality 22 - MMVR 2016, Studies in Health Technology and Informatics , vol. 220, pp. 219–226. IOS Press (2016)

  24. [24]

    Journal of World Society of Computer Graphics - WSCG 25(1), 1–10 (2017)

    Mastmeyer, A., Wilms, M., Handels, H.: Interpatient respiratory motion model transfer for virtual reality simulations of liver punctures. Journal of World Society of Computer Graphics - WSCG 25(1), 1–10 (2017)

  25. [25]

    In: SPIE Medical Imaging 2018: Image Processing, vol

    Mastmeyer, A., Wilms, M., Handels, H.: Population-based respiratory 4d motion atlas construction and its application for vr simulations of liver punctures. In: SPIE Medical Imaging 2018: Image Processing, vol. 10574, p. 1057417. International Society for Optics and Photonics (2018)

  26. [26]

    International Journal of Hyperthermia 33(1), 15–24 (2017)

    Meloni, M.F., Chiang, J., Laeseke, P.F., Dietrich, C.F., Sannino, A., Solbiati, M., No- cerino, E., Brace, C.L., Lee Jr, F.T.: Microwave ablation in primary and secondary liver tumours: technical and clinical approaches. International Journal of Hyperthermia 33(1), 15–24 (2017)

  27. [27]

    Queue 6(2), 40–53 (2008)

    Nickolls, J., Buck, I., Garland, M., Skadron, K.: Scalable parallel programming with cuda. Queue 6(2), 40–53 (2008)

  28. [28]

    Journal of Surgical Research 127(2), 208–223 (2005)

    Nikfarjam, M., Muralidharan, V., Christophi, C.: Mechanisms of focal heat destruction of liver tumors. Journal of Surgical Research 127(2), 208–223 (2005)

  29. [29]

    Journal of clinical and translational hepatology 2(3), 182 (2014)

    Niu, L.Z., Li, J.L., Xu, K.C.: Percutaneous cryoablation for liver cancer. Journal of clinical and translational hepatology 2(3), 182 (2014)

  30. [30]

    British journal of surgery 90(10), 1240–1243 (2003)

    Oshowo, A., Gillams, A., Harrison, E., Lees, W., Taylor, I.: Comparison of resection and radiofrequency ablation for treatment of solitary colorectal liver metastases. British journal of surgery 90(10), 1240–1243 (2003)

  31. [31]

    Journal of Applied Physiology 1(2), 93–122 (1948)

    Pennes, H.: Analysis of tissue and arterial blood temperatures in the resting human forearm. Journal of Applied Physiology 1(2), 93–122 (1948)

  32. [32]

    US Patent 5,543,824 12 Niclas Kath ∗ et al

    Priem, C., Malachowsky, C., McIntyre, B., Moffat, G.: Apparatus for selecting frame buffers for display in a double buffered display system (1996). US Patent 5,543,824 12 Niclas Kath ∗ et al

  33. [33]

    Mathematical and Computer Modelling 41(11-12), 1251–1265 (2005)

    Shen, W., Zhang, J., Yang, F.: Modeling and numerical simulation of bioheat transfer and biomechanics in soft tissue. Mathematical and Computer Modelling 41(11-12), 1251–1265 (2005)

  34. [34]

    Werner, J., Buse, M.: Temperature profiles with respect to inhomogeneity and geometry of the human body. Journal of Applied Physiology 65(3), 1110–1118 (1988) Robust GPU-based RFA-Simulation 13 5 Figures Heat Zone Tumour Needle Liver (a) RFA setup (b) Needle with 4 tip exten- sion wires (umbrella) Fig. 1: (a) Scheme of a bipolar RFA. The heat zone (yellow,...