Pith. sign in

REVIEW

NICMOS Status

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv astro-ph/0604394 v1 pith:575LNB7X submitted 2006-04-19 astro-ph

NICMOS Status

classification astro-ph
keywords nicmosnon-linearityinstrumentbehaviorcalibrationcountdependentdescribe
verification ladder T0 review T1 audit T2 compute T3 formal T4 reserved
0 comments
Share X Bluesky LinkedIn Reddit HN
read the original abstract

We provide an overview of the most important calibration aspects of the NICMOS instrument on board of HST. We describe the performance of the instrument after the installation of the NICMOS Cooling System, and show that the behavior of the instrument has become very stable and predictable. We detail the improvements made to the NICMOS pipeline and outline plans for future developments. The derivation of the absolute photometric zero-point calibration is described in detail. Finally, we describe and quantify a newly discovered count-rate dependent non-linearity in the NICMOS cameras. This new non-linearity is distinctly different from the total count dependent non-linearity that is well known for near-infrared detectors. We show that the non-linearity has a power law behavior, with pixels with high system, or vice versa, pixels with low count rate detecting slightly less than expected. The effect has a wavelength dependence with observations at the shortest wavelengths being the most affected (~0.05-0.1 mag per dex flux change at ~1 micron, 0.03 mag per dex at 1.6 micron).

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.