Tomography of photon-number resolving continuous-output detectors

classification 🪐 quant-ph
keywords approachcontinuous-outputdetectorsdetectortomographyanalysinganalysisapplies
0
0 comments X
read the original abstract

We report a comprehensive approach to analysing continuous-output photon detectors. We employ principal component analysis to maximise the information extracted, followed by a novel noise-tolerant parameterised approach to the tomography of PNRDs. We further propose a measure for rigorously quantifying a detector's photon-number-resolving capability. Our approach applies to all detectors with continuous-output signals. We illustrate our methods by applying them to experimental data obtained from a transition-edge sensor (TES) detector.