REVIEW 4 major objections 5 minor 1 cited by
A new generation of silicon cryogenic detectors reaches sub-eV energy resolution and reveals a consistent ~0.8 eV energy loss for every absorbed photon, bringing phonon-based detectors close to the theoretical limit for reading out single c
Reviewed by Pith at T0; open to challenge. T0 means a machine referee read the full paper against a public rubric. the ladder, T0–T4 →
2026-08-03 08:35 UTC pith:ER3YNCP7
load-bearing objection Sub-eV resolution is real and important; the efficiency and E_loss numbers are plausible but rest on a model that deserves more scrutiny, and the abstract-body numeric mismatch must be fixed. the 4 major comments →
Performance of a SuperCDMS HVeV Detector with Sub-eV Energy Resolution and Single Charge-sensitivity
The pith
A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.
Core claim
The paper's central claim is that lowering the critical temperature of the tungsten transition-edge sensors from ~65 mK to ~40 mK improves a silicon HVeV detector's baseline resolution to sub-eV levels — 612 meV for the CUTE device and 589 meV claimed in the abstract — without sacrificing single-charge sensitivity. With this resolution, the authors resolve quantized electron-hole-pair peaks from individual optical photons and, by fitting a detector-response model simultaneously to four datasets, extract a per-photon energy loss of about 0.8 eV in the body text (0.92 eV per the abstract) that is consistent between the CUTE and SLAC facilities. They further claim that the measured noise is des
What carries the argument
The load-bearing pieces are the QET (quasiparticle-trap-assisted electrothermal-feedback TES) detector with lower-Tc tungsten films, and a detector-response model (Eq. 8) that relates the reconstructed energy of each event to the number of electron-hole pairs produced and the number of photons absorbed, minus an energy-loss term E_loss. A Markov-chain Monte Carlo fit simultaneously constrains E_loss, charge trapping, surface trapping, impact ionization, Poisson photon statistics, and calibration factors. Noise is modeled as the quadrature sum of SQUID readout noise, thermal-fluctuation power noise, and Johnson noise, with only the thermal conductance G treated as a free parameter.
Load-bearing premise
The absolute energy scale and the '0.8 eV loss' value rest on the detector-response model of Ref [21] (Eq. 8), which simultaneously fits many charge-mobility parameters; if those parameters are degenerate, the energy-loss claim and the derived efficiency would change.
What would settle it
Take the same device, illuminate it with photons spanning a range of energies (1.5–4.4 eV) and bias voltages, and test whether the fitted E_loss stays constant at ~0.8 eV or scales with photon energy or bias; a clear dependence would falsify the claim of a single consistent per-photon loss.
If this is right
- If the sub-eV resolution and single-charge sensitivity hold, these detectors can calibrate eV-scale energy deposits in silicon with per-photon accuracy, sharpening low-mass dark matter and coherent neutrino scattering searches.
- The consistent ~0.8 eV per-photon loss implies that calibration of HVeV-style detectors must correct for a bias-dependent energy offset, else reconstructed spectra are systematically shifted.
- The measured 45% phonon collection efficiency, close to the ~60% quasiparticle-generation limit, suggests this design is near-optimal and further gains in efficiency will require addressing the quasiparticle conversion step itself.
- The one-free-parameter noise model, validated across two facilities, provides an unbiased way to measure thermal conductance and to bound parasitic power in TES-based phonon detectors.
- The absence of excess correlated noise in these devices, once scaled for volume and efficiency, tentatively rules out a common intrinsic correlated-phonon source in silicon at the reported sensitivity.
Where Pith is reading between the lines
- The energy-loss value of ~0.8 eV per photon, if confirmed across a wider range of photon energies and bias voltages, would provide a direct handle on surface band-bending and could inform models of charge collection in cryogenic semiconductor detectors.
- The claim that the detector is at the quasiparticle-generation limit could be tested by fabricating devices with different absorber or trap geometries and checking whether the 45% efficiency moves with the 60% limit or with the phonon collection from the substrate.
- One could extend the technique to other absorber materials (e.g., germanium or diamond) to see whether the per-photon loss scales with band gap or with surface state density, separating intrinsic material effects from surface preparation effects.
