3D-Printed Enclosure Wire-Guided Liquid Microfilm for Versatile Spectroscopy

T0 review · 3 major / 5 minor · reviewed 2026-08-06 · deepseek-v4-flash

Pith's one-line read A metal 3D-printed enclosure can replace machined liquid-microfilm sample holders, delivering tunable 25–180 µm films stable for ten hours and usable in Raman, fluorescence, and two-photon absorption spectroscopy.

desk verdict Useful wire-guided film design, but the reproducibility claim rests on a single 3D-printed part. read the letter →

arxiv 2507.02696 v1 pith:EYSE6UHV submitted 2025-07-03 physics.chem-ph

classification physics.chem-ph
keywords wire-guidedliquidmicrofilm3D-printedspectroscopyinstrumentmetal3DprintingRamanfluorescencetwo-photonabsorptionwindowlesssampleBeer–Lambertthicknessmeasurement
verification ladder T0 review T1 audit T2 compute T3 formal

The pith

A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.

The reading

This paper tries to establish that a wire-guided liquid microfilm sample holder — normally a precision-machined instrument — can be replaced by a metal 3D-printed enclosure that any lab can fabricate from a supplied CAD file. The authors report that the resulting microfilm is linearly tunable in thickness from 25 to 180 $\mu$m by pump speed, keeps temporal thickness variation below 1.0% at low flow rates, is spatially homogeneous over an optically useful area of roughly $15\ \mathrm{mm}^2$, and does not break during ten hours of operation. They then show the same device records continuous-wave Raman spectra of three alcohols, fluorescence of Rhodamine B with 50 ms integration, and two-photon-excited fluorescence whose power dependence has order $n = 2.05 \pm 0.03$. The payoff would matter because windowless flowing films avoid the absorption, dispersion, and damage that cuvette windows introduce under intense or low-frequency light, and a printable design could make such films routine in laboratories without machine-shop access.

What carries the argument

The central object is the wire-guided liquid microfilm: a thin liquid sheet held between two thin wires by surface tension, with no windows or container walls in the optical path. The load-bearing part of this design is the 3D-printed metal enclosure, whose internal flow pathway delivers liquid to the wires in a way that the authors say removes most pump pressure and oscillation noise before the liquid descends by gravity; they note this geometry would be hard to make by conventional machining. The supporting components are a low-pressure microannular gear pump providing 0.03–18 mL/min, a 40 $\mathrm{cm}^3$ sample cylinder used as a damper, and a photodiode readout of potassium ferricyanide absorbance at 405 nm, which converts Beer–Lambert transmission into film thickness in micrometers. The enclosure is what transforms an established technique into a replicable one: printed affordably by an external contractor in AlSi10Mg, it needs no custom machining and can be replaced by printing a new copy.

What would settle it

Print two or more enclosures from the supplied CAD file at different vendors using the same AlSi10Mg metal and post-processing, then measure film thickness and temporal deviation at the same pump speeds with the same pump, damper, and tubing. If the thickness-versus-speed curves differ between enclosures by more than the paper's reported run-to-run scatter, or if one enclosure fails to sustain a ten-hour film, the central reproducibility claim would be disproven.

Watch

Extended reading notes

Core claim

On its own terms, the central claim is that a single aluminum (AlSi10Mg) 3D-printed body, combined with a microannular gear pump, a 40 $\mathrm{cm}^3$ damping cylinder, and two 0.003-inch tungsten wires, is sufficient to form a free-standing liquid microfilm with a thickness that increases linearly with pump speed from 25 to 180 $\mu$m. The temporal standard deviation of the thickness is below 1.0% at the low end of the flow range, a central $1\ \mathrm{mm}^2$ region has a mean thickness of 66 $\mu$m with 4.0% pixel-to-pixel variation, and the film ran for ten hours at 2500 RPM without breakage. The same apparatus produced Raman spectra of isopropanol, ethanol, and methanol; a fluorescence spectrum of 10 ppm Rhodamine B; and two-photon photoluminescence with a fitted power-law order of $n = 2.05 \pm 0.03$. The stated advantage is reproducibility: because the enclosure is printed rather than machined, the authors argue that any researcher can download the CAD file and reproduce the instrument affordably and quickly, including replacing a damaged enclosure with an identical copy.

