Pulse shape discrimination performance of Inverted Coaxial Ge detectors

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 1711.01433 v1 pith:E4R2POZP submitted 2017-11-04 physics.ins-det hep-exnucl-ex

classification physics.ins-dethep-exnucl-ex
keywords betadecaydetectorscoaxialexperimentsinvertedadvantagebackground
verification ladder T0 review T1 audit T2 compute T3 formal
0 comments
abstract

We report on the characterization of two inverted coaxial Ge detectors in the context of being employed in future $^{76}$Ge neutrinoless double beta ($0\nu\beta\beta$) decay experiments. It is an advantage that such detectors can be produced with bigger Ge mass as compared to the planar Broad Energy Ge detectors (BEGe) that are currently used in the GERDA $0\nu\beta\beta$ decay experiment. This will result in lower background for the search of $0\nu\beta\beta$ decay due to a reduction of cables, electronics and holders. The measured resolution near the $^{76}$Ge Q-value at 2039 keV is 2.5 keV and their pulse-shape characteristics are similar to BEGe-detectors. It is concluded that this type of Ge-detector is suitable for usage in $^{76}$Ge $0\nu\beta\beta$ decay experiments.