REVIEW 3 major objections 5 minor 112 references
In vivo oxygen measurements argue against oxygen depletion as the mechanism of FLASH radiotherapy's normal-tissue sparing.
Reviewed by Pith at T0; open to challenge. T0 means a machine referee read the full paper against a public rubric. the ladder, T0–T4 →
T0 review · deepseek-v4-flash
2026-08-01 16:27 UTC pith:CY5F4T5M
load-bearing objection A coherent, self-aware review that makes a plausible case against bulk oxygen depletion as the FLASH mechanism, but the abstract overstates the oxygen-window conclusion and the negative argument rests on bulk-probe, mostly skin gO2 values. the 3 major comments →
Review of oxygen measurement and relevance to the mechanisms of FLASH radiotherapy
The pith
A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.
Core claim
The paper's central claim is that measured in vivo radiolytic oxygen consumption during ultra-high dose rate irradiation is too small to induce radiobiological hypoxia. Using phosphorescence-lifetime oxygen probes (Oxyphor) in mice, the measured gO2 values of 0.1–0.5 mmHg per Gray imply that a typical 20 Gy FLASH dose depletes tissue oxygen by only 2–4 mmHg, far below the threshold for hypoxic radioprotection. The review also synthesizes oxygen-modulation experiments showing that FLASH sparing appears only at intermediate baseline pO2—disappearing under both clamping-induced hypoxia and hyperoxia—and pairs this with in vitro radiation chemistry showing UHDR shifts radical yields toward solva
What carries the argument
The gO2 value—radiolytic oxygen consumption per unit dose, measured in mmHg per Gray—is the central quantitative object that carries the argument; its small magnitude and its saturation at baseline pO2 above roughly 15–25 mmHg undermine the oxygen-depletion hypothesis. A second load-bearing idea is the pO2 window: FLASH sparing occurs only when baseline tissue oxygen lies in an intermediate range, vanishing under both hypoxia and hyperoxia. The measurement system that makes the central claim possible is oxygen-quenched phosphorescence with molecular reporters, which provides millisecond-resolution, radiation-insensitive, localized pO2 readings before, during, and after UHDR pulses.
Load-bearing premise
The measured gO2 values from Oxyphor probes in skin accurately represent the oxygen consumed in the irradiated tissue volume; if calibration, probe placement, tissue temperature, or non-concurrent measurement bias those numbers, the claim that depletion cannot cause hypoxia loses its foundation.
What would settle it
An experiment that measures tissue pO2 during a 20 Gy UHDR pulse with an independent fast technique (for instance EPR oximetry or a microelectrode) and finds a transient drop of more than 10 mmHg would contradict the paper's scaling argument. Equally, showing FLASH sparing in clamped hypoxic tissue or in hyperoxic tissue would contradict the pO2-window claim.
If this is right
- If oxygen depletion is not the mechanism, FLASH studies should pivot to measuring reactive oxygen species, scavenging, and downstream damage pathways rather than searching for transient hypoxia.
- Baseline tissue pO2 becomes a critical, often unmeasured variable: anesthesia, inspired oxygen fraction, and biological sex can move tissue into or out of the FLASH window and should be controlled and reported.
- Dose-rate characterization must include pulse structure and total irradiation time, not just mean dose rate, because radiation chemistry completes on nanosecond timescales while mean dose rate is pulse-frequency dominated.
- Direct intra-irradiation oxygen measurement is feasible and provides one of the only real-time windows into radiation chemistry in vivo, which could be extended into clinical settings.
- Tumor and normal-tissue responses may diverge under UHDR: tumor control appears hypoxia-independent, suggesting separate mechanisms for tumor killing and normal tissue sparing.
Where Pith is reading between the lines
- If gO2 is tissue-type independent but baseline dependent, the tumor-versus-normal difference in FLASH response likely comes from the local radical-scavenging pool rather than from oxygen consumption—an inference the paper poses as an open question; a testable prediction is that thiol-rich tissues show smaller effective damage but not smaller ΔpO2.
- The unresolved mismatch between tiny measured depletion and the 5–10 second oxygen-diffusion timescale of the FLASH effect suggests a microdomain effect—intracellular oxygen near DNA or mitochondria could behave differently from bulk extracellular probe readings; this would be testable with intracellular reporters like protoporphyrin IX.
- A clinical translation inference: baseline pO2 imaging could become a patient-selection biomarker for FLASH, predicting which tumors or normal tissues will show a meaningful therapeutic window; this is not stated in the paper but follows from the pO2-window claim.
Editorial analysis
A structured set of objections, weighed in public.
Referee Report
Summary. This review surveys in vivo oxygen measurement techniques and their application to FLASH radiotherapy, with emphasis on phosphorescence-based Oxyphor and PpIX probes. It compiles measurements of radiolytic oxygen consumption (gO2) and oxygen-modulation studies, and argues that measured gO2 values of 0.1–0.5 mmHg/Gy imply a 20 Gy dose depletes only 2–4 mmHg, insufficient to produce transient radiobiological hypoxia. The authors propose instead that FLASH sparing depends on an intermediate baseline pO2 window and on dose-rate-dependent ROS chemistry, modulated by tissue scavenging.