Load-bearing premise

The reproducibility claim assumes that the single aluminum enclosure printed by one external contractor is representative of what any other lab would get from the same CAD file; no second enclosure, inter-part comparison, or lab-to-lab test is reported.

Editorial extensions

If this is right

  • Any lab with access to metal 3D-printing can build a wire-guided liquid microfilm spectrometer from the downloadable CAD file plus a pump, damper, tubing, and thin tungsten wires.
  • Film thickness is set predictably by pump speed over 25–180 $\mu$m, so a user can dial in a desired pathlength without a separate thickness monitor.
  • The film is stable enough for long and spatially resolved measurements: sub-1% temporal deviation at low flow, a large homogeneous area, and ten hours of continuous operation at 2500 RPM.
  • The same printed holder works for linear Raman, one-photon fluorescence, and nonlinear two-photon absorption, including organic solvents and focused femtosecond pulses.
  • Because the body is printed metal, a damaged part can be replaced quickly and the construction metal can in principle be matched to the chemical compatibility of the sample.

Reading between the lines

Editorial extensions of the paper, not claims the author makes directly.

  • Inference: if the reproducibility holds across builds, the main adoption barrier for windowless flowing-sample spectroscopy becomes access to a printer rather than machine-shop expertise; this could carry wire-guided films into terahertz, X-ray, and high-field nonlinear experiments in ordinary labs.
  • Inference: an obvious next test the paper does not report is printing several enclosures from the same CAD file at different vendors and comparing thickness-versus-pump-speed curves; that comparison would directly test the reproducibility claim on which the design's value rests.
  • Inference: the Beer–Lambert thickness readout used for characterization could be repurposed as a live feedback signal, adjusting pump speed to hold the film at a target thickness during long measurements.
  • Inference: the same printed-flow-path strategy could be extended to flat liquid jets and other windowless sample geometries, where rapid CAD iteration would let groups optimize flow stability for their own solvents and pump hardware.
Share X Bluesky LinkedIn Reddit HN

Editorial analysis

A structured set of objections, weighed in public.

Desk editor's note, referee report, and a circularity audit.

Referee Report

3 major / 5 minor

Summary. The paper presents a metal 3D-printed enclosure for producing wire-guided liquid microfilms for spectroscopy. The device is characterized optically: film thickness is measured by Beer-Lambert absorption of potassium ferricyanide at 405 nm as a function of gear-pump speed, giving a tunable range of 25–180 µm with temporal thickness deviations below 1.0% at low flow rates, a claim of 10-hour stability at 2500 RPM, and a 2D spatial thickness map showing a central region with 4.0% pixel-to-pixel standard deviation. The authors then demonstrate Raman spectroscopy of three alcohols, fluorescence of Rhodamine B, and two-photon-excited photoluminescence with a fitted power order of n = 2.05 ± 0.03. The central marketing claim is that the device is highly reproducible because the enclosure is 3D-printed from an included CAD file with no custom-machined parts.

Significance. If the reproducibility claim holds, the device would be a useful, low-barrier addition to the spectroscopy toolkit: it removes windows from the optical path, recycles the sample, and can be adopted without a machine shop. The paper has tangible strengths: the Beer-Lambert thickness method is standard and self-consistent; the static thickness-versus-pump-speed data are plausible; the two-photon order is close to the expected value of 2; and the three spectroscopy demonstrations are direct evidence of versatility. The inclusion of the CAD file is a practical contribution. However, the headline reproducibility advantage currently rests on a single enclosure from one external contractor, and the quantitative thickness claims lack stated uncertainty in the extinction coefficient. These issues are load-bearing for the abstract's central claims and need to be addressed before the paper can be recommended for publication.

