pith. machine review for the scientific record. sign in

arxiv: 1601.01651 · v1 · submitted 2016-01-07 · ⚛️ physics.ins-det · astro-ph.IM· physics.data-an

Recognition: unknown

Processing of X-ray Microcalorimeter Data with Pulse Shape Variation using Principal Component Analysis

Authors on Pith no claims yet
classification ⚛️ physics.ins-det astro-ph.IMphysics.data-an
keywords x-rayshapevariationanalysiscomponentmethodprincipalpulse
0
0 comments X
read the original abstract

We present a method using principal component analysis (PCA) to process x-ray pulses with severe shape variation where traditional optimal filter methods fail. We demonstrate that PCA is able to noise-filter and extract energy information from x-ray pulses despite their different shapes. We apply this method to a dataset from an x-ray thermal kinetic inductance detector which has severe pulse shape variation arising from position-dependent absorption.

This paper has not been read by Pith yet.

discussion (0)