Significance. If the central negative claim holds, the oxygen-depletion hypothesis would be excluded as a general mechanism, redirecting the field toward baseline pO2 windows and ROS/scavenger chemistry. The review usefully brings together independent corroborating data (Grilj, Van Slyke, El Khatib, Koch, Montay-Gruel) alongside the authors' own Oxyphor studies, and it explicitly flags many of its own limitations—non-concurrent pO2 measurements, skin-centric data, and the unresolved split-dose timescale discrepancy. The paper is therefore a valuable synthesis, but its load-bearing scaling argument rests on gO2 values that are predominantly extracellular, cutaneous, and volume-averaged, which the authors themselves acknowledge may not capture micro-local heterogeneity.
major comments (3)
- [Section 3.4, Fig. 7] The central exclusion of oxygen depletion is a direct scaling ΔpO2 = gO2 × D, with gO2 = 0.1–0.5 mmHg/Gy. This assumes that Oxyphor-derived gO2 values represent radiolytic consumption at the radiobiological target. The paper itself states in Section 3.3 that 'it is not clear whether the lack of tissue dependence was due to the measurement process or is an inherent feature,' and Open Question 1 in Section 3.4 asks whether micro-local heterogeneity could matter. Given intercapillary distances of 100–300 μm (Section 2.2), an extracellular probe averaging over such a scale can miss perivascular gradients and localized depletion. The authors should either provide micro-local measurements or model the effect of heterogeneous oxygen distribution to support the 'no transient hypoxia' conclusion.
- [Table 3, Section 4.1] The abstract claims the FLASH effect occurs only at intermediate baseline pO2, yet most studies in Table 3 did not measure pO2 during irradiation, and Section 4.1 explicitly notes the 'absence of real-time tissue pO2 measurements during irradiation.' Hansen et al. used a separate room-air cohort and 25°C water irradiation (Section 4.2). The evidence for a sharp pO2 window is therefore correlative and non-concurrent, and the claim in the abstract is stronger than the data support.
- [Conclusions, final paragraph] The paper acknowledges an unresolved discrepancy: split-dose FLASH sparing is preserved on 5–10 s timescales that match oxygen diffusion, which is hard to reconcile with a 2–4 mmHg depletion being biologically irrelevant. Since this timescale is one of the strongest pieces of evidence for a diffusion-resupply mechanism, the authors need to either propose a quantitative explanation that reconciles the small gO2 with the time-window, or explicitly restrict their conclusion to cases where depletion is the sole mechanism.
minor comments (5)
- [Figure 2 caption] The caption contains corrupted text '9988/*///8988' that should be removed.
- [References] Reference numbering is inconsistent: reference 69 is missing from the list, and ref 117 is cited in Section 1 before many earlier-numbered references.
- [Section 2.3, Key Point 4] ΔpO2 is used without definition at first occurrence; define it in Section 2.3 or earlier in Section 3.2 where it is formally introduced.
- [Table 2] The Koch et al. values include a footnote about the >30 mmHg plateau region; clarify in the header or footnote whether these are plateau values rather than routine gO2 values.
- [Abstract and Conclusions] The abstract uses 'demonstrating that the FLASH effect occurs only at intermediate baseline pO2'; given the limitations acknowledged in the text, 'suggesting' or 'indicating' would be more proportionate.
Circularity Check
Review is a synthesis with acknowledged data gaps; central anti-depletion claim is scaled from external and multi-group gO2 measurements, not a circular derivation.
full rationale
This is a narrative review, not a derivation, so most circularity patterns (fitted-input-called-prediction, ansatz via citation, uniqueness imported from authors) do not apply. The paper's central quantitative claim is that measured gO2 values of 0.1–0.5 mmHg/Gy imply only 2–4 mmHg depletion at 20 Gy, which it concludes is insufficient to cause transient hypoxia (Section 3.4, Fig. 7). This is a linear scaling of empirically measured g-values, and the g-values come from multiple independent groups (Grilj, Van Slyke, El Khatib, Koch, Montay-Gruel, Jansen, plus the authors' own Oxyphor studies), not from a parameter fitted within this paper. The paper does rely heavily on the authors' own Oxyphor-based studies for the gO2 vs pO2 curves (Fig. 6, refs. 5, 72, 74–77, 81), and Table 3's pO2-window conclusion leans on Hunter et al. (2026, bioRxiv, self-citation) plus Tavakkoli and Sunnerberg, but corroboration by independent measurements (Grilj et al., El Khatib et al., Van Slyke et al.) means the central claim is externally supported rather than self-referential. The paper repeatedly flags its own limitations: Section 3.3 states 'It is not clear whether the lack of tissue dependence was due to the measurement process or is an inherent feature'; Section 4.1 notes the 'absence of real-time tissue pO2 measurements during irradiation' in modulation studies; Section 4.2 acknowledges Hansen et al. measured oxygen in a separate cohort not matching the irradiation setup; the Conclusions explicitly state the 5–10 s split-dose sparing timescale 'remains unresolved' against the small depletion estimate. These self-identified gaps weaken evidential strength but are not circularity: no claim is equivalent to its inputs by construction, no fitted parameter is renamed as a prediction, and no load-bearing argument reduces to an unverified self-citation. The only mild self-citation issue is the central pO2-window figure (Fig. 2b, ref. 116) and the gO2 synthesis being dominated by the authors' group; this is a normal self-citation pattern in a field where they perform the relevant measurements, and it is not the sole support. Score 1 reflects that self-citation is present but not load-bearing in a circular sense.
