INTENTAS -- An entanglement-enhanced atomic sensor for microgravity

Pith reviewed 2026-05-23 21:26 UTC · model grok-4.3

classification ⚛️ physics.ins-det cond-mat.quant-gasphysics.atom-phphysics.space-phquant-ph
keywords atomic sensorentangled Bose-Einstein condensatesmicrogravityquantum sensingEinstein-Elevatorall-optical BEC creationSWaP constraintslow-noise environment
0
0 comments X

The pith

INTENTAS designs an atomic sensor that uses entangled Bose-Einstein condensates in microgravity to combine entanglement-enhanced sensitivity with longer interrogation times.

A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.

The paper presents the INTENTAS project, which develops an atomic sensor based on entangled Bose-Einstein condensates for a microgravity environment. The design targets measurements that gain from both the sensitivity boost of entanglement and the extended free-fall times available away from Earth gravity. It solves size-weight-power limits and noise requirements specific to the Einstein-Elevator platform while using an all-optical method to form the condensates. This approach keeps the system flexible across configurations and supports fast experiment cycles. The authors state that a working demonstration on the elevator is the required step before moving the technology to space for high-precision quantum sensing.

Core claim

The central claim is that a compact apparatus can be engineered to create and detect entangled BECs inside the low-noise, SWaP-constrained Einstein-Elevator environment through all-optical methods, thereby enabling entanglement-enhanced atomic sensing with extended interrogation times as a direct path to space-based operation.

What carries the argument

All-optical creation of Bose-Einstein condensates, which supplies the flexibility needed for varied configurations and rapid turnaround while fitting the platform constraints.

If this is right

  • A successful Einstein-Elevator run will establish the technology readiness for space deployment.
  • Microgravity will allow interrogation times long enough to exploit the full sensitivity gain from entanglement.
  • The same all-optical architecture can support multiple sensor configurations without hardware redesign.
  • Rapid turnaround times will enable repeated tests under varying conditions on the elevator platform.

Where Pith is reading between the lines

These are editorial extensions of the paper, not claims the author makes directly.

  • The same low-SWaP entangled-BEC source could be adapted for ground-based portable sensors if vibration isolation improves.
  • Success would motivate similar entanglement upgrades in other microgravity atom interferometers such as gravimeters or clocks.
  • The flexible all-optical design might shorten development cycles for future space missions that require custom BEC geometries.

Load-bearing premise

The apparatus can deliver the low-noise conditions and successful all-optical production of entangled BECs inside the size, weight, and power envelope of the Einstein-Elevator.

What would settle it

Direct measurement during an Einstein-Elevator run showing either noise levels high enough to destroy observable entanglement or failure to produce and detect the condensates within the allotted power and time budget.

Figures

Figures reproduced from arXiv: 2409.01051 by A. Fieguth, A. Heidt, A. Wicht, C. Klempt, C. Lotz, D. Derr, E. Giese, E. M. Rasel, I. Br\"ockel, J. Hamann, J. Kruse, J. Pahl, J. S. Haase, K. M\"uller, L. W\"orner, M. Franzke, M. G\"artner, M. Krutzik, M. Schiemangk, O. Anton, S. Kanthak, S. Kubitza, S. Schwertfeger, W. P. Schleich.

