A Procedure to Calibrate a Multi-Modular Telescope
classification
⚛️ nucl-ex
keywords
energychargemassprocedurecalibrationtechniqueapparatusapplication
read the original abstract
A procedure has been developed for the charge, mass and energy calibration of ions produced in nuclear heavy ion reactions. The charge and mass identification are based on a $\Delta$E-E technique. A computer code determines the conversion from ADC channels into energy values, atomic number and mass of the detected fragments by comparing with energy loss calculations through a minimization routine. The procedure does not need prior measurements with beams of known energy and charge. An application of this technique to the calibration of the MULTICS apparatus is described.