Development and Experimental Characterization of a Compact 228Th–D2O Photoneutron Source for Low-Energy Nuclear Recoil Calibration
The paper says
“Calibrating the few-keV nuclear-recoil response of liquid xenon detectors calls for a low-energy neutron source. We have developed and characterized a 228Th–D2O photoneutron source that converts the 2.615 MeV γ rays from 208Tl into sub-MeV neutrons through the D(γ,n)p reaction.”
more of the abstract
After optimizing the D2O geometry to balance neutron production against escape efficiency, we validated the chosen configuration experimentally. The forward emission rate is 1734 ± 76 (stat.) ± 65 (sys.) h−1, and the spectrum unfolded from Bonner sphere data is concentrated at 0.17–0.22 MeV. In a 220 h exposure of PandaX-20T, the same configuration should provide about 1000 selected nuclear-recoil events. These results support the use of a long-lived 228Th–D2O source for low-energy neutron calibration of large liquid xenon detectors.
from the abstract
Pith says
Machine review · deepseek-v4-flash · 13 Nov 2026 · what was read was not recorded · not checked by a person
It can be wrong. Challenge this review or reply to it.
Holds up, with conditions. Solid engineering paper with a central result that depends on unvalidated simulated efficiency; worth publishing after a clearer discussion of the absolute calibration caveat. This line is wrong
3 major objections · 5 minor
How often objections from reviews like this one are right: not yet measured
Not checked: figures, code, what the 47 references say.
The paper says, at Section 3, Eq. (4)
The review did not quote this passage.
A classifier read this sentence of the paper as acknowledging the concern (probability 0.65):
“A final validation based on the complete PandaX-20T detector simulation or data is still needed. 6.”
Pith says: Objection 1 of 3 · major
The forward neutron emission rate (1734 ± 76 stat. ± 65 sys. h−1) rests entirely on the Geant4-simulated absolute detection efficiency of 7.63% for the moderated 3He counter. The systematic uncertainty (3.76%) is derived from simulation parameter variations, but the absolute efficiency has not been validated against an independent calibrated neutron source.
more of objection 1
The relative position dependence is validated (Fig. 5), but this does not verify the absolute scale. The authors state that final validation with PandaX-20T data is needed, but this limitation should be stated more prominently in the abstract and conclusions, and the potential impact of an uncalibrated absolute efficiency on the emission rate should be discussed quantitatively.
The paper says, at Section 4, Fig. 7
The review did not quote this passage.
A classifier read this sentence of the paper as acknowledging the concern (probability 0.93):
“A quantitative assessment of the statistical significance of any discrepancy is not attempted here because the figure does not show an uncertainty band for the unfolded spectrum.”
Pith says: Objection 2 of 3 · major
The unfolded neutron spectrum is shown without any uncertainty band. The authors acknowledge in the text that 'A quantitative assessment of the statistical significance of any discrepancy is not attempted here because the figure does not show an uncertainty band for the unfolded spectrum' (sentence [b213]).
more of objection 2
For a paper that uses this spectrum as input to the PandaX-20T calibration simulation, the absence of spectral uncertainties is a significant omission. The authors should provide an uncertainty band (e.g., using the covariance matrix from the MAXED unfolding) or explain how the shape uncertainty affects the simulation results. Without this, the reliability of the shape-dependent predictions (e.g., the 4.60 events h−1 NR rate) is unclear.
The paper says, at Section 5, final paragraph
The review did not quote this passage.
A classifier read this sentence of the paper as acknowledging the concern (probability 0.94):
“This coincidence-induced ER contamination should be quantified in a future event-level simulation before claiming a specific NR sample purity.”
Pith says: Objection 3 of 3 · major
The simulation reports neutron-induced NR and gamma-induced ER rates separately but does not evaluate the probability that a 2.615 MeV gamma ray deposits energy in the same event window as a selected NR. The authors correctly note that 'This coincidence-induced ER contamination should be quantified in a future event-level simulation before claiming a specific NR sample purity' (sentence [b280]).
more of objection 3
This omission directly affects the claim that 'about 1000 selected nuclear-recoil events' can be obtained. The paper should either provide a rough estimate of the coincidence contamination or explicitly state that the 1000-event figure refers to raw NR candidates before purity selection. The current presentation could mislead readers.