major comments (3)
  1. [II (Apparatus) and V (Conclusion)] The abstract and conclusion advertise 'highly reproducible' as a unique advantage of the design, but the evidence consists of a single AlSi10Mg enclosure printed by one external contractor. No second enclosure, no inter-part comparison of thickness or flow stability, and no lab-to-lab test is reported. Since film thickness and stability depend on the internal flow geometry shown in Fig. 1(b) and on the wire-guide assembly, build-to-build variation in channel dimensions, surface roughness, or post-processing could alter the tuning curve and the <1.0% deviation claim. This missing evidence is load-bearing for the reproducibility claim; at a minimum, one additional independently printed enclosure should be characterized and compared with the first.
  2. [III.A (Static Position Measurements)] Every quantitative thickness value in the paper—including the 25–180 µm range, the <1.0% temporal deviations, and the spatial map in §III.B—is computed from an in-house molar absorption coefficient of 990 M^-1 cm^-1 for potassium ferricyanide at 405 nm. The paper gives no uncertainty for this coefficient and provides no raw absorbance data. Because the Beer-Lambert inversion makes thickness inversely proportional to ε, a 5% error in ε would shift every quoted thickness by 5% and also change the reported percentage deviations. The authors should report the calibration statistics (replicates, standard deviation, and any concentration dependence) and ideally cross-check against a literature value.
  3. [III.A (Stability statement)] The statement that the microfilm 'ran with a pump speed of 2500 RPM for ten hours with no observed breakage' is presented without the number of replicate runs, the breakage criterion, or a continuous record of the photodiode signal. The abstract elevates ten-hour stability to a headline result. The sentence begins with 'over multiple trials,' but the specific ten-hour observation appears to be a single run. Please clarify the number of runs and define 'no observed breakage' operationally, for example as a continuous transmission signal within a stated tolerance.
minor comments (5)
  1. [Fig. 2(a)] The text and abstract state that the thickness range corresponds to pump speeds of 1500–5500 RPM, but the horizontal axis in Fig. 2(a) ends at 5000 RPM. Please verify the highest data point and the axis label.
  2. [Fig. 3(a)] The text reports a mean thickness of 66 µm with a standard deviation of 4.0% between 50 µm by 50 µm pixels. The abstract describes the film as 'spatially homogeneous'; please clarify how the 4.0% pixel-to-pixel variation is consistent with the qualitative description of homogeneity.
  3. [III.A] The thickness measurements are pooled from nine repetitions at three chromophore concentrations (30, 40, and 50 mM). Please state whether the derived thickness was concentration-independent within the measurement uncertainty, since this is a check on the Beer-Lambert analysis.
  4. [References] References 17 and 25 are formatted incompletely (one is listed as 'Tech. Rep. (2010)' and one has a bare DOI). Please complete the bibliographic details.
  5. [Supplementary materials] The CAD file and the extinction-coefficient measurement are said to be in the supplementary materials, but the manuscript does not state the file format, printing parameters (layer height, orientation, material grade), or post-processing steps. Adding these details would materially improve reproducibility.

Circularity Check

0 steps flagged · score 0.0 of 10

No circularity: the quantitative claims are direct measurements or fits, not reductions to the paper's own inputs.

full rationale

The paper's central quantitative results are empirical characterizations of a single device, not predictions derived from an assumed model. Film thickness is obtained from the Beer-Lambert law using a molar absorption coefficient measured independently in a 2 mm pathlength cuvette (Section III), so the thickness values are calibrated measurements rather than self-defined outputs. The temporal and spatial thickness statistics are computed directly from photodiode transmission time series and a 2D scan; there is no fitted parameter that is later relabeled as a predicted quantity. The two-photon order n = 2.05 ± 0.03 is a fit to the observed power dependence used to validate the expected quadratic response, not an input that the spectroscopic demonstration depends on. The reproducibility claim about 3D printing is a generalization from one aluminum enclosure produced by an external contractor (Section II), which is a real external-validity limitation, but it is not a circularity: the claim does not reduce to its own definition or to a fitted parameter. No load-bearing self-citation, imported uniqueness theorem, or ansatz smuggled in via citation appears in the manuscript. Therefore no specific circular step meeting the required evidence standard can be exhibited, and the appropriate score is 0.

