Deep learning-enhanced chemiluminescence vertical flow assay for high-sensitivity cardiac troponin I testing
T0 review · 3 major / 4 minor · reviewed 2026-08-11 · deepseek-v4-flash
Pith's one-line read A paper-based chemiluminescence vertical flow assay claims a 0.16 pg/mL troponin detection limit and, on 66 blinded samples, matches an FDA-cleared analyzer at r = 0.984.
desk verdict Serious engineering, plausible clinical validation, but the headline LoD is an extrapolation below the lowest calibrator and needs real low-concentration measurements before the sensitivity claim holds. read the letter →
The pith
A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.
The reading
What carries the argument
Three engineered pieces carry the argument. The first is the conjugate: 15 nm gold nanoparticles decorated with PolyHRP-Streptavidin, a polymer that packs many horseradish-peroxidase enzymes per binding event, plus biotinylated anti-cTnI antibodies; the paper measures this as roughly a 210-fold signal gain over a conventional single-HRP conjugate and roughly a 3400-fold gain over the same conjugate's colorimetric readout. The second is the cartridge: a tray that transfers the sensing membrane from the absorbent-pad assembly used for the immunoassay and washing to a flat plastic support stage for imaging, so the chemiluminescence reagent stops flowing and the signal saturates stably; this change cut the imaging coefficient of variation from 22.9% to 4.6%. The third is the computational pipeline: one classifier network sorts each sample into a concentration band (below 40 pg/mL, 40–1000 pg/mL, or above 1000 pg/mL), three dedicated quantifier networks then report the concentration, and a sample is flagged as indeterminate if the classification and quantification stages disagree. The stated detection limit is computed by the standard formula $\mathrm{LoD} = \mathrm{LoB} + 1.645 \times \mathrm{SD}$, where the limit of blank is the mean blank signal plus 1.645 times its standard deviation, with both values converted to concentration through the calibration curve $y = 0.0147 x^{0.4043}$ fitted to spiked serum from 0.5 to $10^5$ pg/mL.
What would settle it
Spike cTnI-free serum at 0.1, 0.16, 0.2, 0.3, and 0.5 pg/mL, run each level in at least five replicates through the CL-VFA under the paper's protocol, and check whether the mean signals follow the calibration curve $y = 0.0147 x^{0.4043}$ and remain statistically separable from the blank. If the power law flattens or low-end variance grows below 0.5 pg/mL, the claimed limit of detection does not hold; if the signals track the curve, the extrapolation is confirmed.
Extended reading notes
Core claim
The paper's claim is that four components — a paper vertical-flow immunoassay, a polymerized-enzyme (PolyHRP) gold-nanoparticle conjugate, a Raspberry Pi-based chemiluminescence reader, and a cascaded set of four fully connected neural networks — bring laboratory-grade high-sensitivity troponin I measurement to a portable, low-cost format. Stated on the assay's own terms: it quantifies cTnI across six orders of magnitude with a limit of detection of 0.16 pg/mL, an average coefficient of variation below 15%, and blinded-test agreement with an FDA-approved clinical analyzer of $r = 0.984$ on 66 patient samples. The paper further claims that the neural-network architecture is what holds accuracy together across that range: the classifier-plus-three-quantifiers cascade outperformed single-model, power-fitting, random-forest, and logistic-regression baselines on the same blind set. The intended consequence is that the formal criteria for a high-sensitivity troponin assay — a coefficient of variation no worse than 10% at the 99th-percentile cutoff and the ability to detect low levels in more than half of healthy people — can be met without a benchtop instrument.
Load-bearing premise
The headline 0.16 pg/mL detection limit, computed in Section 2.5, is an extrapolation: the calibration curve $y = 0.0147 x^{0.4043}$ was fitted to spiked-serum signals from 0.5 to $10^5$ pg/mL, and the same fitted curve is used both to define the blank-equivalent intensity and to convert it into a concentration, with no sample near 0.16 pg/mL ever measured.
Editorial extensions
If this is right
- Point-of-care triage of suspected heart attacks becomes feasible: the paper argues its 0.16 pg/mL sensitivity supports a 1-hour confirmatory re-test, down from the usual 2–3 hours, and enables a 0-hour rule-out for patients with no initial troponin elevation.
