The reviewed record of science sign in
Pith

arxiv: 2409.03356 · v2 · pith:4E5IGVXP · submitted 2024-09-05 · cond-mat.supr-con · astro-ph.IM· physics.ins-det

Magnetic field tunable spectral response of kinetic inductance detectors

Reviewed by Pith T0 review T1 audit T2 compute T3 formal T4 kernel pith:4E5IGVXPrecord.jsonopen to challenge →

classification cond-mat.supr-con astro-ph.IMphysics.ins-det
keywords fieldmagneticresponsedetectorsinductancekineticsuperconductingfactor
0
0 comments X
read the original abstract

We tune the onset of optical response in aluminium kinetic inductance detectors from a natural cutoff frequency of 90 GHz to 60 GHz by applying an external magnetic field. The change in spectral response is due to the decrease of the superconducting gap, from 90 GHz at zero magnetic field to 60 GHz at a magnetic field of around 3 mT. We characterize the variation of the superconducting gap, the detector frequency shift and the internal quality factor as a function of the applied field. In principle, the magnetic field tunable response could be used to make spectroscopic measurements. In practice, the internal quality factor behaves hysteretically with the magnetic field due to the presence of vortices in the thin superconducting film. We conclude by discussing possible solutions to achieve spectroscopy measurements using kinetic inductance detectors and magnetic field.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.