Measurement of reactor thermal neutron fluence of NTD-Ge by activation High-Purity Ge itself

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 2302.12982 v1 pith:KH5P52MQ submitted 2023-02-25 physics.ins-det hep-exnucl-ex

classification physics.ins-dethep-exnucl-ex
keywords neutronfluencethermaldetectoritselfntd-gereactorabsolute
verification ladder T0 review T1 audit T2 compute T3 formal

Signed reviews

No signed human review yet.

0 comments
abstract

In neutron transmutation doped germanium, the thermal neutron fluence of reactor irradiation is as high as 10$^{18}$~n$\cdot$cm$^{-2}$. For radiological safety reasons, general Co or Au neutron flux monitors cannot be easily used. We have experimentally demonstrated the feasibility of measuring the X-rays emitted by the NTD-Ge itself to determine the absolute thermal neutron fluence for the first time. A Micro-Megas Detector (MMD) and a Silicon Drift Detector (SDD) are set up to detect the tagging KX-rays with 9.2 keV and 10.3 keV cascading from the decays of $^{71}$Ge. Combined the detection efficiencies calculated by GEANT4, neutron fluence results given with proper accuracy by MMD and SDD are in a good agreement with each other.