Figure 1
Figure 1. Figure 1: FIG. 1: Sketch of the optical ports of the vacuum chamber (left: [PITH_FULL_IMAGE:figures/full_fig_p004_1.png] view at source ↗
Figure 2
Figure 2. Figure 2: FIG. 2: The INTENTAS setup with its three chambers within the magnetic shielding on [PITH_FULL_IMAGE:figures/full_fig_p006_2.png] view at source ↗
Figure 3
Figure 3. Figure 3: FIG. 3: The CAD representation of the magnetic shield with a sectional view from a [PITH_FULL_IMAGE:figures/full_fig_p008_3.png] view at source ↗
Figure 4
Figure 4. Figure 4: FIG. 4: (a) Measurement of the magnetic field during flight. (b) Simulated magnetic field [PITH_FULL_IMAGE:figures/full_fig_p009_4.png] view at source ↗
Figure 5
Figure 5. Figure 5: FIG. 5: Schematic of the laser system. The system consists out of four subsystems: a) [PITH_FULL_IMAGE:figures/full_fig_p010_5.png] view at source ↗
Figure 6
Figure 6. Figure 6: FIG. 6: Photograph of a science laser module. [PITH_FULL_IMAGE:figures/full_fig_p011_6.png] view at source ↗
Figure 7
Figure 7. Figure 7: FIG. 7: (a) Fiber-coupled optical output power of a science laser vs. injection current into [PITH_FULL_IMAGE:figures/full_fig_p012_7.png] view at source ↗
Figure 8
Figure 8. Figure 8: FIG. 8: (a) Photograph of the reference laser module with the micro-integrated bench [PITH_FULL_IMAGE:figures/full_fig_p013_8.png] view at source ↗
Figure 9
Figure 9. Figure 9: FIG. 9: Photograph of the distribution system used for the INTENTAS flight system. [PITH_FULL_IMAGE:figures/full_fig_p014_9.png] view at source ↗
Figure 10
Figure 10. Figure 10: FIG. 10: Schematic overview of the microwave source. Modified with permission from [ [PITH_FULL_IMAGE:figures/full_fig_p021_10.png] view at source ↗
Figure 11
Figure 11. Figure 11: FIG. 11: Schematic overview of the experimental sequence. [PITH_FULL_IMAGE:figures/full_fig_p023_11.png] view at source ↗
Figure 12
Figure 12. Figure 12: FIG. 12: Expected theoretical squeezing parameter [PITH_FULL_IMAGE:figures/full_fig_p024_12.png] view at source ↗
Figure 13
Figure 13. Figure 13: FIG. 13: Single shot fractional frequency sensitivity as a function of Ramsey time for a [PITH_FULL_IMAGE:figures/full_fig_p025_13.png] view at source ↗
read the original abstract

The INTENTAS project aims to develop an atomic sensor utilizing entangled Bose-Einstein condensates (BECs) in a microgravity environment. This key achievement is necessary to advance the capability for measurements that benefit from both entanglement-enhanced sensitivities and extended interrogation times. The project addresses significant challenges related to size, weight, and power management (SWaP) specific to the experimental platform at the Einstein-Elevator in Hannover. The design ensures a low-noise environment essential for the creation and detection of entanglement. Additionally, the apparatus features an innovative approach to the all-optical creation of BECs, providing a flexible system for various configurations and meeting the requirements for rapid turnaround times. Successful demonstration of this technology in the Einstein-Elevator will pave the way for a future deployment in space, where its potential applications will unlock high-precision quantum sensing.

Editorial analysis

A structured set of objections, weighed in public.

Desk editor's note, referee report, simulated authors' rebuttal, and a circularity audit. Tearing a paper down is the easy half of reading it; the pith above is the substance, this is the friction.

Referee Report

1 major / 0 minor

Summary. The manuscript outlines the INTENTAS project, which aims to develop an atomic sensor utilizing entangled Bose-Einstein condensates (BECs) in microgravity. It focuses on addressing size, weight, and power (SWaP) constraints for the Einstein-Elevator platform in Hannover, ensuring a low-noise environment for entanglement creation and detection, and implementing an all-optical approach to BEC creation for flexibility and rapid turnaround. The paper states that successful demonstration will pave the way for space deployment and high-precision quantum sensing applications.

Significance. If the proposed design proves feasible and is successfully demonstrated, the work could contribute to advancing entanglement-enhanced quantum sensors by enabling longer interrogation times in microgravity. However, the manuscript provides no supporting data, simulations, or analysis, so any significance remains entirely prospective and conditional on future unverified engineering outcomes.

major comments (1)
  1. [Abstract] Abstract: The statements that the design 'ensures a low-noise environment essential for the creation and detection of entanglement' and meets 'requirements for rapid turnaround times' via all-optical BEC creation constitute the central engineering claims, yet they are presented without any quantitative estimates, noise budgets, SWaP calculations, or references to prior validation that would allow assessment of feasibility.

Simulated Author's Rebuttal

1 responses · 1 unresolved

We thank the referee for their review. The manuscript is a design proposal for the INTENTAS sensor, and we address the concern about the abstract below.

read point-by-point responses
  1. Referee: [Abstract] Abstract: The statements that the design 'ensures a low-noise environment essential for the creation and detection of entanglement' and meets 'requirements for rapid turnaround times' via all-optical BEC creation constitute the central engineering claims, yet they are presented without any quantitative estimates, noise budgets, SWaP calculations, or references to prior validation that would allow assessment of feasibility.

