Expanded flow rate range of high-resolution nanoDMAs via improved sample flow injection at the aerosol inlet slit

Instrumentation and Detectors · Submitted 25 Apr 2019

abstract compiled · 2026-09-23 · no language-model review

The authors' main claim By introducing a ring with many perforations upstream of the inlet slit, which favors a more symmetric distribution of the sample flow over the slit perimeter, R values close to ideal are approached with qi/Q ratios as large as 6 to 12%.

how it is stated · asserted 0.96 read the record →

Comment (0) Read on arXiv ↗
Cite ▾
arXiv ID1904.11339 (version 1)
Pith IDXMZDOPTA, Pith's permanent record of this paper and its review
CitationJuan Fernandez de la Mora. "Expanded flow rate range of high-resolution nanoDMAs via improved sample flow injection at the aerosol inlet slit." arXiv:1904.11339 (2019). https://pith.science/paper/XMZDOPTA
Claim paper

The compiled record

The paper's own sentences, arranged. Each label is drawn from a closed vocabulary; the number beside it is the probability the classifier gave that label. Numbers in the tables are the authors', checked by code. No sentence in this record was written by a model.

Open the compiled record

Compiled by classifier-1.13.0 via TypeSafe API, question bank abstract-p1-20260923, 2026-09-23: 1 request, 6,490 input tokens.

What the abstract itself states, as probabilities: claims first 0.49 · states limitation 0.86 · human participants 0.02 · controlled comparison 0.75 · randomized assignment 0.01 · states quantitative result 0.98 · compares with prior results 0.85 · mentions code or data availability 0.02.

The abstract, sentence by sentence

arXiv listing abstract v1. Results and the main claim are underlined; .

High-resolution DMAs requiring hundreds of liters per minute of sheath gas flow Q to classify 1 nm particles have not been previously examined and optimized under the modest Q values (tens of liters per minute) needed to classify particles well above 10 nm. Here we study the resolving power R (based on the relative width of the transfer function) of the Halfmini DMA, at sample flow rates q beyond 1 l/min. The charge-reduced electrosprayed ovalbumin protein used as test aerosol has an accurately determinable Gaussian mobility distribution, which limits the measurable R to at most 25 to 30. Non-ideal DMA response can however be precisely probed by comparing measured peak widths with the convolution of the protein's mobility distribution with the triangular Knutson-Whitby distribution associated to the finite qi/Q ratio. For the unmodified Halfmini DMA, R departs considerably from ideal at qi/Q values as modest as 2%, revealing a problem with the aerosol inlet flow. This imperfection is not removed by either increasing or decreasing the slit width, nor by reorienting axially the sample flow jet as it merges with the sheath gas. By introducing a ring with many perforations upstream of the inlet slit, which favors a more symmetric distribution of the sample flow over the slit perimeter, R values close to ideal are approached with qi/Q ratios as large as 6 to 12%. This improvement enables (in principle) classification of particles larger than 30 nm.

background contribution method result interpretation main claim

What the classifier reports for each sentence. The number is the probability it gave the label.
#sentencerolethe authors' own workstates the main conclusionstrength
s0 High-resolution DMAs requiring hundreds of liters per minute of sheath gas flow Q to classify 1 nm particles have not been previously examined and optimized under the modest Q values (tens of liters per minute) needed to classify particles well above 10 nm. gap 0.94 0.07 0.05 asserted 1.00
s1 Here we study the resolving power R (based on the relative width of the transfer function) of the Halfmini DMA, at sample flow rates q beyond 1 l/min. contribution 0.98 0.93 0.06 asserted 0.96
s2 The charge-reduced electrosprayed ovalbumin protein used as test aerosol has an accurately determinable Gaussian mobility distribution, which limits the measurable R to at most 25 to 30. limitation 0.70 0.71 0.03 asserted 0.92
s3 Non-ideal DMA response can however be precisely probed by comparing measured peak widths with the convolution of the protein's mobility distribution with the triangular Knutson-Whitby distribution associated to the finite qi/Q ratio. method 0.99 0.43 0.04 asserted 0.65
s4 For the unmodified Halfmini DMA, R departs considerably from ideal at qi/Q values as modest as 2%, revealing a problem with the aerosol inlet flow. result 0.98 0.91 0.07 asserted 1.00
s5 This imperfection is not removed by either increasing or decreasing the slit width, nor by reorienting axially the sample flow jet as it merges with the sheath gas. result 0.91 0.90 0.04 asserted 1.00
s6 By introducing a ring with many perforations upstream of the inlet slit, which favors a more symmetric distribution of the sample flow over the slit perimeter, R values close to ideal are approached with qi/Q ratios as large as 6 to 12%. result 0.49 0.94 0.70 asserted 0.96
s7 This improvement enables (in principle) classification of particles larger than 30 nm. interpretation 0.97 0.65 0.48 hedged possibility 0.95

What kind of paper the classifier reports this is: method or tool 0.70 · evidence from observation or experiment 1.00.

What was examined, and what was not

sources read
  • arXiv listing abstract v1
not examined
  • full text
  • figures and tables
  • references and citing sentences
  • relations to other papers
levels run
  • P abstract
not run
  • S sections
  • T body sentences
  • C citations
  • R cross-paper relations

Every quoted sentence on this record is the authors'. Every label is from Pith's closed vocabulary, with the probability shown beside it. Every number in the measurements table is the authors', checked by code. Record as JSON

The paper
Original abstract

High-resolution DMAs requiring hundreds of liters per minute of sheath gas flow Q to classify 1 nm particles have not been previously examined and optimized under the modest Q values (tens of liters per minute) needed to classify particles well above 10 nm. Here we study the resolving power R (based on the relative width of the transfer function) of the Halfmini DMA, at sample flow rates q beyond 1 l/min. The charge-reduced electrosprayed ovalbumin protein used as test aerosol has an accurately determinable Gaussian mobility distribution, which limits the measurable R to at most 25 to 30. Non-ideal DMA response can however be precisely probed by comparing measured peak widths with the convolution of the protein's mobility distribution with the triangular Knutson-Whitby distribution associated to the finite qi/Q ratio. For the unmodified Halfmini DMA, R departs considerably from ideal at qi/Q values as modest as 2%, revealing a problem with the aerosol inlet flow. This imperfection is not removed by either increasing or decreasing the slit width, nor by reorienting axially the sample flow jet as it merges with the sheath gas. By introducing a ring with many perforations upstream of the inlet slit, which favors a more symmetric distribution of the sample flow over the slit perimeter, R values close to ideal are approached with qi/Q ratios as large as 6 to 12%. This improvement enables (in principle) classification of particles larger than 30 nm.