Axiom & Free-Parameter Ledger
free parameters (3)
- Reported gO2 values (radiolytic oxygen consumption per dose) =
0.1–0.5 mmHg/Gy; up to ~0.85 mmHg/Gy in some media
- gO2 saturation threshold pO2 =
≈15–25 mmHg
- Oxygen diffusion/resupply timescale =
5–10 s
axioms (5)
- domain assumption Oxyphor phosphorescence lifetime reports local tissue pO2 accurately during UHDR irradiation.
- domain assumption Inhaled-gas and clamping interventions change tissue pO2 without independently altering radiosensitivity.
- domain assumption In vitro radiation-chemistry measurements (hydrated electron yields, gO2 in BSA solutions) extrapolate to in vivo mechanisms.
- domain assumption Skin oxygen responses are representative of other normal tissues.
- domain assumption A Michaelis-Menten-like saturation of gO2 with pO2 is the correct functional form for ROC.
read the original abstract
FLASH radiotherapy (FLASH-RT) is the phenomenon of relative sparing of normal tissue when ultra-high dose rates (UHDR) are used compared with conventional dose rates (CDR) as clinically used. Despite extensive investigation, the underlying mechanisms remain unexplained. Among the proposed hypotheses, tissue oxygen has consistently been a central theme because oxygen is the most dominant factor known to modulate radiation-induced damage. The factors implicated in FLASH sparing include the baseline partial pressure of oxygen (pO2), transient radiolytic oxygen consumption (ROC), and oxygen-dependent changes in the chemistry of reactive oxygen species (ROS) that vary with dose rate. This review synthesizes current evidence on in vivo oxygen measurement techniques, highlighting their capabilities and limitations in capturing the spatial and temporal heterogeneity of tissue oxygenation. Key experimental studies in skin are summarized and interpreted by modulating oxygen levels via changes in inspired oxygen gas and vascular clamping interventions, demonstrating that the FLASH effect occurs only at intermediate baseline pO2 (normoxic or slightly hypoxic) values, but not at hypoxia or hyperoxia. Direct measurement of oxygen consumption during UHDR irradiation is possible, providing one of the first in situ measurements of radiation chemistry in patients. In parallel, recent advances in fast in vitro radiation chemistry assays indicate that UHDR irradiation alters radical yields, favoring increased production of solvated electrons and reduced hydroxyl radical mediated damage. Taken together, the available data suggest that the FLASH sparing effect arises from an interplay among the delivered dose and dose rate, local oxygen availability, and radiation chemistry, with tissue-specific variation in scavenging, leading to altered biological responses across the CDR-to-UHDR shift.
Figures
Reference graph
Works this paper leans on
-
[3]
Iturri L, Bertho A, Lamirault C, Brisebard E, Juchaux M, Gilbert C, et al. Oxygen supplementation in anesthesia can block FLASH effect and anti -tumor immunity in conventional proton therapy. Commun Med. 2023 Dec 15;3(1):183. doi:10.1038/s43856 - 023-00411-9
doi:10.1038/s43856 2023
-
[4]
Anesthetic Oxygen Use and Sex Are Critical Factors in the FLASH Sparing Effect
Tavakkoli AD, Clark MA, Kheirollah A, Sloop AM, Soderholm HE, Daniel NJ, et al. Anesthetic Oxygen Use and Sex Are Critical Factors in the FLASH Sparing Effect. Adv Radiat Oncol. 2024 Jun 1;9(6):101492. doi:10.1016/j.adro.2024.101492
arXiv 2024
-
[6]
Hunter DI, Sunnerberg JP, Tavakkoli AD, Sloop AM, Allen B, Gui J, et al. Inspired oxygen level and tension modulate the murine skin sparing FLASH effect in electron ultra - high dose rate irradiations [Internet]. bioRxiv; 2025 [cited 2026 Feb 22]. p. 2025.10.06.680759. Available from: https://www.biorxiv.org/content/10.1101/2025.10.06.680759v3 doi:10.1101...