5 minor objections
The paper says, at Section 2.1, Fig. 2 caption
Pith says: minor objection 1 of 5
The caption states 'Because the simulated cross sections are modified with respect to the physical ones, only the relative angular shape is meaningful.' This is a crucial statement that should also appear in the main text, not only in the figure caption. Additionally, the reader needs to know what modifications were made and why, or at least a reference for the modified cross sections.
The paper says, at Section2.2, Fig. 3
Pith says: minor objection 2 of 5
The schematic in Fig. 3 has components 1,2,3,4,5,6,7,8 but the caption lists them as 1, Fused Silica Bottle; 2, Actuating Rod; etc. The numbering in the figure should match the caption list exactly (currently the caption starts with '1. Fused Silica Bottel' – typo 'Fused' is misspelled).
show 3 more
The paper says, at Section 4, Fig. 7 axes
Pith says: minor objection 3 of 5
In Fig. 7, the vertical axis label reads '/;dE φ E d A)'. This appears to be a formatting artifact from the PDF conversion. It should be clearly given as 'E dφ/dE (arb. units)' or similar. Also, the inset panel (B) should have the same axis label or explain th normalization.
The paper says, at Section 5, first paragraph
Pith says: minor objection 4 of 5
The reference to BambooMC [42] is appropriate, but the paper does not state whether the simulation uses the same Geant4 version and physics list as the source optimization. A brief statement of consistency would be helpful.
The paper says, at General
Pith says: minor objection 5 of 5
The paper uses the phrase 'forward end face' repeatedly without defining a coordinate system. A coordinate axis or a sketch showing the forward direction relative to the detector would improve clarity.
What would settle it
Calibrate the entire detection setup (³He counter + PE moderator) using a well-characterised neutron source with known absolute emission rate, e.g., a traceable ²⁵²Cf source or an ²⁴¹Am-Be source with certified output, placed at the same position and geometry as the ²²⁸Th-D₂O source. Compare the measured net count rate with the Geant4-predicted rate to either confirm the 7.63% efficiency within ≤3.76% or quantify a systematic offset.
from the skeptic's reading
How the review went
4 readings by deepseek-v4-flash · 13 Nov 2026
- Reader holds up, with conditions
- Skeptic agrees
- Editor wrote the verdict above
- Referee minor revision: 3 major objections, 5 minor
Not run on this paper: the simulated authors' defence.
Pith did not check this review's objections against the paper.
The review in full
The review in brief, machine-written
This paper reports the first dedicated experimental characterization of a 228Th–D2O photoneutron source for low-energy nuclear recoil calibration in LXe detectors. The core ountputs—a forward emisssion rate of 1734 h⁻¹ and a neutron spectrum peakd at 0.17–0.22 MeV—come from careful measurements with proper D2O/H2O backgound subtraction and Bonner sphere unfolding. The geometry optimization and comparison to simulation are weell done. The systematic uncertaintiy breakdown (Table 1) is transparent and reasonble for a simulation-based efficiency. The PandaX‑20T Monte Caro study demonstrates the source's fesibility for calibration campaignnes.
The load-bearing claim is the forward emission rate, which rests entirely on the Geant4‑simulated absolute detection efficiency of 7.63% for the moderated ³He counter (Section 3). The pap er acknowledges the need for final validation with PandaX‑20T data (last sentence of Section 5), but the absolute rate itself is not cross‑checked against any independent calibrated neutron source. The systematic uncertainty (3.76%) is derived from parameter variations inside the same MC framework—there is no independent absolute measurement to validate that uncertainty envelope. This is the pap er's softest spot. The paper also notes that cooincidence γ‑induced ER contamination is not yet evaluted in the PandaX simuulation (Sect. 5), and that the angular distribution in Fig. 2 is onlly relative because the cross sections were modifed (Sect. 2.1). Those are acknowledged limitations, not flaws.
Overall, this is a competent, honest piece of instrument development. The measurments are sound but the absolute nor malization rests on simulation in a way that the authors don't overstate, but reviewers should push for a clearer discussion of how that efficiency could be validated—or at least an explcitit statement that the rate is model‑dependent until cross‑checked. The paper deserves a serious peer review: it provide s a useful engineering tool for the LXe community, and the data are worth publishing if the caveat on absolute efficiency is properly flagged.
