A Joint Calibration Technique for Improving Measurement Accuracy of Voltage and Current Probes During Synchronous Operation for RF Based Plasma Devices

Plasma Physics · Submitted 26 Dec 2022

abstract compiled · 2026-09-23 · no language-model review

The authors' main claim This paper presents a joint calibration scheme for voltage (V) and current (I) probes that helps resolve accurately voltage-current phase differences even when the difference is very close to $90^{\circ}$.

how it is stated · asserted 0.97 read the record →

Comment (0) Read on arXiv ↗
Cite ▾
arXiv ID2212.13047 (version 1)
Pith IDHIOIFEUI, Pith's permanent record of this paper and its review
CitationA. Ganguli, Debaprasad Sahu, Mahreen, Satyananda Kar, Veda Prakash Gajula. "A Joint Calibration Technique for Improving Measurement Accuracy of Voltage and Current Probes During Synchronous Operation for RF Based Plasma Devices." arXiv:2212.13047 (2022). https://pith.science/paper/HIOIFEUI
Claim paper

The compiled record

The paper's own sentences, arranged. Each label is drawn from a closed vocabulary; the number beside it is the probability the classifier gave that label. Numbers in the tables are the authors', checked by code. No sentence in this record was written by a model.

Open the compiled record

Compiled by classifier-1.13.0 via TypeSafe API, question bank abstract-p1-20260923, 2026-09-23: 1 request, 5,842 input tokens.

What the abstract itself states, as probabilities: claims first 0.04 · states limitation 0.17 · human participants 0.02 · controlled comparison 0.29 · randomized assignment 0.01 · states quantitative result 0.98 · compares with prior results 0.50 · mentions code or data availability 0.02.

The abstract, sentence by sentence

arXiv listing abstract v1. Results and the main claim are underlined; .

This paper presents a joint calibration scheme for voltage (V) and current (I) probes that helps resolve accurately voltage-current phase differences even when the difference is very close to $90^{\circ}$. The latter has been a major issue with V-I probes when used with miniature RF plasma devices like the atmospheric pressure plasma jet (APPJ). Since the impedance of such miniature devices is predominantly capacitive, the phase difference between the voltage and current signals is very nearly $90^{\circ}$. It turns out, however, that when V-I probes are used with such devices without joint calibration, these frequently yield phase shifts over $90^{\circ}$. Also, since power absorption is proportional to the resistive part of the impedance it becomes very sensitive to the phase difference when it is close to $\approx90^{\circ}$. Thus, it is important to be able to resolve the phases accurately. Post-calibration, V-I probes would be indispensable for the electrical characterization of APPJs for determining average RF power $P_{av}$, plasma impedance $Z_p$, etc.Typical post-calibration V-I data yields, $Zp \approx 93.6 - j 1139 {\Omega} (81.5 - j 1173 {\Omega})$ at $P_{av} \approx 9.8 W$ $(\approx7.7 W)$ for helium (argon) gas.

background contribution method result interpretation main claim

What the classifier reports for each sentence. The number is the probability it gave the label.
#sentencerolethe authors' own workstates the main conclusionstrength
s0 This paper presents a joint calibration scheme for voltage (V) and current (I) probes that helps resolve accurately voltage-current phase differences even when the difference is very close to $90^{\circ}$. contribution 1.00 0.87 0.59 asserted 0.97
s1 The latter has been a major issue with V-I probes when used with miniature RF plasma devices like the atmospheric pressure plasma jet (APPJ). background 0.89 0.10 0.04 asserted 1.00
s2 Since the impedance of such miniature devices is predominantly capacitive, the phase difference between the voltage and current signals is very nearly $90^{\circ}$. background 1.00 0.18 0.03 asserted 1.00
s3 It turns out, however, that when V-I probes are used with such devices without joint calibration, these frequently yield phase shifts over $90^{\circ}$. background 0.74 0.59 0.05 asserted 0.91
s4 Also, since power absorption is proportional to the resistive part of the impedance it becomes very sensitive to the phase difference when it is close to $\approx90^{\circ}$. background 0.87 0.28 0.04 asserted 1.00
s5 Thus, it is important to be able to resolve the phases accurately. interpretation 0.56 0.11 0.11 asserted 0.99
s6 Post-calibration, V-I probes would be indispensable for the electrical characterization of APPJs for determining average RF power $P_{av}$, plasma impedance $Z_p$, etc.Typical post-calibration V-I data yields, $Zp \approx 93.6 - j 1139 {\Omega} (81.5 - j 1173 {\Omega})$ at $P_{av} \approx 9.8 W$ $(\approx7.7 W)$ for helium (argon) gas. result 0.98 0.85 0.17 asserted 0.99

What kind of paper the classifier reports this is: method or tool 1.00 · evidence from observation or experiment 0.98.

What was examined, and what was not

sources read
  • arXiv listing abstract v1
not examined
  • full text
  • figures and tables
  • references and citing sentences
  • relations to other papers
levels run
  • P abstract
not run
  • S sections
  • T body sentences
  • C citations
  • R cross-paper relations

Every quoted sentence on this record is the authors'. Every label is from Pith's closed vocabulary, with the probability shown beside it. Every number in the measurements table is the authors', checked by code. Record as JSON

The paper
Original abstract

This paper presents a joint calibration scheme for voltage (V) and current (I) probes that helps resolve accurately voltage-current phase differences even when the difference is very close to $90^{\circ}$. The latter has been a major issue with V-I probes when used with miniature RF plasma devices like the atmospheric pressure plasma jet (APPJ). Since the impedance of such miniature devices is predominantly capacitive, the phase difference between the voltage and current signals is very nearly $90^{\circ}$. It turns out, however, that when V-I probes are used with such devices without joint calibration, these frequently yield phase shifts over $90^{\circ}$. Also, since power absorption is proportional to the resistive part of the impedance it becomes very sensitive to the phase difference when it is close to $\approx90^{\circ}$. Thus, it is important to be able to resolve the phases accurately. Post-calibration, V-I probes would be indispensable for the electrical characterization of APPJs for determining average RF power $P_{av}$, plasma impedance $Z_p$, etc.Typical post-calibration V-I data yields, $Zp \approx 93.6 - j 1139 {\Omega} (81.5 - j 1173 {\Omega})$ at $P_{av} \approx 9.8 W$ $(\approx7.7 W)$ for helium (argon) gas.

Human review

Discussion

Know something about this paper? Leave a comment, or highlight any passage above to comment on it.