- The paper's correlated-noise analysis could be sharpened by operating the same device in the same cryostat with and without a pulse-tube cooler, isolating mechanical coupling as the dominant noise source.
Editorial analysis
A structured set of objections, weighed in public.
Referee Report
Summary. The paper reports the characterization of a new generation of SuperCDMS HVeV silicon detectors with lower- Tc TESs, claiming sub-eV baseline energy resolution (612 meV in the body; 589 meV in the abstract), single-photon sensitivity with a consistent energy loss of ~0.81 eV per absorbed photon across two facilities (0.92 eV in the abstract), a noise description by a standard TES model with one free parameter (G_inferred), and a nominal phonon collection efficiency of 45% at CUTE. The analysis combines TES admittance measurements, noise PSD fitting, LED/background calibrations, and an MCMC detector-response model (Eq. 8), with appendices on efficiency, triggering, cross-talk, and a decorrelation algorithm.
Significance. If the central claims hold, the 0.6 eV baseline resolution and ~45% phonon collection efficiency represent a meaningful advance for single-charge/phonon detectors, with direct implications for low-mass dark matter searches and eV-scale calibration techniques. The cross-facility comparison of detector performance and the validation of the TES noise model with a single free parameter are valuable contributions. The paper also provides a careful treatment of correlated noise and includes a Monte Carlo validation of the decorrelation procedure, which strengthens the noise-related claims. However, the internal numeric inconsistencies and the lack of identifiability diagnostics for the MCMC model currently prevent the results from being taken at face value.
major comments (4)
- [Abstract vs. §1; Table 2] The abstract quotes a baseline resolution of 589±5 meV and E_loss = 0.92±0.02 eV, while the body reports σ_m = 612±4 meV and E_loss = 0.81±0.03 eV (Table 2). §1 further states '612 meV ± 0.04 meV'. In addition, the abstract describes the efficiency as 'place upper bounds on the nominal phonon collection efficiency of 45%', whereas Appendix A and Table 4 report a measurement of 45%±3%±6%. These are the headline numbers of the paper and must be made consistent before the claims can be evaluated.
- [§4.1–4.2; Table 2; Appendix C] The 'consistent energy loss' claim rests on the MCMC fit of Eq. (8), which simultaneously determines E_loss, P_CT, P_II, P_ST, c1, c2, λ, and ACT. No identifiability diagnostics (posterior corner plots, profile likelihoods, or prior sensitivity) are shown. The CUTE posterior is narrow (0.81±0.03 eV), but the SLAC posterior is 0.8+0.2/−0.1 eV, so the two-facility consistency is weak. Furthermore, ACT is assumed energy-independent (Appendix C); if ACT varies with LED drive amplitude, E_loss will absorb the residual. Please provide degeneracy checks and a sensitivity test for the ACT assumption.
- [§3.3; Tables 1 and 2] The predicted resolutions in Table 1 (σ_didv ≈ 0.24 eV at SLAC, 0.34 eV at CUTE) are factors of ~4 and ~2 smaller than the measured resolutions in Table 2 (σ_m ≈ 0.90 eV and 0.612 eV). The paper does not address this discrepancy. If Eq. (7) is an approximation, that should be stated and the direct evaluation of Eq. (6) provided. Without this, the claim that the standard TES noise model describes the detectors is not fully supported by the resolution comparison.
- [Appendix A; Table 4] The quoted systematic uncertainty of 6% on the CUTE efficiency is not documented. Figure 8 shows the efficiency as a function of integration length, but the procedure for deriving the systematic from the plateau variation is not described. Since the central efficiency claim (45%) is load-bearing, the authors should specify which sources contribute to the 6% systematic and how they were estimated.
minor comments (5)
- [§2] There are several typos and formatting issues: 'Moku:Proto' should likely be 'Moku:Pro'; 'Zö 𝑒Smith' contains a stray character; 'e.g.,' often lacks spaces. These should be corrected.
- [Table 1 caption] The sentence 'The theoretical energy resolutions quoted below is estimated' is grammatically incorrect; 'is' should be 'are'.
- [Eq. (7)] The symbol ζ_TES is defined in the text as 'fractional volume of the TES film separate from the Al/W trapping regions', but the definition is not fully clear. Please provide a precise definition or a reference.