In summry: yes to peer review, but the referee should ask the authors to either perform an independent efficiency cross‑check with a known neutron source (if possible) or strengthen the text to make the simulation‑dependence of the central result unmistakable.
The strongest objection
The central claim of a forward neutron emission rate of 1734 ± 76 ± 65 h⁻¹ (Eq. 4) rests on the Geant4-simulated absolute detection efficiency of 7.63% for the moderated ³He counter (Section 3). While the relative position dependence is validated (Fig.5), the absolute efficiency is not cross-checked against any calibrated neutron source. The systematic uncertainty (Table 1) arises from parameter variations within simulation, most notably a 3.27% contribution from a conservatively assumed 1 cm air gap in D₂O. However, the simulation of neutron transport through D₂O, polyethylene moderator, and capture in ³He may harbor unrecognised systematic errors—e.g., in the nuclear data (cross sections, thermalisation) or geometry details—that are not captured by the limited parameter scan. An independent absolute measurement would both reduce and validate this uncertainty. The paper acknowledges the need for final validation with PandaX-20T data, but the absolute rate itself is fundamental to the claim of sufficiency for calibration campaigns.
The reader's weakest_assumption correctly identifies that the absolute ³He detection efficiency is not independently validated. This is indeed the most load-bearing concern because the absolute forward emission rate is a core numerical claim (1734 h⁻¹) and directly determines the utility of the source for calibration campaigns. The paper provides a thorough systematic uncertainty evaluation but omits an external validation step, which would be standard practice for such measurements. The reader's CONDITIONAL verdict is therefore appropriate, and our stress test does not change it. The condition for future work could specifically include an absolute efficiency calibration, but the reader's existing condition on coincidence gamma contamination is also valid.
The referee's report
If the measurements are reliable, the 228Th–D2O source fills a practical need: a long-lived, compact photoneutron source with a clean low-energy spectrum for calibrating the few-keV NR response of large liquid xenon detectors. This is valuable for dark matter and CEνNS experiments where accurate low-energy response characterization is critical. The work includes careful D2O/H2O background subtraction, systematic uncertainty quantification, and comparison with Geant4 simulations. The potential dual-use for high-energy gamma calibration (2.615 MeV line close to the 0νββ Q-value) adds further interest. However, the reliance on a single simulated absolute detection efficiency (7.63%) without independent absolute calibration and the absence of an uncertainty band on the unfolded spectrum limit the strength of the central claims.
Pith's reading of the paper (a machine's summary, not the authors')
This paper develops and characterizes a compact photoneutron source using the 2.615 MeV gamma ray from ²²⁸Th decay to drive the D(γ,n)p reaction in heavy water. The authors optimize the geometry so that the source is placed 1 cm from the forward face of the D₂O volume, maximizing forward neutron emission. They measure the emission rate with a ³He proportional counter and the energy spectrum with a Bonner sphere spectrometer, obtaining a forward rate of 1734 ± 91 h⁻¹ and a spectrum concentrated at 0.17–0.22 MeV, entirely below 1 MeV. Simulating deployment in the PandaX‑20T detector, they predict about 1000 selected nuclear‑recoil events in a 220‑h exposure. If correct, this source offers a longer‑lived alternative to ⁸⁸Y‑Be sources for low‑energy calibration of dark‑matter and neutrino detectors, with the added benefit of a simultaneous 2.615 MeV gamma‑ray line for high‑energy calibration.
What carries the argument. The D(γ,n)p photodisintegration reaction, driven by the 2.615 MeV gamma ray from ²⁰⁸Tl in the ²²⁸Th decay chain. This two‑body reaction’s kinematics, with only 0.391 MeV above the deuteron binding energy, naturally restricts neutron energies to the sub‑MeV range. The measured spectrum is shaped by the reaction kinematics, elastic scattering on deuterium and oxygen in D₂O, and geometry‑dependent escape probabilities, as confirmed by Geant4 track‑by‑track analysis.
