Pith. sign in

REVIEW 4 major objections 4 minor 29 references

Noncovalent force spectroscopy using wide-field optical and diamond-based magnetic imaging

T0 review · 4 major / 4 minor · reviewed 2026-08-14 · deepseek-v4-flash

Pith's one-line read A diamond-based magnetic imaging system resolves individual microspheres and distinguishes specific binding from control surfaces by rupture behavior.

desk verdict The magnetic imaging half is genuinely new and credible, but the quantitative rupture force is not supported; as a proof-of-principle detection demonstration it deserves a serious referee. read the letter →

arxiv 1908.10052 v1 pith:4DW4ES2N submitted 2019-08-27 physics.ins-det physics.app-phphysics.bio-phquant-ph

classification physics.ins-detphysics.app-phphysics.bio-phquant-ph
keywords nitrogen-vacancycentersdiamondmagnetometryforce-inducedremnantmagnetizationspectroscopybiotin-streptavidinruptureforcemagneticmicrosphereswide-fieldimagingopticallydetectedresonance
verification ladder T0 review T1 audit T2 compute T3 formal

The pith

A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.

The reading

This paper reports a proof-of-principle realization of force-induced remnant magnetization spectroscopy (FIRMS) in which the usual vapor-cell magnetometer is replaced by a planar ensemble of nitrogen-vacancy (NV) color centers in diamond, imaged through a wide-field microscope. The paper tries to establish that this detector can resolve individual magnetic microspheres, read their magnetic dipole orientation, and distinguish biotin-streptavidin binding from non-biotin control surfaces by the drive voltage at which beads detach under oscillatory piezoelectric force. If correct, the result matters because it makes FIRMS a micron-scale, background-free, parallel technique: many beads in one field of view can be tracked at once, and the near-field sensor geometry raises the detected signal by nine orders of magnitude over vapor-cell FIRMS. The measured biotin-streptavidin rupture force is 20(8) pN, two to three times below literature values at a comparable loading rate, a difference the authors attribute to their oscillatory force protocol and to chamber-to-chamber variations in force calibration.

What carries the argument

The load-bearing object is the NV ensemble as an imaging magnetometer: a near-surface layer of nitrogen-vacancy color centers whose optically detected magnetic resonance frequency shifts with local magnetic field, converted pixel by pixel into magnetic field images. The force calibration is carried by the inertial relation $F = m\omega^2 x_0$, with $x_0$ the chamber oscillation amplitude measured by a Michelson interferometer through the harmonic content $N = 2\pi x_0/\lambda$ of the photodiode signal. This relation converts a piezo drive voltage into a force per bead and makes the reported rupture force quantitative; the NV layer itself supplies micron-scale, background-free detection that resolves individual beads whose fields are nine orders of magnitude stronger than those detected by vapor-cell FIRMS.

What would settle it

Track a single bead's motion relative to the chamber during piezo driving, for example by high-speed video through the same objective; if the bead's trajectory does not track the interferometrically measured chamber amplitude $x_0$ with the same phase and frequency, then $F = m\omega^2 x_0$ is not the force on the bead and the reported 20(8) pN rupture force is not a true bond strength.

Watch

Extended reading notes

Core claim

The central claim is that an NV-diamond imaging magnetometer can serve as the readout for FIRMS at the level of individual particles. Streptavidin-coated magnetic beads are bound to biotinylated diamond surfaces and driven by piezoelectric oscillation; as the drive voltage is ramped, wide-field optical tracking records when each bead detaches, while magnetic images obtained by fitting optically detected magnetic resonance spectra pixel by pixel show each bead as a magnetic dipole and reveal orientation changes, including a 10(2) degree rotation during an ambiguous motion event. The setup resolves single 2-micron beads, detects the direction of each bead's magnetic dipole, and reproducibly separates biotin-coated from control surfaces; the reported rupture force of 20(8) pN is presented as a force-regime effect of oscillatory loading with an effective loading rate of about 70 pN/s. In the authors' words, at the current stage it is clear that the detection system can distinguish different surface properties.

Load-bearing premise

The load-bearing premise is that each bead feels the same acceleration as the chamber oscillation, so the applied force is $F = m\omega^2 x_0$; the bead and the surrounding fluid must move rigidly with the chamber, and air bubbles or fluid compressibility can break this, which the authors note may explain the 5-40% variation in force-per-volt calibration between chambers.