- The cost structure changes the accessibility argument: roughly $4.25 per test at laboratory scale (projected below $1–2 at production scale) and about $222 for the reader, against benchtop analyzers that cost tens of thousands of dollars.
- The assay runs on 50 µL of serum per test and completes in 25 minutes, with computation adding less than 0.5 seconds per sample.
- Because the sensing membrane carries nine reaction spots by design, the same platform can be extended to multiplexed panels of cardiac biomarkers in a single test.
- The authors state that adding rapid plasma or serum extraction from whole blood is the planned next step for distributed clinics and other point-of-care sites.
Reading between the lines
- Because the 0.16 pg/mL figure is extrapolated below the lowest calibrator, the most direct next experiment is measuring spikes between 0.1 and 0.5 pg/mL; the assay's true low-end behavior is a standing prediction of the paper's calibration curve.
- If the extrapolation holds, the practical limit on sensitivity shifts from the enzyme chemistry to the variance of the blank and negative-control spots — a testable consequence of the paper's own finding that digital quality control and signal averaging materially change classification accuracy.
- The three blind-test samples near the clinical cut-off (ground truth 32–37 pg/mL) were read as 71–105 pg/mL, so the operating point of the classifier around the 40 pg/mL decision boundary is the place to look for clinically meaningful error before deployment.
- The platform's components are largely biomarker-agnostic, so swapping the antibody pair for a different cardiac protein (for example, NT-proBNP or D-dimer) is a cheap test of whether the same sensitivity gains transfer.
Signed reviews
Editorial analysis
A structured set of objections, weighed in public.
Referee Report
Summary. The paper reports a chemiluminescence vertical flow assay (CL-VFA) for cardiac troponin I (cTnI), combining a paper-based sensing membrane, an AuNP-PolyHRP detection conjugate, a Raspberry Pi-based portable reader, a tray-based cartridge for stable chemiluminescence imaging, and a cascade of four neural networks for concentration inference. The authors claim an LoD of 0.16 pg/mL, an average CV below 15%, a six-order-of-magnitude dynamic range, and blinded clinical agreement with an FDA-cleared analyzer (Pearson r = 0.984 on 66 samples). The assay operates on 50 µL of serum in 25 minutes with an estimated per-test cost of $4.25.
Significance. If the analytical claims hold, the engineering contributions are substantial: the tray-based cartridge reduces imaging CV from 22.9% to 4.6%, the portable reader outperforms a benchtop system in detection cut-off, and the blinded clinical evaluation against a clinical-grade analyzer is a strength. The paper also includes detailed assay protocols, a cost model, and a fair comparison between neural-network and power-law quantification. However, the headline LoD is not directly measured but extrapolated below the lowest calibrator, and the clinical validation does not quantitatively confirm low-concentration performance below 4 pg/mL. Therefore, the significance of the sensitivity claim is currently limited by the supporting evidence.
major comments (3)
- [Section 2.5] The LoD of 0.16 pg/mL is derived by extrapolating the power-law calibration y=0.0147x^0.4043, which is fitted to spiked serum from 0.5 to 10^5 pg/mL, down to concentrations below the lowest calibrator. The fitted curve has no blank-offset term, yet the reported mean blank signal is 0.0046, so the model is not constrained to describe the blank; inverting this curve to convert the LoB intensity into concentration is therefore an unvalidated extrapolation. In addition, the SDs used for LoB and LoD come from triplicate measurements (Figure 3c), well below the CLSI EP17 recommendation of at least 20 replicates, leaving the Gaussian assumption and SD stability near the blank untested. The abstract's 'detection limit of 0.16 pg/mL' and the 'order of magnitude' sensitivity claim are not supported by directly measured data; the authors should either measure samples near 0.16 pg/mL with an adequate number of replicates and report a formal limit of quantification, or explicitly present the value as a model-based extrapolation with uncertainty.