Core claim. The central claim is that a ²²⁸Th–D₂O photoneutron source, with the ²²⁸Th disk positioned 1 cm from the forward end face of a D₂O cylinder, produces a neutron field whose forward emission rate is 1734 ± 76 (stat.) ± 65 (sys.) h⁻¹ and whose energy spectrum is peaked at 0.17–0.22 MeV, with all neutrons below 1 MeV. The authors assert that this spectrum is consistent with Geant4 simulations and that the source is suitable for few‑keV nuclear‑recoil calibration of large liquid‑xenon detectors. They further claim that the source’s half‑life (1.91 y) is substantially longer than that of commonly used ⁸⁸Y‑Be sources (106.6 d), and that its spectrum lacks the ~0.95 MeV component present in ⁸⁸Y‑Be, making it cleaner for low‑energy calibrations.
The premise it rests on. The Geant4 simulation of the 3He counter’s absolute detection efficiency (7.63%) is taken as correct, but it is not validated by an independent absolute measurement; the simulation is normalized to data only for relative position dependence, not for absolute normalization.
If the paper is right (Pith's inferences, not the authors')
Large liquid‑xenon detectors (e.g., PandaX‑20T, XENONnT, LZ) can use this source for extended low‑energy nuclear‑recoil calibration campaigns without frequent source replacement due to its 1.91‑year half‑life.
The simultaneous 2.615 MeV gamma ray provides a calibration line near the 0νββ Q‑value of ¹³⁶Xe, enabling cross‑calibration of the high‑energy electron‑recoil response in a separate deployment mode.
The measured emission rate and spectrum serve as a benchmark for validating Geant4 simulations of neutron transport and detection in liquid‑xenon detectors.
The optimized shielding design (tungsten annulus + forward tungsten cylinder) demonstrates a path to achieve a usable NR/ER ratio while staying within data‑acquisition bandwidth.
An exposure of ~220 h in PandaX‑20T is predicted to yield ~1000 selected nuclear‑recoil events, sufficient for characterizing the detector’s few‑keV response.
The source could be adapted for other detector technologies (e.g., solid‑state, organic scintillators) that require a low‑energy neutron field for nuclear‑recoil calibration, possibly with different shielding designs.
The strong dependence of the escaping spectrum on D₂O geometry suggests that varying the source position or D₂O volume could tailor the neutron energy distribution for specific calibration needs, e.g., shifting the peak energy.
An independent absolute measurement of the neutron emission rate (e.g., by manganese‑bath activation) would strengthen confidence in the Geant4‑derived detection efficiency, which currently carries the dominant systematic uncertainty.
The paper’s note that the simulated cross sections were modified for Fig.2 (only the relative angular shape is meaningful) hints that the absolute angular distribution of emitted neutrons is not fully validated, which could affect the simulated event rate in detectors installed at non‑forward positions.
Check this review
Pith stored no record of this review's passes and took no fingerprint of it. Reviews made by code from 13 Nov 2026 on carry both.
How often reviews like this one are right
Reviews like this one: written by deepseek-v4-flash, where what was read was not recorded.
How often objections from reviews like this one are right: not yet measured
Reviews by people
This review was written by a machine. No person has reviewed this paper on Pith yet. A scientist listed on Pith can review it for a fee, and their review is published here whatever it concludes.
The paper
Abstract
Calibrating the few-keV nuclear-recoil response of liquid xenon detectors calls for a low-energy neutron source. We have developed and characterized a 228Th–D2O photoneutron source that converts the 2.615 MeV γ rays from 208Tl into sub-MeV neutrons through the D(γ,n)p reaction. After optimizing the D2O geometry to balance neutron production against escape efficiency, we validated the chosen configuration experimentally. The forward emission rate is 1734 ± 76 (stat.) ± 65 (sys.) h−1, and the spectrum unfolded from Bonner sphere data is concentrated at 0.17–0.22 MeV. In a 220 h exposure of PandaX-20T, the same configuration should provide about 1000 selected nuclear-recoil events. These results support the use of a long-lived 228Th–D2O source for low-energy neutron calibration of large liquid xenon detectors.
The paper's own sentences, arranged, kept beside the review and separate from it. Each label is drawn from a closed vocabulary; the number beside it is the probability the classifier gave that label. No sentence in this record was written by a model.
abstract compiled · 2026-11-13
The authors' main claim, as the classifier picked it out 1.00
“These results support the use of a long-lived $^{228}$Th--D$_2$O source for low-energy neutron calibration of large liquid xenon detectors.”