- [§5.2] When comparing to TESSERACT, the text says 'assuming the same tungsten Tc as that reported in Ref [8]' but does not state the numerical value. Please specify it for clarity.
- [Appendix A] The efficiency for SLAC is based on LED single e−/h+ peaks and therefore depends on E_loss via Eq. (8), while the CUTE efficiency uses background events and does not. This important difference should be stated explicitly when reporting the efficiencies in Table 4.
Circularity Check
No significant circularity: the headline measurements are direct, and cited prior models are independent published work rather than imported conclusions.
full rationale
The paper's central quantities are measured or fitted with new data rather than derived from their own definitions. Baseline resolution (612 ± 4 meV at CUTE; Table 2, Sec. 4.1) comes from optimal-filter widths of LED/background event histograms; the TES noise model is benchmarked by fitting only G_inferred to the measured NEP spectra and comparing to measured resolution, which is a consistency check, not a fitted-input prediction (Sec. 3.2–3.3, Table 1). The headline 45% phonon-collection efficiency uses background single-e/h events at CUTE for which E_expected = V in Eq. 8, independent of the fitted E_loss (Appendix A, Eq. 11). The E_loss parameter (~0.8 eV) is reported as a fit result of the MCMC response model shared across four CUTE datasets, with SLAC as an independent (if less precise) dataset; this is parameter estimation, not a prediction forced by construction. Cross-talk ACT is estimated by extrapolating the 0th-peak position to λ_m = 0 with a stated prior, and is varied in the fit (Appendix C). Self-citations to Refs. [11] and [21] provide detector design and a previously published response model; the current data and external TESSERACT comparison provide independent support, and no uniqueness theorem or ansatz is imported as proof. The abstract/full-text numeric mismatches (589 vs 612 meV; 0.92 vs 0.81 eV) are manuscript consistency issues that must be fixed, but they do not constitute circular reasoning under the required standard.
Axiom & Free-Parameter Ledger
free parameters (7)
- G_inferred (thermal conductance per channel) =
SLAC ~83.4 pW/K; CUTE ~80.7 pW/K
- E_loss (lost photon energy per absorbed photon) =
0.80+0.2-0.1 eV (SLAC); 0.81±0.03 eV (CUTE)
- P_CT, P_II, P_ST (charge trapping, impact ionization, surface trapping probabilities) =
SLAC 10/7/29%; CUTE 27.1/1.2/18.5%
- c1, c2 (calibration factors) =
SLAC 290/27; CUTE 340.6/55.2
- λ (mean photon number per LED pulse) =
varied per dataset
- ACT (cross-talk amplitude shift) =
0.25±0.02 nA
- Integration cutoff for efficiency =
>5 fall times
axioms (6)
- domain assumption Standard TES electro-thermal feedback and thermal fluctuation noise model (four noise components) describes the detector.
- domain assumption Thermal conduction power-law exponent n=5 for the TES-bath coupling.
- domain assumption Crystal is in thermal equilibrium with the mixing chamber stage (T_b = T_mixing chamber).
- ad hoc to paper Cross-talk amplitude ACT is independent of deposited energy.
- domain assumption Detector-response model of Wilson et al. [21] correctly describes charge transport, trapping, and impact ionization.
- domain assumption Theoretical quasiparticle-generation efficiency limit of ~60% (Guruswamy et al.) is accurate.
read the original abstract
We present a detailed characterization of a new generation of athermal-phonon single-charge sensitive Si HVeV detectors, the best of which achieved 589~meV~$\pm$~5~meV baseline resolution. Our sub-eV energy resolution enables precise measurements of single-photon events and reveal consistent energy losses of 0.92~eV~$\pm$~0.02~eV per charge excitation across two facilities. We demonstrate that the noise for these detectors is well described using a standard Transition Edge Sensor noise model. We also measure a nominal phonon collection efficiency of 45\%~$\pm$~3\%~(stat.)~$\pm$~6\%~(syst.) in the best performing device, establishing these detectors as the most efficient athermal phonon detectors to date, limited only by intrinsic limitations of quasiparticle generation.