    Authors: We agree this comment is valid. The abstract uses definitive phrasing for design goals that are elaborated conceptually in the manuscript body but without new quantitative analysis here. As this is a proposal paper, we will revise the abstract to use more precise language (e.g., 'is designed to ensure' and 'aims to meet') and add citations to prior literature on low-noise environments and all-optical BEC production. Detailed budgets remain outside the current scope. revision: yes

standing simulated objections not resolved
  • The manuscript contains no supporting data, simulations, or quantitative analysis, which is inherent to its nature as a forward-looking design proposal rather than a completed experiment report.

Circularity Check

0 steps flagged

No significant circularity; descriptive project proposal

full rationale

The manuscript is a project proposal describing planned development and design features for an atomic sensor rather than reporting completed measurements, derivations, equations, or proofs. No load-bearing steps exist that reduce by the paper's own equations or self-citation to its inputs; the central claim is conditional on future successful demonstration and does not rest on any specific unverified technical assertion or hidden assumption within the text itself.

Axiom & Free-Parameter Ledger

0 free parameters · 0 axioms · 0 invented entities

No free parameters, axioms, or invented entities are introduced because the paper is a high-level project description without mathematical derivations or new physical postulates.

pith-pipeline@v0.9.0 · 5804 in / 1141 out tokens · 27150 ms · 2026-05-23T21:26:27.657066+00:00 · methodology

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.

Reference graph

Works this paper leans on

62 extracted references · 62 canonical work pages · 1 internal anchor

  1. [1]

    The source chamber consists of an atom oven providing the atoms for experimentation and, additionally, a two-dimensional magneto-optical trap including a directional laser beam able to transport pre-cooled atoms into the science chamber (2D +-MOT)

  2. [2]

    The chamber is equipped with a total of ten ports

    The central chamber, which is referred to as the science chamber, is where the actual atom cooling, BEC creation, and state manipulation is performed. The chamber is equipped with a total of ten ports. Seven of the ten ports have an inner diameter of 20 mm and are equipped with windows that feature an anti-reflective (AR) coating for 780 nm light. Two of ...

  3. [3]

    It houses the vacuum pumps and vacuum gauges

    The support chamber is connected to the second large port on the opposite side of the detection lens. It houses the vacuum pumps and vacuum gauges. Furthermore, it is also equipped with a viewport (b) to enable further optical access to the science chamber. 4 The different chambers are made of titanium in order to comply with the requirements on magnetic ...

  4. [4]

    This enables robust science laser modules (see Fig

    Laser sources In order to cope with the requirements of the EE, the science lasers are implemented as fiber-coupled micro-integrated diode laser modules using a technology based on the one described in [31, 32]. This enables robust science laser modules (see Fig. 6) with a footprint of 139 mm × 80 mm and a mass of approximately 0.8 kg. FIG. 6: Photograph ...

  5. [5]

    It consists of a distributed feedback laser (DFB) with a dedicated frequency modulation spectroscopy (FMS) unit based on the design in [35, 36]

    Reference laser As reference for the science lasers, a micro-integrated frequency reference module is used. It consists of a distributed feedback laser (DFB) with a dedicated frequency modulation spectroscopy (FMS) unit based on the design in [35, 36]. The design has been adapted to the needs of the sensor by modifying the laser and spectroscopy units and...

  6. [6]

    The compartment housing this fiber network, is shown in Fig

    Beat detection system The beat detection system is entirely fiber-based. The compartment housing this fiber network, is shown in Fig. 8b. The fiber-based approach, with higher losses but also less need for maintenance, is preferred for long continuous measurement campaigns due to its expected stability gain in comparison with free-space optics. In the sys...

  7. [7]

    To achieve the high efficiencies required to provide sufficient optical power to the sensor head, a free-space approach is chosen

    Distribution System The major part of the output power of the science lasers is sent to the distribution module. To achieve the high efficiencies required to provide sufficient optical power to the sensor head, a free-space approach is chosen. A picture of the system can be found in Fig. 9. The design follows similar system designs, which already proved t...

  8. [8]

    While larger in footprint and not optimized for usage in the Einstein-Elevator, it offers more flexibility in terms of power and modifications for the initial testing on the ground

    Ground laser system In order to test capabilities on the ground another fiber-coupled laser system has been developed. While larger in footprint and not optimized for usage in the Einstein-Elevator, it offers more flexibility in terms of power and modifications for the initial testing on the ground. The laser system provides 485 mW of total power to the 2...

