pith. sign in

arxiv: physics/0703006 · v1 · submitted 2007-03-01 · ⚛️ physics.ins-det · physics.comp-ph

Simulation of Pickup Signal in a Resistive Plate Chamber

classification ⚛️ physics.ins-det physics.comp-ph
keywords pickupsignalbeenelectronfieldaccurateboundarycalculated
0
0 comments X
read the original abstract

The signal induced by an electron on a pickup strip has been calculated in a RPC used in INO calorimeter following Ramo's theorem. An accurate estimation of weighting field has been obtained using a nearly exact Boundary Element Method (neBEM) solver while the electron velocity from the real field values. The calculation has shown a growth and subsequent fall of pickup signal with duration about 90ps.

This paper has not been read by Pith yet.