Assumptions & free parameters 0 free parameters · 4 assumptions · 0 invented entities

The central claims rest on standard optical assumptions plus the unverified reproducibility of the 3D-printed part. No free parameters or invented entities are introduced; the main calibration input, the extinction coefficient, is measured in-house rather than taken as a free fit parameter.

assumptions (4)
  • domain assumption Beer-Lambert law applies to the wire-guided thin film using the extinction coefficient measured in a 2-mm cuvette.
    Section III computes film thickness from photodiode absorbance using epsilon = 990 M^-1 cm^-1 at 405 nm for potassium ferricyanide at 30-50 mM; any concentration or film-geometry dependence would shift all reported thicknesses.
  • domain assumption The sample concentration stays constant during recirculation and over the 10-hour run.
    The apparatus appears to recycle sample from a beaker (Fig. 1e); no evaporation compensation or concentration monitoring is reported, yet thickness is inferred from absorbance.
  • domain assumption The photodiode signal with the chromophore, referenced to signal without chromophore, isolates absorption in the film.
    Section III.A describes only this ratio; scattering by the wires, stray light, and film curvature are not analyzed.
  • ad hoc to paper A single 3D-printed enclosure from one contractor is representative of the design for other users.
    The reproducibility claim in the Abstract and Conclusion rests on the CAD file, but no second build or inter-part variation is tested; Section II notes post-processing by an external contractor without specification.

how reviews work

0 comments
Cite this review

Pith. "Pith review of 3D-Printed Enclosure Wire-Guided Liquid Microfilm for Versatile Spectroscopy." pith.science (2026). https://pith.science/paper/EYSE6UHV

@misc{pith2026250702696,
  author       = {Pith},
  title        = {Pith review of: 3D-Printed Enclosure Wire-Guided Liquid Microfilm for Versatile Spectroscopy},
  year         = {2026},
  howpublished = {\url{https://pith.science/paper/EYSE6UHV}},
  note         = {Machine review of arXiv:2507.02696}
}
abstract

We present a 3D-printing-based design to produce wire-guided liquid microfilms that can be used for versatile spectroscopic applications. We demonstrate the ability of our instrument to provide optically useful liquid microfilms with highly tunable thicknesses over the range 25 - 180 $\mu$m, with standard temporal thickness deviation less than 1.0% on the low end of the range of flow rates, and spatially homogeneous microfilms that remain stable over the course of ten hours. We then show the device's versatility through its use in Raman, fluorescence, and nonlinear spectroscopy. Our approach is highly reproducible as a unique advantage of a 3D-printed enclosure and limited other components. The 3D-printable file for the enclosure is included in the supplementary materials. This innovation in design shows the feasibility of applying 3D-printing to physical and chemical instrumentation for faster adoption of experimental techniques.

Figures

Figures reproduced from arXiv: 2507.02696 by the authors.

Figure 1
Figure 1. FIG. 1 [PITH_FULL_IMAGE:figures/full_fig_p002_1.png] view at source ↗
Figure 2
Figure 2. FIG. 2 [PITH_FULL_IMAGE:figures/full_fig_p003_2.png] view at source ↗
Figure 3
Figure 3. FIG. 3 [PITH_FULL_IMAGE:figures/full_fig_p004_3.png] view at source ↗
Figures from the paper (1 more)
Figure 4
Figure 4. Figure 4: FIG. 4 [PITH_FULL_IMAGE:figures/full_fig_p005_4.png]

Discussion (0). Continue with ORCID to comment.

Reference graph

Works this paper leans on

39 extracted references · 36 canonical work pages

  1. [1]

    Naftaly \ and\ author R

    author author M. Naftaly \ and\ author R. E. \ Miles ,\ title title Terahertz time-domain spectroscopy of silicate glasses and the relationship to material properties , \ https://doi.org/10.1063/1.2771049 journal journal Journal of Applied Physics \ volume 102 ( year 2007 ),\ 10.1063/1.2771049 NoStop

  2. [2]

    author author M. J. \ Tauber , author R. A. \ Mathies , author X. Chen ,\ and\ author S. E. \ Bradforth ,\ title title Flowing liquid sample jet for resonance Raman and ultrafast optical spectroscopy , \ https://doi.org/10.1063/1.1614874 journal journal Review of Scientific Instruments \ volume 74 ,\ pages 4958--4960 ( year 2003 ) NoStop

  3. [3]

    Picchiotti , author V

    author author A. Picchiotti , author V. I. \ Prokhorenko ,\ and\ author R. J. \ Miller ,\ title title A closed-loop pump-driven wire-guided flow jet for ultrafast spectroscopy of liquid samples , \ https://doi.org/10.1063/1.4929860 journal journal Review of Scientific Instruments \ volume 86 ( year 2015 ),\ 10.1063/1.4929860 NoStop

  4. [4]

