Electron pair emission from surfaces: Some general experimental considerations

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 2111.09203 v2 pith:FTSAHPLZ submitted 2021-11-17 cond-mat.mtrl-sci

classification cond-mat.mtrl-sci
keywords experimentalpairdifferentelectroninstrumentapproachesemissionprimary
verification ladder T0 review T1 audit T2 compute T3 formal

Signed reviews

No signed human review yet.

0 comments
read the original abstract

We discuss some experimental facets of electron pair emission from surfaces using two different experimental approaches. In the first case the instrument consists of a pair of hemispherical analyzers which are operated with continuous primary beams of electrons or photons. The second instrument employs a pair of time-of-flight spectrometers which require a pulsed excitation source. A key experimental quantity is the ratio of `true' to `random' coincidences which can be determined in different ways. Regardless of the type of instrument the primary flux has to adopt a much smaller value than in single electron spectroscopy. We describe different approaches to obtain the relevant count rates, in particular the concept of operating with a delayed coincidence circuit. We also address the question on how to compare the two types of spectrometer in terms of their performance.