Intrinsic mirror noise in Fabry-Perot based polarimeters: the case for the measurement of vacuum magnetic birefringence

classification ⚛️ physics.ins-det hep-exphysics.optics
keywords birefringencecavitymagneticnoiseopticalvacuumeffectpath
0
0 comments X
read the original abstract

Although experimental efforts have been active for about 30 years now, a direct laboratory observation of vacuum magnetic birefringence, an effect due to vacuum fluctuations, still needs confirmation. Indeed, the predicted birefringence of vacuum is $\Delta n = 4.0\times 10^{-24}$ @ 1~T. One of the key ingredients when designing a polarimeter capable of detecting such a small birefringence is a long optical path length within the magnetic field and a time dependent effect. To lengthen the optical path within the magnetic field a Fabry-Perot optical cavity is generally used with a finesse ranging from ${\cal F} \approx 10^4$ to ${\cal F} \approx7\times 10^5$. Interestingly, there is a difficulty in reaching the predicted shot noise limit of such polarimeters. We have measured the ellipticity and rotation noises along with a Cotton-Mouton and a Faraday effect as a function of the finesse of the cavity of the PVLAS polarimeter. The observations are consistent with the idea that the cavity mirrors generate a birefringence-dominated noise whose ellipticity is amplified by the cavity itself. The optical path difference sensitivity at $10\;$Hz is $S_{\Delta{\cal D}}=6\times 10^{-19}\;$m$/\sqrt{\rm Hz}$, a value which we believe is consistent with an intrinsic thermal noise in the mirror coatings.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.