- [Section 2.6] The clinical validation does not provide quantitative confirmation of low-concentration performance. Samples with ground truth below 4 pg/mL are excluded from the Pearson r and CV calculations because they lack quantitative labels, and the DNNQ<40 training set assigns such samples a label of 4 pg/mL with an asymmetric MSLE loss that only penalizes overestimates. The blind test also shows that three samples with true cTnI concentrations of 32–37 pg/mL were misclassified and predicted as 71–105 pg/mL, indicating errors of roughly two- to threefold near the clinical threshold. Thus, the claim that the CL-VFA 'accurately measured cTnI concentrations in patient samples' at low levels is not established; the authors should report per-range agreement (e.g., Bland-Altman limits or percentage within ±20% for the 4–40 pg/mL range) and state clearly that no clinical samples below 4 pg/mL had quantitative validation.
- [Section 2.5] The statement that the CL-VFA surpasses traditional benchtop analyzers in sensitivity by an order of magnitude depends on the chosen comparator. The Beckman Access 2 analyzer used for ground truth has a quantification cutoff of 4 pg/mL, but many hs-cTnI assays report LoDs of roughly 1–2 pg/mL; comparing the extrapolated 0.16 pg/mL value to 4 pg/mL conflates LoD with LoQ and does not demonstrate an order-of-magnitude advantage over current hs-cTnI benchtop assays. Please compare against published LoD/LoQ values of specific FDA-cleared hs-cTnI assays and provide uncertainty bounds for the extrapolated LoD.
minor comments (4)
- [Methods] The text repeatedly contains 'm м' (e.g., '10 m м borate buffer'), which should be 'mM'; this encoding artifact appears throughout the Methods and should be corrected.
- [Equation (2)] The normalized signal formula INormalized = 1 - (216-1 - XTest)/(216-1 - Xneg) is ambiguous: the notation '216-1' presumably means 2^16 - 1, and the subtraction order should be clarified.
- [Figure 3] The text states R² = 0.99 for the calibration curve in Figure 3b, while the figure caption reports R2 = 0.9929; please make the reported values consistent.
- [Methods] The description of batch standardization layers in the neural networks should specify whether these are batch normalization layers, and the use of dropout with batch normalization should be justified or clarified.
Circularity Check
No significant circularity: LoD is a standard calibration-curve conversion and clinical validation uses an external FDA-cleared analyzer.
full rationale
The central derivation chain is not circular. In Section 2.5, the LoD is computed as LoB + 1.645 x SD of the lowest cTnI measurement, with LoB from the mean blank plus 1.645 x SD of the blank; the measured blank and low-sample intensities are then converted to concentrations using the calibration curve y=0.0147x^0.4043 fitted to spiked serum from 0.5 to 1e5 pg/mL. This is standard calibration use, not a definitional reduction: the LoD is derived from measured blank and low-end signal variability, and the same fitted curve is used merely to express those signal-level statistics in concentration units. The curve is not defined in terms of the LoD, and the LoD is not an input to the fit. The concern that the 0.16 pg/mL value lies below the lowest calibrator is an extrapolation and statistical robustness issue, not circularity. In Section 2.6, the neural-network predictions are validated against an external gold standard (Beckman Access 2, FDA-cleared), with 66 blinded samples not used in training; samples with ground truth below 4 pg/mL are excluded from Pearson's r and CV calculations, so the correlation is not self-referential. Self-citations to prior VFA hardware, conjugate, and cartridge work support components of the platform but are not load-bearing for the headline LoD or the clinical correlation. No equation reduces to its own inputs, and no fitted parameter is renamed as a prediction. Thus no significant circularity is present.
Assumptions & free parameters
free parameters (5)
- LoD calibration power-law coefficients =
a=0.0147, b=0.4043 in y=a*x^b
- Clinical correlation power-law coefficients =
a=0.0048, b=0.5
- Cascade concentration range borders =
40 pg/mL and 1000 pg/mL
- Outlier exclusion threshold =
>=1 order of magnitude deviation from trendline
- Negative-control QC confidence interval =
95% CI (1.96 SD)
assumptions (6)
- standard math Blank and low-concentration signals are approximately Gaussian, supporting the parametric LoD formula (mean blank + 1.645 SD).
- domain assumption cTnI spiked into cTnI-free human serum behaves like endogenous clinical cTnI.