How the main claim is stated: hedged possibility 0.73
What the abstract itself states, as probabilities: claims first 0.04 · states limitation 0.16 · human participants 0.02 · controlled comparison 0.03 · randomized assignment 0.01 · states quantitative result 0.99 · compares with prior results 0.07 · mentions code or data availability 0.02
The abstract, sentence by sentence
This table labels the abstract only (arXiv listing abstract v1). The number is the probability the classifier gave the label.
| Sentence | Role | The authors' own work | States the main conclusion | Strength |
|---|---|---|---|---|
| Calibrating the few-keV nuclear-recoil response of liquid xenon detectors calls for a low-energy neutron source. | background 0.99 | 0.05 | 0.03 | asserted 0.99 |
| We have developed and characterized a $^{228}$Th--D$_2$O photoneutron source that converts the 2.615~MeV $\gamma$ rays from $^{208}$Tl into sub-MeV neutrons through the D($\gamma$,n)p reaction. | contribution 0.84 | 0.96 | 0.08 | asserted 1.00 |
| After optimizing the D$_2$O geometry to balance neutron production against escape efficiency, we validated the chosen configuration experimentally. | method 0.95 | 0.96 | 0.04 | asserted 1.00 |
| The forward emission rate is $1734 \pm 76~(\mathrm{stat.}) \pm 65~(\mathrm{sys.})~\mathrm{h}^{-1}$, and the spectrum unfolded from Bonner sphere data is concentrated at $0.17$--$0.22~\mathrm{MeV}$. | result 1.00 | 0.96 | 0.06 | asserted 1.00 |
| In a 220~h exposure of PandaX-20T, the same configuration should provide about 1000 selected nuclear-recoil events. | result 0.71 | 0.51 | 0.06 | hedged possibility 0.85 |
| These results support the use of a long-lived $^{228}$Th--D$_2$O source for low-energy neutron calibration of large liquid xenon detectors. | interpretation 0.99 | 0.66 | 0.94 | hedged possibility 0.73 |
What kind of paper the classifier reports this is: method or tool 0.89 · evidence from observation or experiment 1.00
What was examined, and what was not
Sources read
- arXiv listing abstract v1
Not examined
- full text
- figures and tables
- references and citing sentences
- relations to other papers
Levels run
- P abstract
Not run
- S sections
- T body sentences
- C citations
- R cross-paper relations
How the labels were made
Compiled by classifier-1.13.0 via TypeSafe API, question bank abstract-p1-20260923, 2026-11-13: 1 request, 5,192 input tokens.
Every quoted sentence on this record is the authors'. Every label is from Pith's closed vocabulary, with the probability shown beside it. This record as JSON
Figures (10)
Figures are the authors', reproduced from the paper.
References (47)
- [1] D. Z. Freedman, Coherent effects of a weak neutral current, Phys. Rev. D 9 (1974) 1389–1392. doi:10.1103/PhysRevD.9.1389 on Pithdoi:10.1103/physrevd.9.1389
- [2] J. Billard, E. Figueroa-Feliciano, L. Strigari, Implication of neutrino backgrounds on the reach of next generation dark matter direct detection experiments, Phys. Rev. D 89 (2014) 023524. doi:10.1103/PhysRevD.89.023524 on Pithdoi:10.1103/physrevd.89.023524
- [3] C. A. J. O’Hare, New definition of the neutrino floor for direct dark matter searches, Phys. Rev. Lett. 127 (2021) 251802. doi:10.1103/PhysRevLett.127.251802 on Pithdoi:10.1103/physrevlett.127.251802
- [4] E. Aprile, et al., The XENONnT dark matter experiment, Eur. Phys. J. C 84 (2024) 784. doi:10.1140/epjc/s10052-024-12982-5 on Pithdoi:10.1140/epjc/s10052-024-12982-5