    Ekimova , author W

    author author M. Ekimova , author W. Quevedo , author M. Faube , author P. Wernet ,\ and\ author E. T. \ Nibbering ,\ title title A liquid flatjet system for solution phase soft-x-ray spectroscopy , \ https://doi.org/10.1063/1.4928715 journal journal Structural Dynamics \ volume 2 ( year 2015 ),\ 10.1063/1.4928715 NoStop

  5. [5]

    author author J. W. \ Riley , author B. Wang , author M. A. \ Parkes ,\ and\ author H. H. \ Fielding ,\ title title Design and characterization of a recirculating liquid-microjet photoelectron spectrometer for multiphoton ultraviolet photoelectron spectroscopy , \ https://doi.org/10.1063/1.5099040 journal journal Review of Scientific Instruments \ volume ...

  6. [6]

    Yiwen , author Q

    author author E. Yiwen , author Q. Jin , author A. Tcypkin ,\ and\ author X. C. \ Zhang ,\ title title Terahertz wave generation from liquid water films via laser-induced breakdown , \ https://doi.org/10.1063/1.5054599 journal journal Applied Physics Letters \ volume 113 ( year 2018 ),\ 10.1063/1.5054599 NoStop

  7. [7]

    Wang , author P

    author author T. Wang , author P. Klarskov ,\ and\ author P. U. \ Jepsen ,\ title title Ultrabroadband THz time-domain spectroscopy of a free-flowing water film , \ https://doi.org/10.1109/TTHZ.2014.2322757 journal journal IEEE Transactions on Terahertz Science and Technology \ volume 4 ,\ pages 425--431 ( year 2014 ) NoStop

  8. [8]

    Jin , author E

    author author Q. Jin , author E. Yiwen , author K. Williams , author J. Dai ,\ and\ author X. C. \ Zhang ,\ title title Observation of broadband terahertz wave generation from liquid water , \ https://doi.org/10.1063/1.4990824 journal journal Applied Physics Letters \ volume 111 ( year 2017 ),\ 10.1063/1.4990824 NoStop

Show all 39 references
  1. [9]

    Kuramochi , author S

    author author H. Kuramochi , author S. Takeuchi ,\ and\ author T. Tahara ,\ title title Femtosecond time-resolved impulsive stimulated Raman spectroscopy using sub-7-fs pulses: Apparatus and applications , \ https://doi.org/10.1063/1.4945259 journal journal Review of Scientifi...

  2. [10]

    Huang , author X

    author author J. Huang , author X. Lu , author F. Hu , author Y. Deng , author J. Long , author J. Tang , author L. He , author Q. Zhang , author P. Lan ,\ and\ author P. Lu ,\ title title Efficient spectral broadening and few-cycle pulse generation with multiple thin liquid f...

  3. [11]

    \ Loh , author G

    author author Z.-H. \ Loh , author G. Doumy , author C. Arnold , author L. Kjellsson , author S. H. \ Southworth , author A. A. \ Haddad , author Y. Kumagai , author M.-F. \ Tu , author P. J. \ Ho , author A. M. \ March , author R. D. \ Schaller , author M. S. B. M. \ Yusof , ...

  4. [12]

    author author P. J. \ Low , author W. Chu , author Z. Nie , author M. S. B. M. \ Yusof , author O. V. \ Prezhdo ,\ and\ author Z. H. \ Loh ,\ title title Observation of a transient intermediate in the ultrafast relaxation dynamics of the excess electron in strong-field-ionized...

  5. [13]

    Yin , author Y

    author author Z. Yin , author Y. P. \ Chang , author T. Balčiūnas , author Y. Shakya , author A. Djorović , author G. Gaulier , author G. Fazio , author R. Santra , author L. Inhester , author J. P. \ Wolf ,\ and\ author H. J. \ Wörner ,\ title title Femtosecond proton transfe...

  6. [14]

    Zhao , author Y

    author author H. Zhao , author Y. Tan , author L. Zhang , author R. Zhang , author M. Shalaby , author C. Zhang , author Y. Zhao ,\ and\ author X. C. \ Zhang ,\ title title Ultrafast hydrogen bond dynamics of liquid water revealed by terahertz-induced transient birefringence ,...