-
[7]
The oxygen puzzle in FLASH radiotherapy: A comprehensive review and experimental outlook
Scarmelotto A, Delprat V, Michiels C, Lucas S, Heuskin AC. The oxygen puzzle in FLASH radiotherapy: A comprehensive review and experimental outlook. Clin Transl Radiat Oncol. 2024 Nov 1;49:100860. doi:10.1016/j.ctro.2024.100860
arXiv 2024
-
[8]
Model studies of the role of oxygen in the FLASH effect
Favaudon V, Labarbe R, Limoli CL. Model studies of the role of oxygen in the FLASH effect. Med Phys [Internet]. 2021. Available from: https://www.ncbi.nlm.nih.gov/pubmed/34407219 doi:10.1002/mp.15129
arXiv 2021
-
[9]
Reinventing Radiobiology in the Light of FLASH Radiotherapy
Limoli CL, Vozenin MC. Reinventing Radiobiology in the Light of FLASH Radiotherapy. Annu Rev Cancer Biol. 2023 Apr;7:1–21. doi:10.1146/annurev -cancerbio-061421-022217 PubMed PMID: 39421564; PubMed Central PMCID: PMC11486513
doi:10.1146/annurev 2023
-
[10]
Hirayama R, Ito A, Noguchi M, Matsumoto Y, Uzawa A, Kobashi G, et al. OH radicals from the indirect actions of X-rays induce cell lethality and mediate the majority of the oxygen enhancement effect. Radiat Res. 2013. p. 514–23. doi:10.1667/RR13368.1
-
[11]
Oxidative stress, free radicals and protein peroxides
Gebicki JM. Oxidative stress, free radicals and protein peroxides. Arch Biochem Biophys. 2016 Apr 1;595:33 –9. doi:10.1016/j.abb.2015.10.021 PubMed PMID: 27095212
-
[12]
Pulse Radiolysis Using Very-high -energy Ions for Optimizing Cancer Therapy
Getoff N. Pulse Radiolysis Using Very-high -energy Ions for Optimizing Cancer Therapy. Vivo Athens Greece. 2016 Apr;30(2):119–21. PubMed PMID: 26912822
2016
-
[13]
Oxygen as a primary species in radiolysis of water
Daniels M, Wigg E. Oxygen as a primary species in radiolysis of water. Science. 1966 Sep 23;153(3743):1533–4. doi:10.1126/science.153.3743.1533 PubMed PMID: 5331263
arXiv 1966
-
[14]
Chappuis F, Tran HN, Jorge PG, Zein SA, Kyriakou I, Emfietzoglou D, et al. Investigating ultra-high dose rate water radiolysis using the Geant4 -DNA toolkit and a Geant4 model of the Oriatron eRT6 electron linac. Sci Rep. 2024 Nov 4;14(1):26707. doi:10.1038/s41598-024-76769-0 PubMed PMID: 39496703; PubMed Central PMCID: PMC11535405
-
[15]
Historical view of the effects of radiation on cancer cells
Hazout S, Oehler C, Zwahlen DR, Taussky D. Historical view of the effects of radiation on cancer cells. Oncol Rev. 2025;19:1527742. doi:10.3389/or.2025.1527742 PubMed PMID: 40370490; PubMed Central PMCID: PMC12075557
arXiv 2025
-
[16]
Sunada S, Hirakawa H, Fujimori A, Uesaka M, Okayasu R. Oxygen Enhancement Ratio in Radiation-Induced Initial DSBs by an Optimized Flow Cytometry-based Gamma-H2AX Analysis in A549 Human Cancer Cells. Radiat Res. 2017. p. 591–4. doi:10.1667/RR14824.1
-
[17]
Radiobiological Aspects of FLASH Radiotherapy
Hageman E, Che PP, Dahele M, Slotman BJ, Sminia P. Radiobiological Aspects of FLASH Radiotherapy. Biomolecules. 2022 Oct;12(10):1376. doi:10.3390/biom12101376
-
[18]
DNA strand break induction of aqueous plasmid DNA exposed to 30 MeV protons at ultra-high dose rate
Ohsawa D, Hiroyama Y, Kobayashi A, Kusumoto T, Kitamura H, Hojo S, et al. DNA strand break induction of aqueous plasmid DNA exposed to 30 MeV protons at ultra-high dose rate. J Radiat Res. 2022. p. 255–60. doi:10.1093/jrr/rrab114
-
[19]
Biological effects of conventional and ultra high dose rate radiation in human cells
Ryan TL, Escalona MB, O’Brien K, Tan Y, Kanagaraj K, Taveras M, et al. Biological effects of conventional and ultra high dose rate radiation in human cells. Sci Rep. 2026 Jan 2;16(1):3713. doi:10.1038/s41598-025-33817-7 PubMed PMID: 41484408; PubMed Centra l PMCID: PMC12852176
-
[20]
Thomas WS, Kulkarni S, Ilina A, Reed M, Pogue BW. In vitro oxygen concentration alters reactive oxygen species yields with minor plasmid DNA damage change at ultra-high - dose rate. Br J Radiol. 2026 Jan 1;99(1177):169 –75. doi:10.1093/bjr/tqaf245 PubMed PMID: 41003674
-
[21]
Molecular biological mechanisms of radiotherapy-induced skin injury occurrence and treatment
Cui J, Wang TJ, Zhang YX, She LZ, Zhao YC. Molecular biological mechanisms of radiotherapy-induced skin injury occurrence and treatment. Biomed Pharmacother. 2024 Nov 1;180:117470. doi:10.1016/j.biopha.2024.117470
arXiv 2024
-
[23]