- [5] XENON Collaboration, First indication of solar 8B neutrinos via coherent elastic neutrino-nucleus scattering with XENONnT, Phys. Rev. Lett. 133 (2024) 191002. doi:10.1103/PhysRevLett.133.191002 on Pithdoi:10.1103/physrevlett.133.191002
- [6] PandaX Collaboration, First indication of solar 8B neutrinos through coherent elastic neutrino-nucleus scattering in PandaX-4T, Phys. Rev. Lett. 133 (2024) 191001. doi:10.1103/PhysRevLett.133.191001 on Pithdoi:10.1103/physrevlett.133.191001
- [7] D. S. Akerib, et al., Searches for light dark matter and evidence of coherent elastic neutrino-nucleus scat- tering of solar neutrinos with the LUX-ZEPLIN (LZ) experiment, Phys. Rev. Lett. 137 (9) (2026) 091806. arXiv:2512.08065, doi:10.1103/jvqf-njpj. 21 on PithPith's reviewarXiv
- [8] G. Shen, et al., Search for solar boosted dark matter particles at the PandaX-4T experiment, Phys. Rev. Lett. 134 (2025) 161003. doi:10.1103/PhysRevLett.134.161003 on Pithdoi:10.1103/physrevlett.134.161003
- [9] W. Ma, et al., Internal calibration of the PandaX-II detector with radon gaseous sources, JINST 15 (2020) P12038. doi:10.1088/1748-0221/15/12/P12038 on Pithdoi:10.1088/1748-0221/15/12/p12038
- [10] E. Aprile, et al., Low-energy calibration of XENON1T with an internal 37Ar source, Eur. Phys. J. C 83 (2023) on Pith
- [11] X. Guo, et al., Preparation and measurement of an 37Ar source for liquid xenon detector calibration, Nucl. Sci. Tech. 37 (2026) 14. doi:10.1007/s41365-025-01818-w on Pithdoi:10.1007/s41365-025-01818-w
- [12] D. S. Akerib, et al., Low-energy (0.7–74 keV) nuclear recoil calibration of the LUX dark matter experiment using D-D neutron scattering kinematics, arXiv preprint (2016). arXiv:1608.05381, doi:10.48550/arXiv.1608.05381. URLhttps://arxiv.org/abs/1608.05381 on PithPith's reviewarXivdoi:10.48550/arxiv.1608.05381
- [13] E. Aprile, et al., Simultaneous measurement of the light and charge response of liquid xenon to low-energy nuclear recoils at multiple electric fields, Phys. Rev. D 98 (2018) 112003. doi:10.1103/PhysRevD.98.112003 on Pithdoi:10.1103/physrevd.98.112003
- [14] B. Yan, et al., Determination of responses of liquid xenon to low energy electron and nuclear recoils using a PandaX-II detector, Chin. Phys. C 45 (7) (2021) 075001. doi:10.1088/1674-1137/abf6c2 on Pithdoi:10.1088/1674-1137/abf6c2
- [15] D. S. Akerib, et al., Nuclear recoil calibration at sub-keV energies in LUX and its impact on dark matter search sensitivity, Phys. Rev. Lett. 134 (2025) 061002. doi:10.1103/PhysRevLett.134.061002 on Pithdoi:10.1103/physrevlett.134.061002
- [16] Z. Huang, G. Shen, Q. Wang, et al., Neutron-induced nuclear recoil background in the PandaX-4T experiment, Chin. Phys. C 46 (11) (2022) 115001. doi:10.1088/1674-1137/ac8539 on Pithdoi:10.1088/1674-1137/ac8539
- [17] J. Aalbers, et al., Background determination for the LUX-ZEPLIN dark matter experiment, Phys. Rev. D 108 (2023) 012010. doi:10.1103/PhysRevD.108.012010 on Pithdoi:10.1103/physrevd.108.012010
- [18] W. Mannhart, Evaluation of the Cf-252 fission neutron spectrum between 0 MeV and 20 MeV, in: Properties of Neutron Sources, no. 410 in IAEA-TECDOC, International Atomic Energy Agency, Vienna, 1987, pp. 158–170, proceedings of an Advisory Group Meeting held in Leningrad, USSR, ... on Pith