- domain assumption The Beckman Access 2 hs-cTnI analyzer provides accurate quantitative ground truth above 4 pg/mL.
- domain assumption Remnant serum samples stored at -80C and thawed once retain their original cTnI concentration.
- domain assumption Training, validation, and blind-test CL-VFA samples are statistically independent.
- domain assumption Non-specific binding and serum interferents (hemolysis, lipemia, autoantibodies) do not systematically affect test spot signals.
Cite this review
Pith. "Pith review of Deep learning-enhanced chemiluminescence vertical flow assay for high-sensitivity cardiac troponin I testing." pith.science (2026). https://pith.science/paper/SWY77IIG
@misc{pith2026241208945,
author = {Pith},
title = {Pith review of: Deep learning-enhanced chemiluminescence vertical flow assay for high-sensitivity cardiac troponin I testing},
year = {2026},
howpublished = {\url{https://pith.science/paper/SWY77IIG}},
note = {Machine review of arXiv:2412.08945}
}
read the original abstract
Democratizing biomarker testing at the point-of-care requires innovations that match laboratory-grade sensitivity and precision in an accessible format. Here, we demonstrate high-sensitivity detection of cardiac troponin I (cTnI) through innovations in chemiluminescence-based sensing, imaging, and deep learning-driven analysis. This chemiluminescence vertical flow assay (CL-VFA) enables rapid, low-cost, and precise quantification of cTnI, a key cardiac protein for assessing heart muscle damage and myocardial infarction. The CL-VFA integrates a user-friendly chemiluminescent paper-based sensor, a polymerized enzyme-based conjugate, a portable high-performance CL reader, and a neural network-based cTnI concentration inference algorithm. The CL-VFA measures cTnI over a broad dynamic range covering six orders of magnitude and operates with 50 uL of serum per test, delivering results in 25 min. This system achieves a detection limit of 0.16 pg/mL with an average coefficient of variation under 15%, surpassing traditional benchtop analyzers in sensitivity by an order of magnitude. In blinded validation, the computational CL-VFA accurately measured cTnI concentrations in patient samples, demonstrating a robust correlation against a clinical-grade FDA-cleared analyzer. These results highlight the potential of CL-VFA as a robust diagnostic tool for accessible, rapid cardiac biomarker testing that meets the needs of diverse healthcare settings, from emergency care to underserved regions.
Figures
Reference graph
Works this paper leans on
-
[1]
World Health Organization, World health statistics 2023: monitoring health for the SDGs, sustainable development goals, https://www.who.int/publications/i/item/9789240074323, accessed: November, 2023
arXiv 2023
-
[2]
C. W. Tsao, A. W. Aday, Z. I. Almarzooq, C. A. M. Anderson, P. Arora, C. L. Avery, C. M. Baker -Smith, A. Z. Beaton, A. K. Boehme, A. E. Buxton, Y. Commodore -Mensah, M. S. V. Elkind, K. R. Evenson, C. Eze-Nliam, S. Fugar, G. Generoso, D. G. Heard, S. Hiremath, J. E. Ho, R. Kalani, D. S. Kazi, D. Ko, D. A. Levine, J. Liu, J. Ma, J. W. Magnani, E. D. Micho...
work page 2023
- [3]
-
[4]
A. Gheorghe, U. Griffiths, A. Murphy, H. Legido-Quigley, P. Lamptey, P. Perel, BMC Public Health 2018, 18, 975
work page 2018
-
[5]
X. Du, X. Su, W. Zhang, S. Yi, G. Zhang, S. Jiang, H. Li, S. Li, F. Xia, Anal. Chem. 2022, 94, 442
work page 2022
-
[6]
J. L. Anderson, E. W. Campion, D. A. Morrow, N. Engl. J. Med. 2017, 376, 2053
work page 2017
-
[7]
K. C. Park, D. C. Gaze, P. O. Collinson, M. S. Marber, Cardiovasc. Res. 2017, 113, 1708
work page 2017
- [8]
Show all 64 references
-
[9]
Westermann, J
D. Westermann, J. T. Neumann, N. A. Sörensen, S. Blankenberg, Nat. Rev. Cardiol. 2017, 14, 472
2017
-
[10]
Thygesen, J
K. Thygesen, J. S. Alpert, A. S. Jaffe, B. R. Chaitman, J. J. Bax, D. A. Morrow, H. D. White, Circulation 2018, 138, e618
2018
-
[11]
A. H. B. Wu, R. H. Christenson, D. N. Greene, A. S. Jaffe, P. A. Kavsak, J. Ordonez -Llanos, F. S. Apple, Clin. Chem. 2018, 64, 645
2018
-
[12]
J. T. Neumann, R. Twerenbold, F. Ojeda, N. A. Sörensen, A. R. Chapman, A. S. V. Shah, A. Anand, J. Boeddinghaus, T. Nestelberger, P. Badertscher, A. Mokhtari, J. W. Pickering, R. W. Troughton, J. Greenslade, W. Parsonage, M. Mueller-Hennessen, T. Gori, T. Jernberg, N. Morris, ...