Editorial extensions

If this is right

  • Individual 2-micron beads can be imaged magnetically at diffraction-limited resolution, so bond rupture can be followed one particle at a time rather than as an ensemble average.
  • Magnetic dipole orientation is resolved, so events such as sliding, rolling, or detach-and-reattach become visible even when the optical image is ambiguous.
  • Because the NV layer sits microns from the beads, the detected field is about nine orders of magnitude larger than in vapor-cell FIRMS, opening the way to nanoscale samples and to imaging through opaque media.
  • Biotin-functionalized and control surfaces give reproducible, distinguishable detachment curves, supporting the use of the method for screening surface functionalizations.
  • Rupture forces measured under oscillatory loading come out two to three times below constant-loading literature values, so force-regime corrections such as the effective loading rate of about 70 pN/s are needed when comparing results.

Reading between the lines

Editorial extensions of the paper, not claims the author makes directly.

  • Beyond the paper: if force application is switched to a non-inertial mechanism such as dielectrophoresis, the same NV readout could turn every bead in the field of view into an independent force sensor, giving per-bead rupture-force distributions instead of ensemble detachment curves.
  • Beyond the paper: the demonstrated orientation tracking could classify each detachment event as rupture, rolling, or sliding; counting rolling and sliding events separately may explain part of the apparent difference between biotin and control surfaces without invoking a change in bond strength.
  • Beyond the paper: because the near-field geometry provides a nine-order-of-magnitude signal gain over vapor-cell FIRMS, a direct next test is whether single 100-nm or sub-micron magnetic labels become detectable, which would extend the technique to intracellular and blood-based assays.
Share X Bluesky LinkedIn Reddit HN

Signed reviews

No signed human review yet.

Editorial analysis

A structured set of objections, weighed in public.

Desk editor's note, referee report, and a circularity audit.

Referee Report

4 major / 4 minor

Summary. The manuscript reports a proof-of-principle realization of force-induced remnant magnetization spectroscopy (FIRMS) using wide-field optical microscopy and nitrogen-vacancy (NV) diamond magnetometry. Streptavidin-coated magnetic microbeads were allowed to bind to biotinylated and control diamond surfaces in sealed microfluidic chambers, which were then shaken by piezoelectric actuators at 5–10 kHz with a slowly ramping voltage. A Michelson interferometer measured the mechanical oscillation amplitude x0, and the force on a bead was computed as F = mω²x0. Detachment was tracked optically; magnetic images were used to resolve individual bead dipoles and a 10(2)-degree orientation change of one bead. The authors report a biotin-streptavidin rupture force of 20(8) pN, about 2–3 times below literature values, and claim that the system can distinguish biotin from control surfaces, with an estimated effective loading rate of ≈70 pN/s.

Significance. If the results hold, this would be a useful step toward parallel, background-free force spectroscopy at the single-microparticle level: the NV magnetometer provides micron-scale spatial resolution, direct magnetic-field images with uncertainty maps, and field measurements of about 0.3 G from single beads, a large gain over vapor-cell FIRMS. The optical and magnetic images (Figs. 5–7) are presented as actual data, and the paper ships direct experimental images rather than simulations. However, the significance of the quantitative claims is limited by an unverified force-calibration model and by chamber-to-chamber variability; the demonstrated capability is currently stronger at the level of single-bead magnetic imaging and voltage-relative detachment than at the level of absolute rupture-force determination.