Figures
Forward citations
Cited by 1 Pith paper
-
Photon Calibration Techniques for High Resolution Cryogenic Detectors
Realistic detector models violate Poisson-calibration assumptions and introduce bias whose size depends on specific performance parameters.
Reference graph
Works this paper leans on
-
[1]
Y. Kahn, T. Lin, Searches for light dark matter using condensed mattersystems,Rep.Prog.Phys.85(6)(2022)066901. doi:10.1088/ 1361-6633/ac5f63
2022
-
[2]
K. M. Zurek, Dark Matter Candidates of a Very Low Mass, Annu. Rev. Nucl. Part. Sci. 74 (2024) 287–319.doi:https://doi.org/10. 1146/annurev-nucl-101918-023542. : Preprint submitted to Elsevier Page 11 of 16
2024
-
[3]
H.Murayama,S.Asai,K.Heeger,A.Ballarino,T.Bose,K.Cranmer, F.-Y. Cyr-Racine, S. Demers, C. Geddes, Y. Gershtein, et al., Explor- ing the Quantum Universe: Pathways to Innovation and Discovery in Particle Physics, 2023.doi:10.2172/2368847
doi:10.2172/2368847 2023
- [4]
-
[5]
C.Goupy,H.Abele,G.Angloher,A.Bento,L.Canonica,F.Cappella, L. Cardani, N. Casali, R. Cerulli, I. Colantoni, et al., Exploring coherent elastic neutrino-nucleus scattering of reactor neutrinos with the NUCLEUS experiment, SciPost Phys. Proc. 12 (2023) 053.doi: 10.21468/SciPostPhysProc.12.053
-
[6]
L. Baudis, A. Bismark, N. Brugger, C. Capelli, I. Charaev, J. C. García, G. D. Hadas, Y. Hochberg, J. K. Hohmann, A. Kavner, et al., First Sub-MeV Dark Matter Search with the QROCODILE Exper- iment Using Superconducting Nanowire Single-Photon Detectors, Phys. Rev. Lett. 135 (2025) 081002.doi:10.1103/4hb6-f6jl
-
[7]
J.Billard,J.Gascon,S.Marnieros,S.Scorza,TransitionEdgeSensors with Sub-eV Resolution And Cryogenic Targets (TESSERACT) at the underground laboratory of Modane (LSM), Nucl. Phys. B 1003 (2024) 116465, special Issue of Nobel Symposium 182 on Dark Matter. doi:https://doi.org/10.1016/j.nuclphysb.2024.116465
arXiv 2024
-
[8]
R.Anthony-Petersen,C.L.Chang,Y.-Y.Chang,L.Chaplinsky,C.W. Fink, M. Garcia-Sciveres, W. Guo, S. A. Hertel, X. Li, J. Lin, et al., Low energy backgrounds and excess noise in a two-channel low- threshold calorimeter, Appl. Phys. Lett. 126 (10) (Mar. 2025).doi: 10.1063/5.0247343
-
[9]
R. K. Romani, P. L. Brink, B. Cabrera, M. Cherry, T. Howarth, N. Kurinsky, R. A. Moffatt, R. Partridge, F. Ponce, M. Pyle, et al., Thermal detection of single e-h pairs in a biased silicon crystal detector, Appl. Phys. Lett. 112 (4) (2018) 043501.doi:10.1063/1. 5010699
doi:10.1063/1 2018
-
[10]
Z. Hong, R. Ren, N. Kurinsky, E. Figueroa-Feliciano, L. Wills, S.Ganjam,R.Mahapatra,N.Mirabolfathi,B.Nebolsky,H.D.Pinck- ney, et al., Single electron–hole pair sensitive silicon detector with surface event discrimination, Nucl. Instrum. Methods A 963 (2020) 163757. doi:10.1016/j.nima.2020.163757
arXiv 2020
-
[11]
R. Ren, C. Bathurst, Y. Y. Chang, R. Chen, C. W. Fink, Z. Hong, N. A. Kurinsky, N. Mast, N. Mishra, V. Novati, et al., Design and characterization of a phonon-mediated cryogenic particle detector with an eV-scale threshold and 100 keV-scale dynamic range, Phys. Rev. D 104 (3) (2021) 032010.doi:10.1103/PhysRevD.104.032010