- [19] J. W. Marsh, D. J. Thomas, M. Burke, High resolution measurements of neutron energy spectra from AmBe and AmB neutron sources, Nucl. Instrum. Methods Phys. Res. A 366 (2-3) (1995) 340–348. doi:10.1016/0168- 9002(95)00613-3 on Pithdoi:10.1016/0168-
- [20] H. Liskien, A. Paulsen, Neutron production cross sections and energies for the reactions T(p,n) 3He, D(d,n)3He, and T(d,n)4He, At. Data Nucl. Data Tables 11 (7) (1973) 569–619. doi:10.1016/S0092-640X(73)80081-6. 22 on Pithdoi:10.1016/s0092-640x(73)80081-6
show 27 more
- [21] J. I. Collar, Applications of an 88Y/Be photoneutron calibration source to dark matter and neutrino experiments, Phys. Rev. Lett. 110 (2013) 211101. doi:10.1103/PhysRevLett.110.211101 on Pithdoi:10.1103/physrevlett.110.211101
- [22] XENON Collaboration, Low-energy nuclear recoil calibration of XENONnT with a 88YBe photoneutron source, Phys. Rev. D 113 (2026) 112017. doi:10.1103/rsl9-j1ky on Pithdoi:10.1103/rsl9-j1ky
- [23] LZ Collaboration, Low-energy nuclear recoil calibration of the LUX-ZEPLIN experiment with a photoneutron source, arXiv preprint (2025). arXiv:2509.16281, doi:10.48550/arXiv.2509.16281. URLhttps://arxiv.org/abs/2509.16281 on Pithdoi:10.48550/arxiv.2509.16281
- [24] M. J. Martin, Nuclear Data Sheets forA=208, Nucl. Data Sheets 108 (8) (2007) 1583–1806. doi:10.1016/j.nds.2007.07.001 on Pithdoi:10.1016/j.nds.2007.07.001
- [25] M.-M. Bé, V . Chisté, C. Dulieu, X. Mougeot, V . P. Chechev, F. G. Kondev, A. L. Nichols, X. Huang, B. Wang, Table of Radionuclides, V olume 7:A=14 to 245, Bureau International des Poids et Mesures, Sèvres, France, 2013 on Pith
- [26] P. Marin, G. R. Bishop, H. Halban, The absolute standardization of the 2.615 MeVγ-rays of ThC” and the cross section for the photodisintegration of the deuteron at this energy, Proc. Phys. Soc. A 66 (1953) 608–616. doi:10.1088/0370-1298/66/7/305 on Pithdoi:10.1088/0370-1298/66/7/305
- [27] V . P. Chechev, The evaluation of half-lives and other decay data used in nuclear astrophysics and cosmochronol- ogy, Phys. At. Nucl. 74 (12) (2011) 1713–1717. doi:10.1134/S106377881111007X on Pithdoi:10.1134/s106377881111007x
- [28] M. Redshaw, E. Wingfield, J. McDaniel, E. G. Myers, Mass and double-beta-decay Q value of136Xe, Phys. Rev. Lett. 98 (2007) 053003. doi:10.1103/PhysRevLett.98.053003 on Pithdoi:10.1103/physrevlett.98.053003
- [29] S. Agostinelli, et al., Geant4—a simulation toolkit, Nucl. Instrum. Methods Phys. Res. A 506 (2003) 250–303. doi:10.1016/S0168-9002(03)01368-8 on Pithdoi:10.1016/s0168-9002(03)01368-8
- [30] J. Allison, et al., Recent developments in Geant4, Nucl. Instrum. Methods Phys. Res. A 835 (2016) 186–225. doi:10.1016/j.nima.2016.06.125 on Pithdoi:10.1016/j.nima.2016.06.125
- [31] A. Abdukerim, et al., PandaX-xT—a deep underground multi-ten-tonne liquid xenon observatory, Sci. China Phys. Mech. Astron. 68 (2) (2025) 221011. arXiv:2402.03596, doi:10.1007/s11433-024-2539-y on PithPith's reviewarXiv
- [32] J. Aalbers, et al., The design, implementation, and performance of the LZ calibration systems, JINST 19 (2024) P08027. doi:10.1088/1748-0221/19/08/P08027 on Pithdoi:10.1088/1748-0221/19/08/p08027
- [33] C. Li, et al., Precise measurement of 216Po half-life with exact parent-daughter pairing in PandaX-4T, Phys. Rev. C 113 (1) (2026) 014610. doi:10.1103/6r8d-qxb8. 23 on Pithdoi:10.1103/6r8d-qxb8