Demidov V, Zhao X, Demidova O, Pang HYM, Flueraru C, Liu FF, et al. Preclinical quantitative in vivo assessment of skin tissue vascularity in radiation-induced fibrosis with optical coherence tomography. J Biomed Opt. 2018 Oct;23(10):1–9. doi:10.1117/1.JBO.23.10.106003 PubMed PMID: 30315644
-
[24]
Relevance of oxygen in radiation oncology
Molls M, Stadler P, Becker A, Feldmann HJ, Dunst J. Relevance of oxygen in radiation oncology. Mechanisms of action, correlation to low hemoglobin levels. Strahlenther Onkol Organ Dtsch Rontgengesellschaft Al. 1998 Dec;174 Suppl 4:13–6. PubMed PMID: 9879341
1998
-
[25]
Cooper CR, Jones D, Jones GD, Petersson K. FLASH irradiation induces lower levels of DNA damage ex vivo, an effect modulated by oxygen tension, dose, and dose rate. Br J Radiol. 2022. p. 20211150. doi:10.1259/bjr.20211150
-
[26]
Effects of the Oxygen depletion in FLASH irradiation investigated through Geant4 -DNA toolkit
Farokhi F, Shirani B, Fattori S, Ali Asgarian M, Cuttone G, Jia SB, et al. Effects of the Oxygen depletion in FLASH irradiation investigated through Geant4 -DNA toolkit. Radiat Phys Chem. 2023 Nov 1;212:111184. doi:10.1016/j.radphyschem.2023.111184
arXiv 2023
-
[27]
Van den Heuvel F, Vella A, Fiorini F, Brooke M, Hill MA, Maughan T. Incorporating oxygenation levels in analytical DNA-damage models -quantifying the oxygen fixation mechanism. Phys Med Biol [Internet]. 2021. Available from: https://www.ncbi.nlm.nih.gov/pubmed/34130265 doi:10.1088/1361 -6560/ac0b80
arXiv 2021
-
[28]
The oxygen fixation hypothesis: a reevaluation
Ewing D. The oxygen fixation hypothesis: a reevaluation. Am J Clin Oncol. 1998. p. 355 –
1998
-
[29]
A mechanistic investigation of the oxygen fixation hypothesis and oxygen enhancement ratio
Grimes DR, Partridge M. A mechanistic investigation of the oxygen fixation hypothesis and oxygen enhancement ratio. Biomed Phys Eng Express. 2015. p. 045209. doi:10.1088/2057-1976/1/4/045209
-
[30]
Impact of oxygen deprivation on the FLASH effect for skin toxicity in a murine model
Hansen AH, Poulsen PR, Kristensen L, Møller VV, Sinha PM, Johansen JG, et al. Impact of oxygen deprivation on the FLASH effect for skin toxicity in a murine model. Radiother Oncol J Eur Soc Ther Radiol Oncol. 2026 Jan;214:111277. doi:10.1016/j.radonc.2025.111277 PubMed PMID: 41232779
arXiv 2026
-
[31]
Investigating the oxygen dependence of FLASH -RT using electron paramagnetic resonance imaging
Waldrop T, Murley G, Velasquez B, Konradsson E, Neill D, Connell L, et al. Investigating the oxygen dependence of FLASH -RT using electron paramagnetic resonance imaging. Radiother Oncol. 2026 Feb 1;215. doi:10.1016/j.radonc.2025.111329 PubMed PMID: 41389858
arXiv 2026
-
[32]
Prognostic value of tumor oxygenation in 397 head and neck tumors after primary radiation therapy
Nordsmark M, Bentzen SM, Rudat V, Brizel D, Lartigau E, Stadler P, et al. Prognostic value of tumor oxygenation in 397 head and neck tumors after primary radiation therapy. An international multi-center study. Radiotherapy and oncology. 2005. p. 18 –24
2005
-
[33]
Nozue M, Lee I, Yuan F, Teicher BA, Brizel DM, Dewhirst MW, et al. Interlaboratory variation in oxygen tension measurement by Eppendorf “Histograph” and comparison with hypoxic marker. J Surg Oncol. 1997. p. 30–8. doi:10.1002/(SICI)1096- 9098(199709)66:1<30::AID-JSO7>3.0.CO;2-O [pii]
-
[34]
Measuring Tumor Hypoxia
Raleigh JA, Dewhirst MW, Thrall DE. Measuring Tumor Hypoxia. Semin Radiat Oncol
-
[35]
Oxygen transport in tumors: characteristics and clinical implications
Vaupel P. Oxygen transport in tumors: characteristics and clinical implications. Adv Exp Med Biol. 1996. p. 341 –51
1996
-
[36]
Swartz HM, Flood AB, Schaner PE, Halpern H, Williams BB, Pogue BW, et al. How best to interpret measures of levels of oxygen in tissues to make them effective clinical tools for care of patients with cancer and other oxygen -dependent pathologies. Physiol R ep. 2020. p. e14541. doi:10.14814/phy2.14541
-
[37]
Measuring real levels of oxygen in vivo: opportunities and challenges
Swartz HM. Measuring real levels of oxygen in vivo: opportunities and challenges. Biochem Soc Trans. 2002. p. 248 –52. doi:10.1042/
2002
-
[38]
Oxygenation of human tumors
Vaupel P. Oxygenation of human tumors. Strahlenther Onkol. 1990. p. 377–86