-
[12]
R. Agnese, T. Aralis, T. Aramaki, I. Arnquist, E. Azadbakht, W. Baker, S. Banik, D. Barker, D. Bauer, T. Binder, et al., First Dark Matter Constraints from a SuperCDMS Single-Charge Sensitive De- tector,Phys.Rev.Lett.121(5)(Aug.2018). doi:10.1103/physrevlett. 121.051301
-
[13]
D.Amaral,T.Aralis,T.Aramaki,I.Arnquist,E.Azadbakht,S.Banik, D. Barker, C. Bathurst, D. Bauer, L. Bezerra, et al., Constraints on low-mass, relic dark matter candidates from a surface-operated SuperCDMS single-charge sensitive detector, Phys. Rev. D 102 (9) (Nov. 2020).doi:10.1103/physrevd.102.091101
-
[14]
M. F. Albakry, I. Alkhatib, D. Alonso-González, D. W. P. Amaral, J.Anczarski,T.Aralis,T.Aramaki,I.J.Arnquist,I.AtaeeLangroudy, E. Azadbakht, et al., Light dark matter constraints from SuperCDMS HVeV detectors operated underground with an anticoincidence event selection, Phys. Rev. D 111 (2025) 012006.doi:10.1103/PhysRevD. 111.012006
doi:10.1103/physrevd 2025
-
[15]
M. F. Albakry, I. Alkhatib, D. Alonso-González, J. Anczarski, T. Ar- alis, T. Aramaki, I. A. Langroudy, C. Bathurst, R. Bhattacharyya, A. J. Biffl, et al., Search for low-mass electron-recoil dark matter using a single-charge sensitive SuperCDMS-HVeV Detector (2025). arXiv:2509.03608
arXiv 2025
-
[16]
Arnquist, I
M.Albakry,I.Alkhatib,D.Alonso,D.Amaral,T.Aralis,T.Aramaki, I. Arnquist, I. Ataee Langroudy, E. Azadbakht, S. Banik, et al., First Measurement of the Nuclear-Recoil Ionization Yield in Silicon at 100 eV, Phys. Rev. Lett. 131 (9) (Aug. 2023). doi:10.1103/ physrevlett.131.091801
2023
-
[17]
doi:10.1140/epjc/s10052-025-13844-4
G.Bratrud,C.Chang,R.e.a.Chen,Firstdemonstrationofatesbased cryogenic limoo detector for neutrinoless double beta decay search, Eur.Phys.J.C85(2025)118. doi:10.1140/epjc/s10052-025-13844-4
-
[18]
M. F. Albakry, I. Alkhatib, D. Alonso-Gonźalez, D. W. P. Amaral, J. Anczarski, T. Aralis, T. Aramaki, I. A. Langroudy, C. Bathurst, R. Bhattacharyya, et al., Low-Energy Calibration of SuperCDMS HVeVCryogenicSiliconCalorimetersUsingComptonSteps(2025). arXiv:2508.02402
arXiv 2025
-
[19]
Ataee Langroudy, E
M.Albakry,I.Alkhatib,D.Amaral,T.Aralis,T.Aramaki,I.Arnquist, I. Ataee Langroudy, E. Azadbakht, S. Banik, C. Bathurst, et al., Investigating the sources of low-energy events in a SuperCDMS- HVeV detector, Phys. Rev. D 105 (11) (Jun. 2022).doi:10.1103/ physrevd.105.112006
2022
-
[21]
M. Wilson, A. Zaytsev, B. von Krosigk, I. Alkhatib, M. Buchanan, R. Chen, M. Diamond, E. Figueroa-Feliciano, S. Harms, Z. Hong, et al., Improved modeling of detector response effects in phonon- based crystal detectors used for dark matter searches, Phys. Rev. D 109 (11) (Jun. 2024).doi:10.1103/physrevd.109.112018
-
[22]
