TiAu TES 32$\times$32 pixel array: uniformity, thermal crosstalk and performance at different X-ray energies

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 2102.09814 v1 pith:TZ6VLFFX submitted 2021-02-19 astro-ph.IM physics.ins-det

classification astro-ph.IMphysics.ins-det
keywords x-rayenergytimesarrayarrayscrosstalkmathrmpixels
verification ladder T0 review T1 audit T2 compute T3 formal
0 comments
abstract

Large format arrays of transition edge sensor (TES) are crucial for the next generation of X-ray space observatories. Such arrays are required to achieve an energy resolution of $\mathrm{\Delta}E<$3 eV full-width-half-maximum (FWHM) in the soft X-ray energy range. We are currently developing X-ray microcalorimeter arrays as a backup option for the X-IFU instrument on board of ATHENA space telescope, led by ESA and foreseen to be launched in 2031. In this contribution, we report on the development and the characterization of a uniform 32$\times$32 pixel array with (length$\times $width) 140$\times$30 $\mu$m$^2$ TiAu TESs, which have \textcolor{black}{a 2.3 $\mu$m} thick Au absorber for X-ray photons. The pixels have a typical normal resistance $R_\mathrm{n}$ = 121 m$\Omega$ and a critical temperature $T_\mathrm{c}\sim$ 90 mK. We performed extensive measurements on 60 pixels out of the array in order to show the uniformity of the array. We obtained an energy resolutions between 2.4 and 2.6 eV (FWHM) at 5.9 keV, measured in a single-pixel mode at AC bias frequencies ranging from 1 to 5 MHz, with a frequency domain multiplexing (FDM) readout system, which is developed at SRON/VTT. We also present the detector energy resolution at X-ray with different photon energies generated by a modulated external X-ray source from 1.45 keV up to 8.9 keV. Multiplexing readout across several pixels has also been performed to evaluate the impact of the thermal crosstalk to the instrument's energy resolution budget requirement. This value results in a derived requirement, for the first neighbour, that is less than 1$\times$10$^{-3}$ when considering the ratio between the amplitude of the crosstalk signal to an X-ray pulse (for example at 5.9 keV)

Discussion (0). Sign in to comment.

Pith tools