1990
-
[39]
Concepts of oxygen transport at the microcirculatory level
Dewhirst MW. Concepts of oxygen transport at the microcirculatory level. Semin Radiat Oncol. 1998 Jul;8(3):143–50. doi:10.1016/s1053-4296(98)80040 -4 PubMed PMID: 9634491
-
[41]
Pulse oximetry: analysis of theory, technology, and practice
Wukitsch MW, Petterson MT, Tobler DR, Pologe JA. Pulse oximetry: analysis of theory, technology, and practice. J Clin Monit. 1988. p. 290–301
1988
-
[42]
Measurement of hemoglobin flow and blood flow by near -infrared spectroscopy
Edwards AD, Richardson C, van der Zee P, Elwell C, Wyatt JS, Cope M, et al. Measurement of hemoglobin flow and blood flow by near -infrared spectroscopy. J Appl Physiol. 1993. p. 1884 –9
1993
-
[43]
Oxygen distributions within R3230Ac tumors growing in dorsal flap window chambers in rats
Wilson DF, Evans SM, Jenkins WT, Vinogradov SA, Ong E, Dewhirst MW. Oxygen distributions within R3230Ac tumors growing in dorsal flap window chambers in rats. Adv Exp Med Biol. 1998. p. 603–9
1998
-
[44]
Blood flow, oxygen and nutrient supply, and metabolic microenvironment of human tumors: a review
Vaupel P, Kallinowski F, Okunieff P. Blood flow, oxygen and nutrient supply, and metabolic microenvironment of human tumors: a review. Cancer Res. 1989. p. 6449 –65
1989
-
[45]
Pogue BW, Paulsen KD, O’Hara JA, Wilmot CM, Swartz HM. Estimation of oxygen distribution in RIF-1 tumors by diffusion model -based interpretation of pimonidazole hypoxia and eppendorf measurements. Radiat Res. 2001 Jan;155(1 Pt 1):15 –25. doi:10.1667/0033- 7587(2001)155[0015:eoodir]2.0.co;2 PubMed PMID: 11121211
-
[46]
Effect of hyperoxia, hypoxia and hypercapnia on O2 supply of malignant tumors in situ
Vaupel P. Effect of hyperoxia, hypoxia and hypercapnia on O2 supply of malignant tumors in situ. Bibl Anat. 1977. p. 288– 90
1977
-
[47]
Blood flow and oxygenation status of head and neck carcinomas
Vaupel P. Blood flow and oxygenation status of head and neck carcinomas. Adv Exp Med Biol. 1997. p. 89–95
1997
-
[48]
An integrated approach to measuring tumor oxygen status using human melanoma xenografts as a model
Menon C, Polin GM, Prabakaran I, Hsi A, Cheung C, Culver JP, et al. An integrated approach to measuring tumor oxygen status using human melanoma xenografts as a model. Cancer Res. 2003. p. 7232–40
2003
-
[49]
Review of methods used to study oxygen transport at the microcirculatory level
Dewhirst MW, Klitzman B, Braun RD, Brizel DM, Haroon ZA, Secomb TW. Review of methods used to study oxygen transport at the microcirculatory level. Int J Cancer. 2000 Oct 20;90(5):237–55. PubMed PMID: 11091348
2000
-
[50]
Indwelling blood compatible chemical sensors
Eberhart RC. Indwelling blood compatible chemical sensors. Surg Clin North Am. 1985 Aug;65(4):1025–40. doi:10.1016/s0039 -6109(16)43692-3 PubMed PMID: 3901342
-
[51]
Noninvasive imaging of the distribution in oxygen in tissue in vivo using near-infrared phosphors
Vinogradov SA, Lo LW, Jenkins WT, Evans SM, Koch C, Wilson DF. Noninvasive imaging of the distribution in oxygen in tissue in vivo using near-infrared phosphors. Biophys J. 1996. p. 1609–17. doi:S0006-3495(96)79764-3 [pii] 10.1016/S0006 -3495(96)79764-3
-
[52]
Measuring oxygen in living tissue: intravascular, interstitial, and “tissue” oxygen measurements
Wilson DF, Finikova OS, Lebedev AY, Apreleva S, Pastuszko A, Lee WM, et al. Measuring oxygen in living tissue: intravascular, interstitial, and “tissue” oxygen measurements. Adv Exp Med Biol. 2011. p. 53–9. doi:10.1007/978 -1-4419-7756-4_8
doi:10.1007/978 2011
-
[53]
In vivo EPR: when, how and why? NMR Biomed
Gallez B, Swartz HM. In vivo EPR: when, how and why? NMR Biomed. 2004. p. 223 –5. doi:10.1002/nbm.913
-
[54]
A method for measuring oxygen distributions in tissue using frequency domain phosphorometry
Vinogradov SA, Fernandez-Seara MA, Dupan BW, Wilson DF. A method for measuring oxygen distributions in tissue using frequency domain phosphorometry. Comp Biochem Physiol A Mol Integr Physiol. 2002. p. 147 –52. doi:S1095643301005414 [pii]
2002
-
[55]
Oxygen distributions in tissue measured by phosphorescence quenching
Vinogradov SA, Grosul P, Rozhkov V, Dunphy I, Shuman L, Dugan BW, et al. Oxygen distributions in tissue measured by phosphorescence quenching. Adv Exp Med Biol. 2003. p. 181–5
