Effectiveness Study of Calibration and Correction Algorithms on the Prototype of the POLAR-2/LPD Detector
Not yet reviewed by Pith; the record is open.
This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.
SPECIMEN: schema-true, not a live event
T0 review · schema-true
One-sentence machine reading of the paper's core claim.
pith:XXXXXXXX · record.json · timestamp
Signed reviews
read the original abstract
Gaseous X-ray polarimetry refers to a class of detectors used for measuring the polarization of soft X-rays. The systematic effects of such detectors introduce residual modulation, leading to systematic biases in the polarization detection results of the source. This paper discusses the systematic effects and their calibration and correction using the Gas Microchannel Plate-Pixel Detector (GMPD) prototype for POLAR-2/Low-Energy X-ray Polarization Detector (LPD). Additionally, we propose an algorithm that combines parameterization with Monte Carlo simulation and Bayesian iteration to eliminate residual modulation. The residual modulation after data correction at different energy points has been reduced to below 1%, and a good linear relationship is observed between the polarization degree and modulation degree. The improvement in modulation degree after correction ranges from 2% to 15%, and the results exceed those of the Imaging X-Ray Polarimetry Explorer (IXPE) above 5 keV.