  9. [9]

    Self-built telescopes produce collimated beams with circular or linear polarization

    Optical system The light from the laser system is guided to the science and source chamber by polar- ization maintaining fibers. Self-built telescopes produce collimated beams with circular or linear polarization. In the 2D +-MOT , the pusher beam is propagating towards the center of the differential pumping stage, pushing the atoms towards the 3D-MOT. Th...

  10. [10]

    The 2D +-MOT is surrounded by four coils

    Magnetic field coils Next to the optical capabilities, there are several coils added to provide magneto-optical trapping and manipulation. The 2D +-MOT is surrounded by four coils. Those are rect- angular shaped, have a size of 47 mm × 83 mm, and are separated by 62 mm. Each coil comprises 64 windings of Kapton-insulated copper wire with a diameter of 0.7...

  11. [11]

    The resulting focal point has a diameter of 10.5 µm (1/e2) and a Rayleigh range of 320 µm

    to the intersection point and cross under an angle of 30 °. The resulting focal point has a diameter of 10.5 µm (1/e2) and a Rayleigh range of 320 µm. With this system it is possible to bring up to 10 W per beam to the location of the atoms. To increase the trapping volume of the cODT, both beams can be spatially modulated in two directions. Given the cap...

  12. [12]

    Alonso et al., Cold atoms in space: community workshop summary and proposed road-map, EPJ Quantum Technology 9, 30 (2022)

    I. Alonso et al., Cold atoms in space: community workshop summary and proposed road-map, EPJ Quantum Technology 9, 30 (2022)

  13. [13]

    Herrmann et al., Test of the Gravitational Redshift with Galileo Satellites in an Eccentric Orbit, Phys

    S. Herrmann et al., Test of the Gravitational Redshift with Galileo Satellites in an Eccentric Orbit, Phys. Rev. Lett. 121, 231102 (2018)

  14. [14]

    S. Peil, S. Crane, J. L. Hanssen, T. B. Swanson, and C. R. Ekstrom, Tests of local position invariance using continuously running atomic clocks, Phys. Rev. A 87, 010102 (2013)

  15. [15]

    Battelier et al., Exploring the foundations of the physical universe with space tests of the equivalence principle, Experimental Astronomy 51, 1695 (2021)

    B. Battelier et al., Exploring the foundations of the physical universe with space tests of the equivalence principle, Experimental Astronomy 51, 1695 (2021)

  16. [16]

    Dimopoulos, P

    S. Dimopoulos, P. W. Graham, J. M. Hogan, M. A. Kasevich, and S. Rajendran, Gravitational wave detection with atom interferometry, Phys. Lett. B 678, 37 (2009). 26

  17. [17]

    Douch, H

    K. Douch, H. Wu, C. Schubert, J. M¨ uller, and F. Pereira dos Santos, Simulation-based evalu- ation of a cold atom interferometry gradiometer concept for gravity field recovery, Advances in Space Research 61, 1307 (2018)

  18. [18]

    Carraz, C

    O. Carraz, C. Siemes, L. Massotti, R. Haagmans, and P. Silvestrin, A spaceborne gravity gradiometer concept based on cold atom interferometers for measuring earth’s gravity field, Microgravity Science and Technology 26, 139 (2014)

  19. [19]

    Chiow, J

    S.-w. Chiow, J. Williams, and N. Yu, Laser-ranging long-baseline differential atom interfer- ometers for space, Physical Review A 92, 063613 (2015)

  20. [20]

    L´ ev` equeet al

    T. L´ ev` equeet al. , Correlated atom accelerometers for mapping the Earth gravity field from Space, in International Conference on Space Optics , Vol. 11180 (SPIE, 2019) p. 344

  21. [21]

    Jekeli, Navigation Error Analysis of Atom Interferometer Inertial Sensor, NAVIGATION 52, 1 (2005)

    C. Jekeli, Navigation Error Analysis of Atom Interferometer Inertial Sensor, NAVIGATION 52, 1 (2005)

  22. [22]

    Canuel et al

    B. Canuel et al. , Six-axis inertial sensor using cold-atom interferometry, Physical Review Letters 97, 010402 (2006)

  23. [23]

    Cheiney et al

    P. Cheiney et al. , Navigation-Compatible Hybrid Quantum Accelerometer Using a Kalman Filter, Physical Review Applied 10, 034030 (2018)

  24. [24]

    Barrett, P

    B. Barrett, P. Cheiney, B. Battelier, F. Napolitano, and P. Bouyer, Multidimensional Atom Optics and Interferometry, Physical Review Letters 122, 043604 (2019)