2003
-
[56]
Development of Ruthenium Nanophotocages with Red or Near -Infrared Light -Responsiveness
Zhang Z, He M, Wang R, Fan J, Peng X, Sun W. Development of Ruthenium Nanophotocages with Red or Near -Infrared Light -Responsiveness. Chembiochem Eur J Chem Biol. 2023 Dec 14;24(24):e202300606. doi:10.1002/cbic.202300606 PubMed PMID: 37837285
-
[57]
Chu F, Yang J, Cai H, Qu R, Fang Z. Characterization of a dissolved oxygen sensor made of plastic optical fiber coated with ruthenium-incorporated solgel. Appl Opt. 2009 Jan 10;48(2):338–42. doi:10.1364/ao.48.000338 PubMed PMID: 19137045
-
[58]
Xiang HJ, An L, Tang WW, Yang SP, Liu JG. Photo-controlled targeted intracellular delivery of both nitric oxide and singlet oxygen using a fluorescence -trackable ruthenium nitrosyl functional nanoplatform. Chem Commun (Camb). 2015. p. 2555–8. doi:10.1039/c4cc09869b
-
[59]
Mitochondrial PO2 measured by delayed fluorescence of endogenous protoporphyrin IX
Mik EG, Stap J, Sinaasappel M, Beek JF, Aten JA, van Leeuwen TG, et al. Mitochondrial PO2 measured by delayed fluorescence of endogenous protoporphyrin IX. Nat Methods
-
[60]
Petusseau AF, Ochoa M, Reed M, Doyley MM, Hasan T, Bruza P, et al. Pressure- enhanced sensing of tissue oxygenation via endogenous porphyrin: Implications for dynamic visualization of cancer in surgery. Proc Natl Acad Sci U S A. 2024 Aug 20;121(34):e2405628121. doi:10.1073/pnas.2405628121 PubMed PMID: 39141355; PubMed Central PMCID: PMC11348300
-
[61]
doi:10.1097/00000421-199808000-00008
-
[62]
Oxyphor R2 and G2: phosphors for measuring oxygen by oxygen-dependent quenching of phosphorescence
Dunphy I, Vinogradov SA, Wilson DF. Oxyphor R2 and G2: phosphors for measuring oxygen by oxygen-dependent quenching of phosphorescence. Anal Biochem. 2002 Nov 15;310(2):191–8. doi:10.1016/s0003 -2697(02)00384-6 PubMed PMID: 12423638
-
[63]
Esipova TV, Karagodov A, Miller J, Wilson DF, Busch TM, Vinogradov SA. Two new “protected” oxyphors for biological oximetry: properties and application in tumor imaging. Anal Chem. 2011. p. 8756–65. doi:10.1021/ac2022234
-
[65]
In vivo mitochondrial oxygen tension measured by a delayed fluorescence lifetime technique
Mik EG, Johannes T, Zuurbier CJ, Heinen A, Houben -Weerts JH, Balestra GM, et al. In vivo mitochondrial oxygen tension measured by a delayed fluorescence lifetime technique. Biophys J. 2008. p. 3977–90. doi:10.1529/biophysj.107.126094 S0006 -3495(08)78536-9 [pii]
-
[66]
Mik EG. Special article: measuring mitochondrial oxygen tension: from basic principles to application in humans. Anesth Analg. 2013. p. 834 –46. doi:10.1213/ANE.0b013e31828f29da ANE.0b013e31828f29da [pii]
-
[67]
Oxyphor 2P: A High-Performance Probe for Deep -Tissue Longitudinal Oxygen Imaging
Esipova TV, Barrett MJP, Erlebach E, Masunov AE, Weber B, Vinogradov SA. Oxyphor 2P: A High-Performance Probe for Deep -Tissue Longitudinal Oxygen Imaging. Cell Metab
-
[68]
Does FLASH deplete oxygen? Experimental evaluation for photons, protons, and carbon ions
Jansen J, Knoll J, Beyreuther E, Pawelke J, Skuza R, Hanley R, et al. Does FLASH deplete oxygen? Experimental evaluation for photons, protons, and carbon ions. Med Phys
-
[70]
Van Slyke AL, El Khatib M, Velalopoulou A, Diffenderfer E, Shoniyozov K, Kim MM, et al. Oxygen Monitoring in Model Solutions and In vivo in Mice During Proton Irradiation at Conventional and FLASH Dose Rates. Radiat Res. 2022. p. 181 –9. doi:10.1667/RADE-21- 00232.1
-
[71]
Scholz M, Cao X, Gunn JR, Bruza P, Pogue B. pO(2)-weighted imaging in vivo by delayed fluorescence of intracellular protoporphyrin IX: publisher’s note. Opt Lett. 2020 Feb 1;45(3):664. doi:10.1364/OL.387641 PubMed PMID: 32004278
-
[73]
Quantification of Oxygen Depletion During FLASH Irradiation In vitro and In vivo
Cao X, Zhang R, Esipova TV, Allu SR, Ashraf R, Rahman M, et al. Quantification of Oxygen Depletion During FLASH Irradiation In vitro and In vivo. Int J Radiat Oncol. 2021 Sep 1;111(1):240–8. doi:10.1016/j.ijrobp.2021.03.056
-
[74]
Van Slyke AL, El Khatib M, Velalopoulou A, Diffenderfer E, Shoniyozov K, Kim MM, et al. Oxygen Monitoring in Model Solutions and In vivo in Mice During Proton Irradiation at Conventional and FLASH Dose Rates. Radiat Res. 2022 Aug 1;198(2):181 –9. doi:10.1667/RADE-21-00232.1 PubMed PMID: 35640166; PubMed Central PMCID: PMC10176203
