Characterization of bolometric Light Detectors for rare event searches

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 1304.6289 v4 pith:65OH4REC submitted 2013-04-23 physics.ins-det nucl-ex

classification physics.ins-detnucl-ex
keywords lightdetectorbackgroundbolometersbolometricdetectorsenergyperformances
verification ladder T0 review T1 audit T2 compute T3 formal
0 comments
read the original abstract

Bolometers have proven to be very good detectors to search for rare processes thanks to their excellent energy resolution and their low intrinsic background. Further active background rejection can be obtained by the simultaneous readout of the heat and light signals produced by particles interacting in scintillating bolometers, as proposed by the LUCIFER experiment. In this framework, the choice of the light detector and the optimization of its working conditions play a crucial role. In this paper, we report a study of the performances of a Germanium bolometric light detector in terms of signal amplitude, energy resolution and signal time development. The impact of various operational parameters on the detector performances is discussed.