  25. [25]

    Gersemann, M

    M. Gersemann, M. Gebbe, S. Abend, C. Schubert, and E. M. Rasel, Differential interferometry using a Bose-Einstein condensate, The European Physical Journal D 74, 203 (2020)

  26. [26]

    Abend et al., Technology roadmap for cold-atoms based quantum inertial sensor in space, AVS Quantum Science 5, 019201 (2023)

    S. Abend et al., Technology roadmap for cold-atoms based quantum inertial sensor in space, AVS Quantum Science 5, 019201 (2023)

  27. [27]

    M¨ untingaet al

    H. M¨ untingaet al. , Interferometry with Bose-Einstein Condensates in Microgravity, Phys. Rev. Lett. 110, 093602 (2013)

  28. [28]

    Becker et al., Space-borne Bose–Einstein condensation for precision interferometry, Nature 562, 391–395 (2018)

    D. Becker et al., Space-borne Bose–Einstein condensation for precision interferometry, Nature 562, 391–395 (2018)

  29. [29]

    Condon et al., All-optical bose-einstein condensates in microgravity, Phys

    G. Condon et al., All-optical bose-einstein condensates in microgravity, Phys. Rev. Lett. 123, 240402 (2019)

  30. [30]

    Pelluet et al

    C. Pelluet et al. , Atom interferometry in an Einstein Elevator 10.48550/arXiv.2407.07183 (2024). 27

  31. [31]

    Feldmann et al

    P. Feldmann et al. , Optimal squeezing for high-precision atom interferometers 10.48550/arXiv.2311.10241 (2023)

  32. [32]

    R. Roy, A. Green, R. Bowler, and S. Gupta, Rapid cooling to quantum degeneracy in dynam- ically shaped atom traps, Phys. Rev. A 93, 043403 (2016)

  33. [33]

    L¨ uckeet al., Twin matter waves for interferometry beyond the classical Limit, Science 334, 773 (2011)

    B. L¨ uckeet al., Twin matter waves for interferometry beyond the classical Limit, Science 334, 773 (2011)

  34. [34]

    Kruse et al

    I. Kruse et al. , Improvement of an Atomic Clock using Squeezed Vacuum, Physical Review Letters 117, 143004 (2016)

  35. [35]

    Cassens, B

    C. Cassens, B. Meyer-Hoppe, E. Rasel, and C. Klempt, An entanglement-enhanced atomic gravimeter 10.48550/arXiv.2404.18668

  36. [36]

    All-optical production of Bose-Einstein condensates with 2 Hz repetition rate

    M. Hetzel, M. Quensen, J. S. Haase, and C. Klempt, All-optical production of Bose-Einstein condensates with 2 Hz repetition rate 10.48550/arXiv.2406.16488 (2024)

  37. [37]

    Rudolph et al

    J. Rudolph et al. , A high-flux BEC source for mobile atom interferometers, New Journal of Physics 17, 065001 (2015)

  38. [38]

    Lotz, Untersuchungen zu Einflussfaktoren auf die Qualit¨ at von Experimenten unter Mikro- gravitation im Einstein-Elevator, Ph.D

    C. Lotz, Untersuchungen zu Einflussfaktoren auf die Qualit¨ at von Experimenten unter Mikro- gravitation im Einstein-Elevator, Ph.D. thesis, Gottfried Wilhelm Leibniz Universit¨ at Han- nover (2022)

  39. [39]

    Raudonis et al., Microgravity facilities for cold atom experiments, Quantum Science and Technology 8, 044001 (2023)

    M. Raudonis et al., Microgravity facilities for cold atom experiments, Quantum Science and Technology 8, 044001 (2023)

  40. [40]

    S. T. Seidel, Eine Quelle f¨ ur die Interferometrie mit Bose-Einstein-Kondensaten auf Hoehen- forschungsraketen, Ph.D. thesis, Gottfried Wilhelm Leibniz Universit¨ at Hannover (2014)

  41. [41]

    Altarev et al

    I. Altarev et al. , Minimizing magnetic fields for precision experiments, Journal of Applied Physics 117, 233903 (2015)

  42. [42]

    K¨ urbiset al

    C. K¨ urbiset al. , Extended cavity diode laser master-oscillator-power-amplifier for operation of an iodine frequency reference on a sounding rocket, Applied Optics 59, 253 (2020)