- [34] D. J. Thomas, A. V . Alevra, Bonner sphere spectrometers—a critical review, Nucl. Instrum. Methods Phys. Res. A 476 (1-2) (2002) 12–20. doi:10.1016/S0168-9002(01)01379-1 on Pithdoi:10.1016/s0168-9002(01)01379-1
- [35] R. L. Bramblett, R. I. Ewing, T. W. Bonner, A new type of neutron spectrometer, Nucl. Instrum. Methods 9 (1960) 1–12. doi:10.1016/0029-554X(60)90043-4 on Pithdoi:10.1016/0029-554x(60)90043-4
- [36] M. Reginatto, P. Goldhagen, S. Neumann, Spectrum unfolding, sensitivity analysis and propagation of uncertain- ties with the maximum entropy deconvolution code MAXED, Nucl. Instrum. Methods Phys. Res. A 476 (1-2) (2002) 242–246. doi:10.1016/S0168-9002(01)01439-5 on Pithdoi:10.1016/s0168-9002(01)01439-5
- [37] M. Matzke, Propagation of uncertainties in unfolding procedures, Nucl. Instrum. Methods Phys. Res. A 476 (1-2) (2002) 230–241. doi:10.1016/S0168-9002(01)01438-3 on Pithdoi:10.1016/s0168-9002(01)01438-3
- [38] M. Luszik-Bhadra, M. Reginatto, H. Wershofen, B. Wiegel, A. Zimbal, New PTB thermal neutron calibration facility: first results, Radiat. Prot. Dosim. 161 (1-4) (2014) 352–356. doi:10.1093/rpd/nct354 on Pithdoi:10.1093/rpd/nct354
- [39] P. Wang, J. Yang, H. Zhang, Z. Ren, R. Wang, J. Zhang, Measurement and evaluation of neutron spectra in nuclear power plants, J. Harbin Eng. Univ. 43 (11) (2022) 1621–1629. doi:10.11990/jheu.202206016 on Pithdoi:10.11990/jheu.202206016
- [40] M. Reginatto, B. Wiegel, A. Zimbal, UMG 3.1: Unfolding with MAXED and GRA VEL, Physikalisch- Technische Bundesanstalt (2002) on Pith
- [41] D. J. Thomas, Use of a priori information in Bonner sphere unfolding, in: Modern Neutron Detection, IAEA- TECDOC-1935, International Atomic Energy Agency, Vienna, 2020, pp. 131–136 on Pith
- [42] X. Chen, et al., BambooMC—a Geant4-based simulation program for the PandaX experiments, JINST 16 (2021) T09004. arXiv:2107.05935, doi:10.1088/1748-0221/16/09/T09004 on PithPith's reviewarXiv
- [43] A. Manzur, et al., Scintillation efficiency and ionization yield of liquid xenon for monoenergetic nuclear recoils down to 4 keV, Phys. Rev. C 81 (2010) 025808. doi:10.1103/PhysRevC.81.025808 on Pithdoi:10.1103/physrevc.81.025808
- [44] G. Plante, et al., New measurement of the scintillation efficiency of low-energy nuclear recoils in liquid xenon, Phys. Rev. C 84 (2011) 045805. doi:10.1103/PhysRevC.84.045805 on Pithdoi:10.1103/physrevc.84.045805
- [45] B. Lenardo, et al., Low-energy physics reach of xenon detectors for nuclear-recoil-based dark matter and neutrino experiments, Phys. Rev. Lett. 123 (2019) 231106. doi:10.1103/PhysRevLett.123.231106 on Pithdoi:10.1103/physrevlett.123.231106
- [46] A. E. Robinson, Reanalysis of radioisotope measurements of the 9Be(γ,n)8Be cross section, Phys. Rev. C 94 (2016) 024613. doi:10.1103/PhysRevC.94.024613. 24 on Pithdoi:10.1103/physrevc.94.024613
- [542] doi:10.1140/epjc/s10052-023-11512-z on Pithdoi:10.1140/epjc/s10052-023-11512-z
Challenges and discussion
Disagree with the review, or know something it missed? Leave a comment, or challenge this review formally: a challenge becomes a permanent entry in this paper's record, and its decision is published here.
Come back by email
A short list of the week's newly reviewed papers. Nothing is sent until you confirm from your inbox, and every email has a link that stops it. No account.
No comments yet.
Sign in to comment