-
[75]
Ultrafast Tracking of Oxygen Dynamics During Proton FLASH
El Khatib M, Van Slyke AL, Velalopoulou A, Kim MM, Shoniyozov K, Allu SR, et al. Ultrafast Tracking of Oxygen Dynamics During Proton FLASH. Int J Radiat Oncol. 2022 Jul 1;113(3):624–34. doi:10.1016/j.ijrobp.2022.03.016
-
[78]
Analysis of the heterogeneity of pO2 dynamics during photodynamic therapy with verteporfin
Pogue BW, Braun RD, Lanzen JL, Erickson C, Dewhirst MW. Analysis of the heterogeneity of pO2 dynamics during photodynamic therapy with verteporfin. Photochem Photobiol. 2001 Nov;74(5):700–6. doi:10.1562/0031 -8655(2001)074<0700:aothop>2.0.co;2 PubMed PMID: 11723798
-
[79]
Does FLASH deplete oxygen? Experimental evaluation for photons, protons, and carbon ions
Jansen J, Knoll J, Beyreuther E, Pawelke J, Skuza R, Hanley R, et al. Does FLASH deplete oxygen? Experimental evaluation for photons, protons, and carbon ions. Med Phys. 2021;48(7):3982–90. doi:10.1002/mp.14917
-
[80]
Koch CJ, Kim MM, Wiersma RD. Radiation-Chemical Oxygen Depletion Depends on Chemical Environment and Dose Rate: Implications for the FLASH Effect. Int J Radiat Oncol. 2023 Sep 1;117(1):214–22. doi:10.1016/j.ijrobp.2023.04.001
-
[81]
Petusseau AF, Clark M, Bruza P, Gladstone D, Pogue BW. Intracellular Oxygen Transient Quantification in vivo During Ultra-High Dose Rate FLASH Radiation Therapy. Int J Radiat Oncol. 2024 Nov 1;120(3):884 –93. doi:10.1016/j.ijrobp.2024.04.068
-
[82]
Ultra-High Dose Rate (FLASH) Radiotherapy: Silver Bullet or Fool’s Gold? Front Oncol
Wilson JD, Hammond EM, Higgins GS, Petersson K. Ultra-High Dose Rate (FLASH) Radiotherapy: Silver Bullet or Fool’s Gold? Front Oncol. 2020 Jan 17;9. doi:10.3389/fonc.2019.01563
arXiv 2020
-
[83]
Ling CC. Time Scale of Radiation-Induced Oxygen Depletion and Decay Kinetics of Oxygen-Dependent Damage in Cells Irradiated at Ultrahigh Dose Rates. Radiat Res. 1975 Sep;63(3):455. doi:10.2307/3574097
-
[84]
Kinetic model of radiochemical oxygen depletion (ROD) in FLASH radiotherapy
Seco J, Freitas H. Kinetic model of radiochemical oxygen depletion (ROD) in FLASH radiotherapy. Med Phys. 2026 Apr;53(4):e70383. doi:10.1002/mp.70383
-
[85]
Long - term neurocognitive benefits of FLASH radiotherapy driven by reduced reactive oxygen species
Montay-Gruel P, Acharya MM, Petersson K, Alikhani L, Yakkala C, Allen BD, et al. Long - term neurocognitive benefits of FLASH radiotherapy driven by reduced reactive oxygen species. Proc Natl Acad Sci U S A. 2019 May 28;116(22):10943– 51. doi:10.1073/pnas.1901777116 PubMed PMID: 31097580; PubMed Central PMCID: PMC6561167
-
[86]
Deep-tissue oxygen monitoring in the brain of rabbits for stroke research
Khan N, Hou H, Eskey CJ, Moodie K, Gohain S, Du G, et al. Deep-tissue oxygen monitoring in the brain of rabbits for stroke research. Stroke. 2015 Mar;46(3):e62 -66. doi:10.1161/STROKEAHA.114.007324 PubMed PMID: 25613304; PubMed Central PMCID: PMC4342327
-
[87]
Favaudon V, Caplier L, Monceau V, Pouzoulet F, Sayarath M, Fouillade C, et al. Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice. Sci Transl Med. 2014 Jul 16;6(245):245ra93. doi:10.1126/scitranslmed.3008973 PubMed PMID: 25031268
-
[88]
Design, Implementation, and in vivo Validation of a Novel Proton FLASH Radiation Therapy System
Diffenderfer ES, Verginadis II, Kim MM, Shoniyozov K, Velalopoulou A, Goia D, et al. Design, Implementation, and in vivo Validation of a Novel Proton FLASH Radiation Therapy System. Int J Radiat Oncol Biol Phys. 2020 Feb 1;106(2):440– 8. doi:10.1016/j.ijrobp.2019.10.049 PubMed PMID: 31928642; PubMed Central PMCID: PMC7325740
-
[89]
Hypofractionated FLASH-RT as an Effective Treatment against Glioblastoma that Reduces Neurocognitive Side Effects in Mice - PubMed [Internet]. [cited 2026 Jan 23]. Available from: https://pubmed.ncbi.nlm.nih.gov/33060122/
arXiv 2026
-
[90]
Levy K, Natarajan S, Wang J, Chow S, Eggold JT, Loo PE, et al. Abdominal FLASH irradiation reduces radiation -induced gastrointestinal toxicity for the treatment of ovarian cancer in mice. Sci Rep. 2020 Dec 10;10(1):21600. doi:10.1038/s41598 -020-78017-7 PubMed PMID: 33303827; PubMed Central PMCID: PMC7728763
doi:10.1038/s41598 2020
discussion (0)
Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.