major comments (4)
  1. [Section III, Fig. 6] The force scale F = mω²x0 uses an unspecified mass m; the Methods section provides only the effective weight (3.2×10^-14 N) of a 2 µm particle, not its mass or whether buoyancy is included. If m is the actual bead mass (≈7.5×10^-15 kg), the buoyancy-corrected effective mass is lower by a factor ρ_p/(ρ_p−ρ_f) ≈ 2.25, which changes the reported 20(8) pN rupture force by about that factor. The paper should state which mass is used and justify the choice.
  2. [Sections II.D and IV, Fig. 6] The conversion from interferometer amplitude to bead force assumes that the bead moves rigidly with the chamber, i.e., F = mω²x0. This assumption is not validated at the bead level, and Section IV itself flags compressibility and air bubbles as possible violations; Fig. 6 shows 5–40% chamber-to-chamber variation in pN/V. The reported 20(8) pN rupture force and the estimated 70 pN/s effective loading rate are therefore not supported unless a direct calibration (for example, bead tracking under oscillation or a known viscous-drag model) is provided. In the absence of such validation, the quantitative force claims should be removed or explicitly labeled preliminary.
  3. [Section III, Fig. 5] The central claim that the system 'can distinguish different surface properties' is based on comparing only two biotin and two control chambers, while the text states that results 'often deviated significantly' when a new chamber was assembled. The detachment curves are plotted against drive voltage, not force, and no error bars or statistical test are given; since the voltage-to-force conversion itself depends on chamber and frequency, the biotin/control difference is confounded by chamber-to-chamber mechanical variability. A quantitative comparison (for example, per-chamber force thresholds with uncertainties, or a test against the null hypothesis that the two groups arise from the same chamber-to-chamber distribution) is needed to support the claim.
  4. [Section III, Fig. 5] The drive frequency used for the detachment experiments in Fig. 5 is not stated. Because Fig. 6 shows that force-per-volt varies strongly with frequency and chamber, the voltage axis in Fig. 5 is not an unambiguous force scale unless the frequency is specified for each run and the frequency response is accounted for.
minor comments (4)
  1. [Section III, Fig. 7] The method used to fit the bead orientation from the magnetic images is not described; the reported 10(2)-degree rotation should be accompanied by a description of the fitting procedure (for example, a dipole model and the number of fit parameters).
  2. [Section III] The sentence comparing the measured 0.3 G bead field with the 3×10^-14 T vapor-cell value conflates the measured field from a single bead with the sensor sensitivity; rephrasing would avoid implying equal noise floors.
  3. [Section II.B] The paper would benefit from stating the typical pixel size and field of view of the imaging magnetometer, since these are relevant to the claim of diffraction-limited single-bead resolution.
  4. [Fig. 5 caption] The numbers labeling the individual curves (36, 32, 19, etc.) are not identified in the caption until the following sentence; clarify that these are the numbers of tracked particles per experiment.

Circularity Check

0 steps flagged · score 0.0 of 10

No significant circularity: the force calibration is independently measured and the biotin/control discrimination is a direct observation, not a fit to the claim.

full rationale

The paper's derivation chain is self-contained for its central claim. The force axis is constructed from an independent Michelson interferometer measurement of the chamber oscillation amplitude x0, combined with F = mω²x0; the harmonic-count fit for N = 2πx0/λ is a standard physical model, and no detachment statistic is used to calibrate the force axis. The biotin/control comparison is a direct empirical observation of detachment curves; the 20(8) pN estimate is read off the calibrated axis and then compared with, not fitted to, literature values. The authors' citations to their own prior FIRMS work (Refs. 6, 7, 20, 21) supply technique context and a dielectrophoresis outlook, but they are not used to define the measured force or to validate the claimed surface discrimination. Possible concerns about inertial coupling and chamber-to-chamber variation are correctness risks, not circularity, and the paper itself flags them in Section IV. No equation reduces to another by construction, and no fitted parameter is renamed as a prediction.

Assumptions & free parameters 0 free parameters · 3 assumptions · 0 invented entities

The central result rests on three domain assumptions: NV magnetometry performance, the inertial force model for bead detachment, and the interpretation of differential detachment as specific versus nonspecific binding. No free parameters are fitted to produce the headline rupture force; the voltage-to-force conversion comes from a separate interferometric calibration.

assumptions (3)
  • domain assumption NV centers in diamond provide a linear magnetic field response with gyromagnetic ratio 2.8 MHz/G and can be read out via ODMR.
    Invoked in Section II.B to convert measured resonance frequencies into magnetic field images. Standard and well-established, but not independently verified in this paper.
  • domain assumption The force experienced by a bound bead is inertial, F = m omega squared x0, with the bead rigidly following the chamber oscillation.
    Introduced in Section II.D and used in Section III to convert interferometric displacements into rupture forces. The authors later question this due to air bubbles and fluid compressibility (Section IV).
  • domain assumption Detachment events observed on biotin-functionalized surfaces are dominated by specific biotin-streptavidin bond rupture, while control surfaces show only nonspecific interactions.
    Underpins the claim that the system distinguishes specific binding. Not directly verified at the molecular level; the low-force feature could also arise from nonspecific adhesion.

how reviews work

0 comments
Cite this review

Pith. "Pith review of Noncovalent force spectroscopy using wide-field optical and diamond-based magnetic imaging." pith.science (2026). https://pith.science/paper/4DW4ES2N