2019
-
[13]
Farmakis, C
D. Farmakis, C. Mueller, F. S. Apple, Eur. Heart J. 2020, 41, 4050
2020
-
[14]
F. S. Apple, C. R. Fantz, P. O. Collinson, t. I. C. o. C. A. o. C. Bio-Markers, Clin. Chem. 2020, 67, 70
2020
-
[15]
P. K. Drain, E. P. Hyle, F. Noubary, K. A. Freedberg, D. Wilson, W. R. Bishai, W. Rodriguez, I. V. Bassett, Lancet Infect. Dis. 2014, 14, 239
2014
-
[16]
A. R. Chapman, S. Stewart, N. L. Mills, Heart 2019, 105, 740
2019
-
[17]
I. V. L. Thulin, G. M. S. Myrmel, S. M. F. Jordalen, P. Collinson, K. M. Aakre, J. Lab. Precis. Med. 2023, 8
2023
-
[18]
J. A. Leigh, M. Alvarez, C. J. Rodriguez, Curr. Atheroscler. Rep. 2016, 18, 9
2016
-
[19]
Suzuki, K
K. Suzuki, K. Komukai, K. Nakata, R. Kang, Y. Oi, E. Muto, Y. Kashiwagi, M. Tominaga, S. Miyanaga, T. Ishikawa, K. Okuno, M. Uzura, M. Yoshimura, Intern. Med. 2018, 57, 1673
2018
-
[20]
Regan, R
B. Regan, R. O’Kennedy, D. Collins, in Advances in Clinical Chemistry, Vol. 104 (Ed: G. S. Makowski), Elsevier, Netherland 2021, Ch. 1
2021
-
[21]
Cullen, P
L. Cullen, P. O. Collinson, E. Giannitsis, Emerg. Med. 2022, 39, 861
2022
-
[22]
F. V. Plapp, A. Dasgupta, Ann. Clin. Lab. Sci. 2023, 53, 322
2023
-
[23]
W. F. Peacock, D. Diercks, R. Birkhahn, A. J. Singer, J. E. Hollander, R. Nowak, B. Safdar, C. D. Miller, M. Peberdy, F. Counselman, A. Chandra, J. Kosowsky, J. Neuenschwander, J. Schrock, E. Lee - Lewandrowski, W. Arnold, J. Nagurney, Ann. Lab. Med. 2016, 36, 405
2016
-
[24]
Xiong-Hang, K
K. Xiong-Hang, K. Schulz, Y. Sandoval, S. W. Smith, A. K. Saenger, F. S. Apple, Clin. Biochem. 2023, 114, 79
2023
-
[25]
W. U. Dittmer, T. H. Evers, W. M. Hardeman, W. Huijnen, R. Kamps, P. de Kievit, J. H. Neijzen, J. H. Nieuwenhuis, M. J. Sijbers, D. W. Dekkers, M. H. Hefti, M. F. Martens, Clin. Chim. Acta. 2010, 411, 868
2010
-
[26]
D. W. Kemper, V. Semjonow, F. de Theije, D. Keizer, L. van Lippen, J. Mair, B. Wille, M. Christ, F. Geier, P. Hausfater, D. Pariente, V. Scharnhorst, J. Curvers, J. Nieuwenhuis, Clin. Biochem. 2017, 50, 174
2017
-
[27]
A. H. Wu, A. Smith, R. H. Christenson, M. M. Murakami, F. S. Apple, Clin. Chim. Acta. 2004, 346, 211
2004
-
[28]
Osredkar, K
J. Osredkar, K. Krivic, T. Fabjan, K. Kumer, J. Tršan, L. Poljančič, P. Finderle, H. Možina, Med. Access Point Care 2021, 5, 23992026211055095
2021
-
[29]
Ahmed, L
S. Ahmed, L. Jafri, A. Raheem, S. Shakeel, I. Siddiqui, Cureus 2019, 11, e4231
2019
-
[30]
Han, M.-G
G.-R. Han, M.-G. Kim, Sensors 2020, 20, 2593. 24
2020
-
[31]