  7. [15]

    Stemer , author T

    author author D. Stemer , author T. Buttersack , author H. Haak , author S. Malerz , author H. C. \ Schewe , author F. Trinter , author K. Mudryk , author M. Pugini , author B. Credidio , author R. Seidel , author U. Hergenhahn , author G. Meijer , author S. Thürmer ,\ and\ au...

  8. [16]

    author author D. H. \ Kang , author M. Koga , author N. Haldar ,\ and\ author D. M. \ Neumark ,\ title title Dynamics of photoexcited 5-bromouracil and 5-bromo-2'-deoxyuridine studied by extreme ultraviolet time-resolved photoelectron spectroscopy in liquid flat jets , \ https...

  9. [17]

    author author C. A. \ Rivera , author S. E. \ Bradforth ,\ and\ author G. Tempea ,\ https://doi.org/10.1364/OA_License_v1#VOR title Gires-Tournois interferometer type negative dispersion mirrors for deep ultraviolet pulse compression , \ type Tech. Rep. \ ( year 2010 ) NoStop

  10. [18]

    author author T. T. \ Luu , author Z. Yin , author A. Jain , author T. Gaumnitz , author Y. Pertot , author J. Ma ,\ and\ author H. J. \ Wörner ,\ title title Extreme–ultraviolet high–harmonic generation in liquids , \ https://doi.org/10.1038/s41467-018-06040-4 journal journal...

  11. [19]

    Alexander , author J

    author author O. Alexander , author J. C. \ Barnard , author E. W. \ Larsen , author T. Avni , author S. Jarosch , author C. Ferchaud , author A. Gregory , author S. Parker , author G. Galinis , author A. Tofful , author D. Garratt , author M. R. \ Matthews ,\ and\ author J. P...

  12. [20]

    author author Y. H. \ Lo , author S. C. \ Li ,\ and\ author H. Hiramatsu ,\ title title Sampling unit for efficient signal detection and application to liquid chromatography-Raman spectroscopy , \ https://doi.org/10.1039/d0nj06054b journal journal New Journal of Chemistry \ vo...

  13. [21]

    Srivastava , author S

    author author T. Srivastava , author S. Chaudhuri , author C. C. \ Rich , author G. C. \ Schatz , author R. R. \ Frontiera ,\ and\ author P. Bruggeman ,\ title title Probing time-resolved plasma-driven solution electrochemistry in a falling liquid film plasma reactor: Identifi...

  14. [22]

    author author Y. Mizutani ,\ title title Time-resolved resonance Raman spectroscopy and application to studies on ultrafast protein dynamics , \ https://doi.org/10.1246/bcsj.20170218 journal journal Bulletin of the Chemical Society of Japan \ volume 90 ,\ pages 1344--1371 ( ye...

  15. [23]

    Perlík , author J

    author author V. Perlík , author J. Seibt , author L. J. \ Cranston , author R. J. \ Cogdell , author C. N. \ Lincoln , author J. Savolainen , author F. Šanda , author T. Mančal ,\ and\ author J. Hauer ,\ title title Vibronic coupling explains the ultrafast carotenoid-to-bacte...

  16. [24]

    author author M. L. \ Donten \ and\ author P. Hamm ,\ title title pH -jump induced -helix folding of poly-L-glutamic acid , \ https://doi.org/10.1016/j.chemphys.2012.11.023 journal journal Chemical Physics \ volume 422 ,\ pages 124--130 ( year 2013 ) NoStop

  17. [25]

    author author H. Y. \ Seferyan , author M. B. \ Nasr , author V. Senekerimyan , author R. Zadoyan , author P. Collins ,\ and\ author V. A. \ Apkarian ,\ title title Transient grating measurements of excitonic dynamics in single-walled carbon nanotubes: The dark excitonic bottl...

  18. [26]

    Sperling , author A

    author author J. Sperling , author A. Nemeth , author J. Hauer , author D. Abramavicius , author S. Mukamel , author H. F. \ Kauffmann ,\ and\ author F. Milota ,\ title title Excitons and disorder in molecular nanotubes: A 2D electronic spectroscopy study and first comparison ...

  19. [27]

    Lee , author M

    author author C. Lee , author M. N. \ Pohl , author I. A. \ Ramphal , author W. Yang , author B. Winter , author B. Abel ,\ and\ author D. M. \ Neumark ,\ title title Evaporation and molecular beam scattering from a flat liquid jet , \ https://doi.org/10.1021/acs.jpca.2c01174 ...