@misc{pith2026190810052,
  author       = {Pith},
  title        = {Pith review of: Noncovalent force spectroscopy using wide-field optical and diamond-based magnetic imaging},
  year         = {2026},
  howpublished = {\url{https://pith.science/paper/4DW4ES2N}},
  note         = {Machine review of arXiv:1908.10052}
}
read the original abstract

A realization of the force-induced remnant magnetization spectroscopy (FIRMS) technique of specific biomolecular binding is presented where detection is accomplished with wide-field optical and diamond-based magnetometry using an ensemble of nitrogen-vacancy (NV) color centers. The technique may be adapted for massively parallel screening of arrays of nanoscale samples.

Figures

Figures reproduced from arXiv: 1908.10052 by the authors.

Figure 1
Figure 1. FIG. 1. Operating principle of NV-based magnetometry. (a) The NV [PITH_FULL_IMAGE:figures/full_fig_p002_1.png] view at source ↗
Figure 2
Figure 2. FIG. 2. Optical schematics for (a) imaging magnetometer and (b) [PITH_FULL_IMAGE:figures/full_fig_p003_2.png] view at source ↗
Figure 3
Figure 3. FIG. 3. 3-D rendering of the chamber. The chamber is composed of an imaging spacer sandwiched between two glass coverslips. Before [PITH_FULL_IMAGE:figures/full_fig_p004_3.png] view at source ↗
Figures from the paper (4 more)
Figure 4
Figure 4. Figure 4: FIG. 4. Power spectral density (PSD) of experimental photodiode data (a,b,c) alongside PSD of simulated data (d,e,f) with parameters chosen [PITH_FULL_IMAGE:figures/full_fig_p005_4.png]
Figure 5
Figure 5. Figure 5: FIG. 5. Normalized number of particles detached under increasingly [PITH_FULL_IMAGE:figures/full_fig_p005_5.png]
Figure 6
Figure 6. Figure 6: FIG. 6. Amplitude of the force experienced by a microsphere with di [PITH_FULL_IMAGE:figures/full_fig_p006_6.png]
Figure 7
Figure 7. Figure 7: FIG. 7. Images of particles before (top), during (middle), and after (bottom) application of force. Wide-field optical images are displayed on [PITH_FULL_IMAGE:figures/full_fig_p007_7.png]

Discussion (0). Continue with ORCID to comment.

Reference graph

Works this paper leans on

29 extracted references · 21 canonical work pages

  1. [1]

    merlin.mbs aapmrev4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked

    FUNCTION id.bst "merlin.mbs aapmrev4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked" ENTRY address archive archivePrefix author bookaddress booktitle chapter collaboration doi edition editor eid eprint howpublished institution isbn issn journal key language month note number organization pages primaryClass publisher school SLACcitation series title translat...

  2. [2]

    merlin.mbs aipauth4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked

    FUNCTION id.bst "merlin.mbs aipauth4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked" ENTRY address archive archivePrefix author bookaddress booktitle chapter collaboration doi edition editor eid eprint howpublished institution isbn issn journal key language month note number organization pages primaryClass publisher school SLACcitation series title translat...

  3. [3]

    merlin.mbs aipnum4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked

    FUNCTION id.bst "merlin.mbs aipnum4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked" ENTRY address archive archivePrefix author bookaddress booktitle chapter collaboration doi edition editor eid eprint howpublished institution isbn issn journal key language month note number organization pages primaryClass publisher school SLACcitation series title translati...

  4. [4]

    merlin.mbs apsrev4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked

    FUNCTION id.bst "merlin.mbs apsrev4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked" ENTRY address archive archivePrefix author bookaddress booktitle chapter collaboration doi edition editor eid eprint howpublished institution isbn issn journal key language month note number organization pages primaryClass publisher school SLACcitation series title translati...

  5. [5]

    merlin.mbs apsrmp4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked

    FUNCTION id.bst "merlin.mbs apsrmp4-1.bst 2010-07-25 4.21a (PWD, AO, DPC) hacked" ENTRY address archive archivePrefix author bookaddress booktitle chapter collaboration doi edition editor eid eprint howpublished institution isbn issn journal key language month note number organization pages primaryClass publisher school SLACcitation series title translati...