M. Song, J. Xing, H. Cai, X. Gao, C. Li, C. Liu, X. Li, X. Fu, S. Ding, W. Cheng, R. Chen, ACS Nano 2023, 17, 10748
2023
-
[32]
G.-R. Han, H. Ki, M.-G. Kim, ACS Appl. Mater. Interfaces 2020, 12, 1885
2020
-
[33]
G.-R. Han, H. J. Koo, H. Ki, M.-G. Kim, ACS Appl. Mater. Interfaces 2020, 12, 34564
2020
-
[34]
Khachornsakkul, W
K. Khachornsakkul, W. Dungchai, ACS Sens. 2021, 6, 1339
2021
-
[35]
T. Zhan, Y. Su, W. Lai, Z. Chen, C. Zhang, Biosens. Bioelectron. 2022, 214, 114494
2022
-
[36]
Hong, E.-J
D. Hong, E.-J. Jo, K. Kim, M.-B. Song, M.-G. Kim, Small 2020, 16, 2004535
2020
-
[37]
F. Li, L. Guo, Y. Hu, Z. Li, J. Liu, J. He, H. Cui, Talanta 2020, 207, 120346
2020
-
[38]
R. Yang, F. Li, W. Zhang, W. Shen, D. Yang, Z. Bian, H. Cui, Anal. Chem. 2019, 91, 13006
2019
-
[39]
I. H. Cho, E. H. Paek, Y. K. Kim, J. H. Kim, S. H. Paek, Anal. Chim. Acta. 2009, 632, 247
2009
-
[40]
G.-R. Han, A. Goncharov, M. Eryilmaz, H. -A. Joung, R. Ghosh, G. Yim, N. Chang, M. Kim, K. Ngo, M. Veszpremi, K. Liao, O. B. Garner, D. Di Carlo, A. Ozcan, ACS Nano 2024, 18, 27933
2024
-
[41]
Cheng, Y
K. Cheng, Y. Ding, C. Liu, Y. Ding, S. Xie, X. Zhu, H. Liu, W. Yue, Microfluid.Nanofluidics 2023, 27, 39
2023
-
[42]
H. Zhao, Q. Lin, L. Huang, Y. Zhai, Y. Liu, Y. Deng, E. Su, N. He, Nanoscale 2021, 13, 3275
2021
-
[43]
J. Liu, L. Zhang, Y. Wang, Y. Zheng, S. Sun, Measurement 2014, 47, 200
2014
-
[44]
D. Sun, Z. Luo, J. Lu, S. Zhang, T. Che, Z. Chen, L. Zhang, Biosens. Bioelectron. 2019, 134, 49
2019
-
[45]
Sarangadharan, S.-L
I. Sarangadharan, S.-L. Wang, R. Sukesan, P.-c. Chen, T.-Y. Dai, A. K. Pulikkathodi, C. -P. Hsu, H.-H. K. Chiang, L. Y.-M. Liu, Y.-L. Wang, Anal. Chem. 2018, 90, 2867
2018
-
[46]
Y. Wang, T. Liu, M. Yang, C. Wu, W. Zhang, Z. Chu, W. Jin, Biosens. Bioelectron. 2021, 193, 113554
2021
-
[47]
X. Yang, Y. Zhao, L. Sun, H. Qi, Q. Gao, C. Zhang, Sensor. Actuat. B-Chem 2018, 257, 60
2018
-
[48]
W. Xu, L. Wang, R. Zhang, X. Sun, L. Huang, H. Su, X. Wei, C. -C. Chen, J. Lou, H. Dai, K. Qian, Nat. Commun. 2020, 11, 1654
2020
-
[49]
Westwood, B
M. Westwood, B. Ramaekers, S. Grimm, G. Worthy, D. Fayter, N. Armstrong, T. Buksnys, J. Ross, M. Joore, J. Kleijnen, Health Technol. Assess. 2021, 25, 1
2021
-
[50]
International Federation of Clinical Chemistry and Laboratory Medicine, High-Sensitivity Cardiac Troponin I and T Assay Analytical Characteristics Designated by Manufacturer. IFCC Committee on Clinical Applications of Cardiac Bio -Markers, https://ifccfiles.com/2024/03/High-Se...