M. Kelsey, R. Agnese, Y. Alam, I. A. Langroudy, E. Azadbakht, D. Brandt, R. Bunker, B. Cabrera, Y.-Y. Chang, et al., G4CMP: Condensed matter physics simulation using the Geant4 toolkit, Nucl. Instrum. Methods A 1055 (2023) 168473.doi:10.1016/j.nima.2023. 168473
-
[23]
Agostinelli, J
S. Agostinelli, J. Allison, K. Amako, J. Apostolakis, H. Araujo, P. Arce, M. Asai, D. Axen, S. Banerjee, G. Barrand, et al., Geant4 - A Simulation Toolkit, Nucl. Instrum. Meth. A 506 (2003) 250 – 303
2003
-
[24]
Allison, K
J. Allison, K. Amako, J. Apostolakis, H. Araujo, P. Arce Dubois, M. Asai, G. Barrand, R. Capra, S. Chauvie, R. Chytracek, et al., Geant4 Developments and Applications, IEEE Trans. Nucl. Sci. 53 (2006) 270 – 278
2006
-
[25]
J. Allison, K. Amako, J. Apostolakis, P. Arce, M. Asai, T. Aso, E. Bagli, A. Bagulya, S. Banerjee, G. Barrand, et al., Recent devel- opments in Geant4, Nucl. Instrum. Methods A 835 (2016) 186–225. doi:10.1016/j.nima.2016.06.125
-
[26]
Camus, A
P. Camus, A. Cazes, A. Dastgheibi-Fard, K. Dering, G. Gerbier, W.Rau,S.Scorza,X.Zhang,CUTE:ALowBackgroundFacilityfor TestingCryogenicDarkMatterDetectors,J.LowTemp.Phys.193(5-
-
[27]
(2018) 813–818.doi:10.1007/s10909-018-2014-0
-
[28]
P.Camus,J.Corbett,S.Crawford,K.Dering,E.Fascione,G.Gerbier, R.Germond,M.Ghaith,J.Hall,Z.Hong,etal.,CUTE:ACryogenic Underground TEst facility at SNOLAB, Front. Phys. 11 (Jan. 2024). doi:10.3389/fphy.2023.1319879
arXiv 2024
-
[29]
P. Adari, A. A. Aguilar-Arevalo, D. Amidei, G. Angloher, E. Ar- mengaud, C. Augier, L. Balogh, S. Banik, D. Baxter, C. Beaufort, etal., EXCESSworkshop:Descriptions ofrisinglow-energy spectra, SciPost Phys. Proc. 9 (2022) 001.doi:10.21468/scipostphysproc.9. 001
-
[30]
K. D. Irwin, S. W. Nam, B. Cabrera, B. Chugg, B. A. Young, A quasiparticle-trap-assisted transition-edge sensor for phonon- mediated particle detection, Rev. Sci. Instrum. 66 (11) (1995) 5322–
1995
-
[31]
M. E. Huber, P. A. Neil, R. G. Benson, D. A. Burns, A. Corey, C. S. Flynn, Y. Kitaygorodskaya, O. Massihzadeh, J. M. Martinis, G. Hilton, DC SQUID series array amplifiers with 120 MHz band- width(corrected),IEEETrans.Appl.Supercond.11(2)(2001)4048–
2001
-
[32]
Hansen, F
S. Hansen, F. DeJongh, J. Hall, B. A. Hines, M. E. Huber, T. Kiper, V. Mandic, W. Rau, T. Saab, D. Seitz, K. Sundqvist, The cryogenic darkmattersearchteststandwarmelectronicscard,in:IEEENuclear ScienceSymposium&MedicalImagingConference,2010,pp.1392–
2010
-
[33]
C. W. Fink, S. L. Watkins, QETpy (Jan. 2022).doi:10.5281/zenodo. 5903353
doi:10.5281/zenodo 2022
-
[34]
K. Irwin, G. Hilton, Cryogenic Particle Detection, Springer Berlin Heidelberg, Berlin, Heidelberg, 2005, Ch. Transition-Edge Sensors, pp. 63–150. doi:10.1007/10933596_3
-
[35]
C. W. Fink, S. L. Watkins, T. Aramaki, P. L. Brink, S. Ganjam, B. A. Hines, M. E. Huber, N. A. Kurinsky, R. Mahapatra, N. Mirabolfathi, et al., Characterizing TES power noise for future single optical- phonon and infrared-photon detectors, AIP Adv. 10 (8) (Aug. 2020). doi:10.1063/5.0011130
-
