pith. sign in

arxiv: 1609.09778 · v1 · pith:KRD6VSMMnew · submitted 2016-09-30 · ❄️ cond-mat.mtrl-sci · cond-mat.mes-hall

Bandgap measurement of high refractive index materials by off-axis EELS

classification ❄️ cond-mat.mtrl-sci cond-mat.mes-hall
keywords lossrangeeelsmaterialsoff-axisaccelerationbandgapdetermined
0
0 comments X
read the original abstract

In the present work, Cs aberration corrected and monochromated scanning transmission electron microscopy electron energy loss spectroscopy STEM-EELS has been used to explore experimental set-ups that allows bandgaps of high refractive index materials to be determined. Semi-convergence and -collection angles in the micro-radian range were combined with off-axis or dark field EELS to avoid relativistic losses and guided light modes in the low loss range to contribute to the acquired EEL spectra. Off-axis EELS further suppressed the zero loss peak and the tail of the zero loss peak. The bandgap of several GaAs-based materials were successfully determined by direct inspection and without any background subtraction of the EEL spectra. The presented set-up does not require that the acceleration voltage is set to below the Cerenkov limit and can be applied over the entire acceleration voltage range of modern TEMs and for a wide range of specimen thicknesses.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.