2024
-
[51]
M. Yang, J. Huang, J. Fan, J. Du, K. Pu, X. Peng, Chem. Soc. Rev. 2020, 49, 6800
2020
-
[52]
H. A. Joung, Z. S. Ballard, A. Ma, D. K. Tseng, H. Teshome, S. Burakowski, O. B. Garner, D. Di Carlo, A. Ozcan, Lab Chip 2019, 19, 1027
2019
-
[53]
Z. S. Ballard, H. A. Joung, A. Goncharov, J. Liang, K. Nugroho, D. Di Carlo, O. B. Garner, A. Ozcan, NPJ Digit. Med. 2020, 3, 66
2020
-
[54]
Goncharov, H
A. Goncharov, H. A. Joung, R. Ghosh, G. R. Han, Z. S. Ballard, Q. Maloney, A. Bell, C. T. Z. Aung, O. B. Garner, D. D. Carlo, A. Ozcan, Small 2023, 19, e2300617
2023
-
[55]
Eryilmaz, A
M. Eryilmaz, A. Goncharov, G. R. Han, H. A. Joung, Z. S. Ballard, R. Ghosh, Y. Zhang, D. Di Carlo, A. Ozcan, ACS Nano 2024, 18, 16819
2024
-
[56]
Ghosh, H.-A
R. Ghosh, H.-A. Joung, A. Goncharov, B. Palanisamy, K. Ngo, K. Pejcinovic, N. Krockenberger, E. J. Horn, O. B. Garner, E. Ghazal, A. O’Kula, P. M. Arnaboldi, R. J. Dattwyler, A. Ozcan, D. Di Carlo, Nat. Commun. 2024, 15, 7124
2024
-
[57]
H. J. Jang, H. A. Joung, A. Goncharov, A. G. Kanegusuku, C. W. Chan, K. J. Yeo, W. Zhuang, A. Ozcan, J. Chen, ACS Nano 2024, 18, 24792
2024
-
[58]
Joung, Z
H.-A. Joung, Z. S. Ballard, J. Wu, D. K. Tseng, H. Teshome, L. Zhang, E. J. Horn, P. M. Arnaboldi, R. J. Dattwyler, O. B. Garner, D. Di Carlo, A. Ozcan, ACS Nano 2020, 14, 229
2020
-
[59]
N. M. Green, Adv. Protein Chem. 1975, 29, 85
1975
-
[60]
D. A. Armbruster, T. Pry, Clin. Biochem. Rev. 2008, 29 Suppl 1, S49
2008
-
[61]
D. S. Herman, P. A. Kavsak, D. N. Greene, Am. J. Clin. Pathol. 2017, 148, 281
2017
-
[62]
J. Mair, O. Hammarsten, J. Lab. Precis. Med. 2023, 8
2023
-
[63]
M. T. H. Lowry, A. Anand, N. L. Mills, Heart 2021, 107, 1912
2021
-
[64]
J. W. Pickering, J. M. Young, P. M. George, A. S. Watson, S. J. Aldous, R. W. Troughton, C. J. Pemberton, A. M. Richards, L. A. Cullen, M. P. Than, JAMA Cardiol. 2018, 3, 1108
2018
Reviewed August 11, 2026 · model on record in the stance chip above.
Discussion (0). Continue with ORCID to comment.