pith. sign in

arxiv: 1103.0823 · v2 · pith:QVDYXBKHnew · submitted 2011-03-04 · ⚛️ physics.ins-det · nucl-ex

Development of portable NMR polarimeter system for polarized HD target

classification ⚛️ physics.ins-det nucl-ex
keywords systemportablepolarimeterconventionalpolarizationpolarizedtargetperformance
0
0 comments X
read the original abstract

A portable NMR polarimeter system has been developed to measure the polarization of a polarized Hydrogen-Deuteride (HD) target for hadron photoproduction experiments at SPring-8. The polarized HD target is produced at the Research Center for Nuclear Physics (RCNP), Osaka university and is transported to SPring-8. The HD polarization should be monitored at both places. We have constructed the portable NMR polarimeter system by replacing the devices in the conventional system with the software system with PCI eXtensions for Instrumentation (PXI). The weight of the NMR system is downsized from 80 kg to 7 kg, and the cost is reduced to 25%. We check the performance of the portable NMR polarimeter system. The signal-to-noise (S/N) ratio of the NMR signal for the portable system is about 50% of that for the conventional NMR system. This performance of the portable NMR system is proved to be compatible with the conventional NMR system for the polarization measurement.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.