Time-resolved one-dimensional detection of x-ray scattering in pulsed magnetic fields
classification
⚛️ physics.ins-det
cond-mat.mtrl-sci
keywords
fieldspulseddetectordiffractionfieldmagneticone-dimensionalreciprocal
read the original abstract
We have developed an application of a one-dimensional micro-strip detector for capturing x-ray diffraction data in pulsed magnetic fields. This detector consists of a large array of 50 \mu m-wide Si strips with a full-frame read out at 20 kHz. Its use substantially improves data-collection efficiency and quality as compared to point detectors, because diffraction signals are recorded along an arc in reciprocal space in a time-resolved manner. By synchronizing with pulsed fields, the entire field dependence of a two-dimensional swath of reciprocal space may be determined using a small number of field pulses.