  43. [43]

    Hirsch et al

    J. Hirsch et al. , Micro-integrated diode laser modules for operation in quantum technology applications, Proc. SPIE Int. Soc. Opt. Eng. 12912, 129120B (2024)

  44. [44]

    Schiemangk et al., Accurate frequency noise measurement of free-running lasers, Applied Optics 53, 7138 (2014)

    M. Schiemangk et al., Accurate frequency noise measurement of free-running lasers, Applied Optics 53, 7138 (2014)

  45. [45]

    Di Domenico, S

    G. Di Domenico, S. Schilt, and P. Thomann, Simple approach to the relation between laser frequency noise and laser line shape, Applied Optics 49, 4801 (2010). 28

  46. [46]

    Strangfeld et al

    A. Strangfeld et al. , Prototype of a compact rubidium-based optical frequency reference for operation on nanosatellites, J. Opt. Soc. Am. B 38, 1885 (2021)

  47. [47]

    Strangfeld, B

    A. Strangfeld, B. Wiegand, J. Kluge, M. Schoch, and M. Krutzik, Compact plug and play optical frequency reference device based on doppler-free spectroscopy of rubidium vapor, Opt. Express 30, 12039 (2022)

  48. [48]

    Wiegand, B

    B. Wiegand, B. Leykauf, R. J¨ ordens, and M. Krutzik, Linien: A versatile, user-friendly, open- source FPGA-based tool for frequency stabilization and spectroscopy parameter optimization, Review of Scientific Instruments 93 (2022)

  49. [49]

    Pahl et al

    J. Pahl et al. , Compact and robust diode laser system technology for dual-species ultracold atom experiments with rubidium and potassium in microgravity, Appl. Opt. 58, 5456 (2019)

  50. [50]

    M. A. Popp, Compact, low-noise current drivers for quantum sensors with atom chips, Ph.D. thesis, Gottfried Wilhelm Leibniz Universit¨ at Hannover (2018)

  51. [51]

    Herbst et al

    A. Herbst et al. , High-flux source system for matter-wave interferometry exploiting tunable interactions, Physical Review Research 6 (2024)

  52. [52]

    Ammann and N

    H. Ammann and N. Christensen, Delta Kick Cooling: A New Method for Cooling Atoms, Phys. Rev. Lett. 78, 2088 (1997)

  53. [53]

    Meyer-Hoppe et al

    B. Meyer-Hoppe et al. , Dynamical low-noise microwave source for cold-atom experiments, Review of Scientific Instruments 94, 074705 (2023)

  54. [54]

    Bourdeauducq et al., Artiq: Advanced real-time infrastructure for quantum physics (2021)

    S. Bourdeauducq et al., Artiq: Advanced real-time infrastructure for quantum physics (2021)

  55. [55]

    Sinara Hardware Project, Sinara: Open-Source Hardware for Quantum Physics (2024), ac- cessed: 2024-07-30

  56. [56]

    Kasprowicz et al

    G. Kasprowicz et al. , Urukul – Open-source Frequency Synthesizer Module for Quantum Physics, International Journal of Electronics and Telecommunications 68, 123 (2022)

  57. [57]

    Grafana Labs, Grafana documentation (2018)

  58. [58]

    Pezz` e, A

    L. Pezz` e, A. Smerzi, M. K. Oberthaler, R. Schmied, and P. Treutlein, Quantum metrology with nonclassical states of atomic ensembles, Rev. Mod. Phys. 90, 035005 (2018)

  59. [59]

    C. D. Hamley, C. S. Gerving, T. M. Hoang, E. M. Bookjans, and M. S. Chapman, Spin-nematic squeezed vacuum in a quantum gas, Nature Physics 8, 305–308 (2012)

  60. [60]

    Szmuk, V

    R. Szmuk, V. Dugrain, W. Maineult, J. Reichel, and P. Rosenbusch, Stability of a trapped- atom clock on a chip, Phys. Rev. A 92, 012106 (2015). 29

  61. [61]

    Laurent et al., Qualification and frequency accuracy of the space-based primary frequency standard PHARAO, Metrologia 57, 055005 (2020)

    P. Laurent et al., Qualification and frequency accuracy of the space-based primary frequency standard PHARAO, Metrologia 57, 055005 (2020)

  62. [62]

    Anders et al

    F. Anders et al. , Momentum Entanglement for Atom Interferometry, Phys. Rev. Lett. 127, 140402 (2021). 30