  6. [6]

    author author V. T. \ Moy , author E.-L. \ Florin , \ and\ author H. E. \ Gaub ,\ title title Intermolecular forces and energies between ligands and receptors , \ http://www.jstor.org/stable/2884765 journal journal Science \ volume 266 ,\ pages 257--259 ( year 1994 ) NoStop

  7. [7]

    \ Kim , author Y.-B

    author author M.-G. \ Kim , author Y.-B. \ Shin , author J.-M. \ Jung , author H.-S. \ Ro , \ and\ author B. H. \ Chung ,\ title title Enhanced sensitivity of surface plasmon resonance ( SPR ) immunoassays using a peroxidase-catalyzed precipitation reaction and its application to a protein microarray , \ 10.1016/j.jim.2004.12.007 journal journal Journal o...

  8. [8]

    author author P. M. \ Yangyuoru , author S. Dhakal , author Z. Yu , author D. Koirala , author S. M. \ Mwongela , \ and\ author H. Mao ,\ title title Single-molecule measurements of the binding between small molecules and DNA aptamers , \ 10.1021/ac300427d journal journal Analytical Chemistry \ volume 84 ,\ pages 5298--5303 ( year 2012 ) ,\ note pMID: 227...

Show all 29 references
  1. [9]

    Masetti , author H.-n

    author author M. Masetti , author H.-n. \ Xie , author Z. Krpeti\' c , author M. Recanatini , author R. A. \ Alvarez-Puebla , \ and\ author L. Guerrini ,\ title title Revealing DNA interactions with exogenous agents by surface-enhanced raman scattering , \ 10.1021/ja511398w jo...

  2. [10]

    Sitters , author D

    author author G. Sitters , author D. Kamsma , author G. Thalhammer , author M. Ritsch-Marte , author E. Peterman , \ and\ author G. Wuite ,\ title title Acoustic force spectroscopy , \ 10.1038/nmeth.3183 journal journal Nature Methods \ volume 12 ,\ pages 47--50 ( year 2015 ) NoStop

  3. [11]

    De Silva , author L

    author author L. De Silva , author L. Yao , author Y. Wang , \ and\ author S. Xu ,\ title title Well-defined and sequence-specific noncovalent binding forces of DNA , \ 10.1021/jp403817b journal journal The Journal of Physical Chemistry B \ volume 117 ,\ pages 7554--7558 ( yea...

  4. [12]

    Jia , author Y

    author author H. Jia , author Y. Wang , \ and\ author S. Xu ,\ title title Super-resolution force spectroscopy reveals ribosomal motion at sub-nucleotide steps , \ 10.1039/C8CC02658K journal journal Chem. Commun. \ volume 54 ,\ pages 5883--5886 ( year 2018 ) NoStop

  5. [13]

    Ota , author T

    author author T. Ota , author T. Sugiura , \ and\ author S. Kawata ,\ title title Rupture force measurement of biotin-streptavidin bonds using optical trapping , \ 10.1063/1.1999855 journal journal Applied Physics Letters \ volume 87 ,\ pages 043901 ( year 2005 ) NoStop

  6. [14]

    Wong , author E

    author author S. Wong , author E. Joselevich , author A. Woolley , author C. Cheung , \ and\ author C. Lieber ,\ title title Covalently functionalized nanotubes as nanometresized probes in chemistry and biology , \ 10.1038/27873 journal journal Nature \ volume 394 ,\ pages 52-...

  7. [15]

    Wong , author A

    author author J. Wong , author A. Chilkoti , \ and\ author V. T. \ Moy ,\ title title Direct force measurements of the streptavidin--biotin interaction , \ 10.1016/S1050-3862(99)00035-2 journal journal Biomolecular Engineering \ volume 16 ,\ pages 45--55 ( year 1999 ) NoStop

  8. [16]

    author author S. B. \ van Dam , author M. Walsh , author M. J. \ Degen , author E. Bersin , author S. L. \ Mouradian , author A. Galiullin , author M. Ruf , author M. IJspeert , author T. H. \ Taminiau , author R. Hanson , \ and\ author D. R. \ Englund ,\ title title Optical c...

  9. [17]

    Fescenko , author A

    author author I. Fescenko , author A. Laraoui , author J. Smits , author N. Mosavian , author P. Kehayias , author J. Seto , author L. Bougas , author A. Jarmola , \ and\ author V. M. \ Acosta ,\ title title Diamond magnetic microscopy of malarial hemozoin nanocrystals , \ 10....