[36]
K. D. Irwin, G. C. Hilton, Transition-edge sensors, in: Cryogenic particle detection, Springer, 2005, pp. 63–150
2005
-
[37]
K. Ramanathan, N. Kurinsky, Ionization yield in silicon for eV-scale electron-recoil processes, Phys. Rev. D 102 (6) (2020) 063026.doi: 10.1103/PhysRevD.102.063026
-
[38]
T.Guruswamy,D.J.Goldie,S.Withington,Quasiparticlegeneration efficiency in superconducting thin films, Supercond. Sci. Technol. 27 (5) (2014) 055012.doi:10.1088/0953-2048/27/5/055012
-
[39]
F.Ponce,W.Page,P.Brink,B.Cabrera,M.Cherry,C.Fink,N.Kurin- sky,R.Partridge,M.Pyle,B.Sadoulet,etal.,ModelingofImpactIon- izationandChargeTrappinginSuperCDMSHVeVDetectors,J.Low Temp. Phys. 199 (2020) 598–605.doi:10.1007/s10909-020-02349-x
-
[40]
I. J. Maasilta, Complex impedance, responsivity and noise of transition-edgesensors:Analyticalsolutionsfortwo-andthree-block thermal models, AIP Advances 2 (2012) 042110
2012
-
[41]
C. L. Chang, Y.-Y. Chang, M. Garcia-Sciveres, W. Guo, S. A. Her- tel, X. Li, J. Lin, M. Lisovenko, R. Mahapatra, W. Matava, et al., Spontaneousgenerationofathermalphononburstswithinbulksilicon causingexcessnoise,lowenergybackgroundeventsandquasiparticle poisoning in superconducting sensors (2025).arXiv:2505.16092
arXiv 2025
-
[42]
Fink, A Gram-Scale low-Tc Low-Surface-Coverage Athermal- Phonon Sensitive Dark Matter Detector, Ph.D
C. Fink, A Gram-Scale low-Tc Low-Surface-Coverage Athermal- Phonon Sensitive Dark Matter Detector, Ph.D. thesis, UC Berkeley (2022). Appendix A. Phonon Collection Efficiency The phonon collection efficiency refers to the fraction of phonon energy that is absorbed by the TES, and can be thoughtofastheQETtotalquantumefficiency,asitincludes both phonon and i...
2022
-
[44]
F. Ponce, C. Stanford, S. Yellin, W. Page, C. Fink, M. Pyle, B. Sadoulet, B. Serfass, S. L. Watkins, P. L. Brink, et al., Measuring the impact ionization and charge trapping probabilities in Super- CDMS HVeV phonon sensing detectors, Phys. Rev. D 101 (2020) 031101. doi:10.1103/PhysRevD.101.031101
-
[47]
Phonon collection efficiency
-
[48]
Phonon to quasiparticle conversion efficiency
-
[49]
QP collection efficiency
-
[50]
Efficiency Statistical Systematic SLAC 61% ± 16% ± 7% CUTE 45% ± 3% ± 6% bias voltage
Trapped QP to TES thermal energy conversion effi- ciency Theabsorbedenergyiscomputedastheintegralofthepower pulse over the time it takes to return to baseline𝑇 as 𝐸abs= ∫ 𝑇 0 Δ𝑃𝑑𝑡 (9) = ∫ 𝑇 0 [(𝑉𝑏−2𝐼0𝑅𝑙 )Δ𝐼(𝑡)−Δ 𝐼(𝑡)2𝑅𝑙 ]𝑑𝑡, (10) whereΔ𝐼 refers to the change in TES current from a small change in temperature,𝑅𝑙 = 𝑅𝑝 + 𝑅shunt, and𝑉𝑏 is the Table 4 Phonon co...
-
[1395]
doi:10.1109/NSSMIC.2010.5874000
arXiv 2010
-
[4053]
: Preprint submitted to Elsevier Page 12 of 16
doi:10.1109/77.919577. : Preprint submitted to Elsevier Page 12 of 16
-
[5326]
doi:10.1063/1.1146105
This paper was first reviewed by deepseek-v4-flash on August 3, 2026.
discussion (0)
Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.