  20. [28]

    Yang , author M

    author author W. Yang , author M. M. \ Foreman , author S. Saric , author A. M. \ Wodtke , author K. R. \ Wilson ,\ and\ author D. M. \ Neumark ,\ title title Molecular beam scattering from flat jets of liquid dodecane and water , \ https://doi.org/10.1002/ntls.20240009 journa...

  21. [29]

    author author C. T. \ Dewberry , author R. B. \ Mackenzie , author S. Green ,\ and\ author K. R. \ Leopold ,\ title title 3D -printed slit nozzles for fourier transform microwave spectroscopy , \ https://doi.org/10.1063/1.4922852 journal journal Review of Scientific Instrument...

  22. [30]

    author author E. K. \ Grasse , author M. H. \ Torcasio ,\ and\ author A. W. \ Smith ,\ title title Teaching uv-vis spectroscopy with a 3D -printable smartphone spectrophotometer , \ https://doi.org/10.1021/acs.jchemed.5b00654 journal journal Journal of Chemical Education \ vol...

  23. [31]

    author author T. C. \ Wilkes , author A. J. S. \ McGonigle , author J. R. \ Willmott , author T. D. \ Pering ,\ and\ author J. M. \ Cook ,\ title title Low-cost 3D printed 1 nm resolution smartphone sensor-based spectrometer: instrument design and application in ultraviolet sp...

  24. [32]

    Baumgartner , author S

    author author B. Baumgartner , author S. Freitag ,\ and\ author B. Lendl ,\ title title 3D printing for low-cost and versatile attenuated total reflection infrared spectroscopy , \ https://doi.org/10.1021/acs.analchem.9b04043 journal journal Analytical Chemistry \ volume 92 ,\...

  25. [33]

    author author J. H. \ da Silva Junior , author J. V. \ de Melo ,\ and\ author P. S. \ Castro ,\ title title Lab-made 3D -printed accessories for spectroscopy and spectroelectrochemistry: a proof of concept to investigate dynamic interfacial and surface phenomena , \ https://do...

  26. [34]

    author author M. A. \ Elsawy , author K.-H. \ Kim , author J.-W. \ Park ,\ and\ author A. Deep ,\ title title Hydrolytic degradation of polylactic acid (PLA) and its composites , \ https://doi.org/https://doi.org/10.1016/j.rser.2017.05.143 journal journal Renewable and Sustain...

  27. [35]

    Ambrosi , author J

    author author A. Ambrosi , author J. G. S. \ Moo ,\ and\ author M. Pumera ,\ title title Helical 3D -printed metal electrodes as custom-shaped 3D platform for electrochemical devices , \ https://doi.org/10.1002/adfm.201503902 journal journal Advanced Functional Materials \ vol...

  28. [36]

    Wei , author H

    author author Q. Wei , author H. Li , author G. Liu , author Y. He , author Y. Wang , author Y. E. \ Tan , author D. Wang , author X. Peng , author G. Yang ,\ and\ author N. Tsubaki ,\ title title Metal 3D printing technology for functional integration of catalytic system , \ ...

  29. [37]

    Battula , author S

    author author H. Battula , author S. Bommi , author Y. Bobde , author T. Patel , author B. Ghosh ,\ and\ author S. Jayanty ,\ title title Distinct rhodamine B derivatives exhibiting dual effect of anticancer activity and fluorescence property , \ https://doi.org/10.1016/j.jpap...

  30. [38]

    author author A. S. \ Kristoffersen , author S. R. \ Erga , author B. Hamre ,\ and\ author Øyvind Frette ,\ title title Testing fluorescence lifetime standards using two-photon excitation and time-domain instrumentation: Rhodamine B , coumarin 6 and lucifer yellow , \ https://...

  31. [39]

    Alifu , author L

    author author N. Alifu , author L. Yan , author H. Zhang , author A. Zebibula , author Z. Zhu , author W. Xi , author A. W. \ Roe , author B. Xu , author W. Tian ,\ and\ author J. Qian ,\ title title Organic dye doped nanoparticles with NIR emission and biocompatibility for ul...

Pith tools

Reviewed August 6, 2026 · model on record in the stance chip above.