  10. [18]

    https://www.adamasnano.com/ NoStop

    note Ad \'a mas Nanotechnologies. https://www.adamasnano.com/ NoStop

  11. [19]

    Jensen , author P

    author author K. Jensen , author P. Kehayias , \ and\ author D. Budker ,\ title Magnetometry with nitrogen-vacancy centers in diamond , \ in\ 10.1007/978-3-319-34070-8_18 booktitle High Sensitivity Magnetometers ,\ editor edited by\ editor A. Grosz , editor M. J. \ Haji-Sheikh...

  12. [20]

    Bar-Gill , author L

    author author N. Bar-Gill , author L. Pham , author A. Jarmola , author D. Budker , \ and\ author R. Walsworth ,\ title title Solid-state electronic spin coherence time approaching one second , \ 10.1038/ncomms2771 journal journal Nature communications \ volume 4 ,\ pages 1743...

  13. [21]

    Budker ,\ title title Diamond nanosensors: The sense of colour centres , \ 10.1038/nphys1989 journal journal Nature Physics \ volume 7 ,\ pages 453--454 ( year 2011 ) NoStop

    author author D. Budker ,\ title title Diamond nanosensors: The sense of colour centres , \ 10.1038/nphys1989 journal journal Nature Physics \ volume 7 ,\ pages 453--454 ( year 2011 ) NoStop

  14. [22]

    Hecht ,\ https://books.google.com/books?id=7aG6QgAACAAJ title Optics ,\ Pearson education\ ( publisher Addison-Wesley ,\ year 2002 )\ pp.\ pages 291--302 NoStop

    author author E. Hecht ,\ https://books.google.com/books?id=7aG6QgAACAAJ title Optics ,\ Pearson education\ ( publisher Addison-Wesley ,\ year 2002 )\ pp.\ pages 291--302 NoStop

  15. [23]

    author author E. A. \ Evans \ and\ author K. Ritchie ,\ title title Dynamic strength of molecular adhesion bonds. \ 10.1016/S0006-3495(97)78802-7 journal journal Biophysical journal \ volume 72 ,\ pages 1541--55 ( year 1997 ) NoStop

  16. [24]

    Merkel , author P

    author author R. Merkel , author P. Nassoy , author A. Leung , author K. Ritchie , \ and\ author E. Evans ,\ title title Energy landscapes of receptor-ligand bonds explored with dynamic force spectroscopy , \ 10.1038/16219 journal journal Nature \ volume 397 ,\ pages 50--53 ( ...

  17. [25]

    Yao \ and\ author S

    author author L. Yao \ and\ author S. Xu ,\ title title Force-induced remnant magnetization spectroscopy for specific magnetic imaging of molecules , \ 10.1002/anie.201007297 journal journal Angew. Chem. Int. Ed. \ volume 50 ,\ pages 4407--4409 ( year 2011 ) NoStop

  18. [26]

    Kayci , author S

    author author M. Kayci , author S. Lourette , author T. Wang , author L. Bougas , \ and\ author D. Budker ,\ title title Quantitative measurements of non-covalent interactions with diamond based magnetic imaging , \ 10.1063/1.5041513 journal journal Applied Physics Letters \ v...

  19. [27]

    Bustamante , author Z

    author author C. Bustamante , author Z. Bryant , \ and\ author S. B Smith ,\ title title Ten years of tension: Single-molecule DNA mechanics , \ 10.1038/nature01405 journal journal Nature \ volume 421 ,\ pages 423--7 ( year 2003 ) NoStop

  20. [28]

    \ Chen , author A

    author author Y.-T. \ Chen , author A. Kolhatkar , author O. Zenasni , author S. Xu , \ and\ author T. Randall Lee ,\ title title Biosensing using magnetic particle detection techniques , \ 10.3390/s17102300 journal journal Sensors \ volume 17 ,\ pages 2300 ( year 2017 ) NoStop

  21. [29]

    Liang , author X.-B

    author author H. Liang , author X.-B. \ Zhang , author Y. Lv , author L. Gong , author R. Wang , author X. Zhu , author R. Yang , \ and\ author W. Tan ,\ title title Functional DNA -containing nanomaterials: Cellular applications in biosensing, imaging, and targeted therapy , ...

Pith tools

Reviewed August 14, 2026 · model on record in the stance chip above.