Laura Baudis, Sofia Calgaro, Jose Javier Cuenca-Garcia, Florian Joerg, Margherita Noia, Francesco Piastra, Aravind Remesan Sreekala

The paper says

“We report on a readout upgrade of Gator, a low-background high-purity germanium spectrometer operated at the Laboratori Nazionali del Gran Sasso for material radioassay.”

more of the abstract

The upgraded data acquisition records digitised waveforms, enabling noise rejection on an event-level basis, and offline energy reconstruction. Measurements before and after the upgrade show an improved energy resolution of 17% at the 2614.5 keV line of 208-Tl, with (2.14 +- 0.22) keV at FWHM, and suppression of electronic noise in the low-energy region. We assess the resulting sensitivity for a specified sample geometry, and present a new background measurement. The background rate in the (100-2700) keV range is (86.6 +- 0.5) counts/(kg day). We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment.

from the abstract

Pith says

Machine review · deepseek-v4-flash · 13 Nov 2026 · what was read was not recorded · not checked by a person

It can be wrong. Challenge this review or reply to it.

Holds up. Solid technical report on a readout upgrade and PMT radioassay; the results are useful for the XLZD collaboration and the low-background community, but the novelty is incremental and the secular equilibrium assumption is standard and acknowledged. This line is wrong

3 major objections · 6 minor

How often objections from reviews like this one are right: not yet measured

Are you an author? What you can do.

Not checked: figures, code, what the 25 references say.

The paper says, at §3.5, Eq. (3)

The review did not quote this passage.

Pith says: Objection 1 of 3 · major

The likelihood analysis assumes secular equilibrium within the 238U and 232Th decay-chain segments (sentence [b118]). While this is a common assumption in gamma-ray spectrometry, the paper does not provide evidence that secular equilibrium holds for the specific PMT materials. If chemical processing during stem fabrication has separated parent and daughter isotopes, the activity estimates for chain segments (e.g., 226Ra, 228Ra, 228Th) could be biased.

more of objection 1

The authors should discuss the validity of this assumption for the PMT stems, or at least acknowledge this limitation more explicitly. The current text only states the assumption without justification.

The paper says, at §4.2, Fig. 6 and §4.3

The review did not quote this passage.

Pith says: Objection 2 of 3 · major

The background comparison shows a 9% reduction in the integrated rate (100–2700 keV) relative to the pre-upgrade dataset, but the authors note that the present background level does not show a reduction relative to the earlier published value of (82.0 ± 0.7) events/(kg day) from Ref.

more of objection 2

[3]. This discrepancy is not fully explained. The paper should clarify whether the 9% reduction is relative to a different pre-upgrade dataset (the 42.4-day run shown in Fig. 6) and why the current background is higher than the 2011/2022 value. This is important for understanding the stability and performance of the facility.

The paper says, at §5.2, Table 2

The review did not quote this passage.

Pith says: Objection 3 of 3 · major

For 226Ra, 228Ra, and 228Th, the counting analysis yields upper limits while the likelihood analysis returns finite activities. The paper states that the central values are close and agree within uncertainties for nuclides with finite activities in both methods, but it does not discuss the systematic difference for these three nuclides.

more of objection 3

The authors should explain why the likelihood analysis is able to extract finite values where the counting analysis cannot, and whether this indicates a limitation of the counting method or a potential bias in the likelihood model (e.g., due to the secular equilibrium assumption).

6 minor objections

The paper says, at §1, Abstract

Pith says: minor objection 1 of 6

The abstract states 'improved energy resolution of 17% at the 2614.5 keV line of 208Tl, with (2.14 ± 0.22) keV at FWHM'. The 17% improvement should be explicitly referenced to the pre-upgrade value (2.59 ± 0.24) keV from Table 1. Consider adding a brief comparison for clarity.

The paper says, at §3.2, Fig. 3

Pith says: minor objection 2 of 6

The right panel of Fig. 3 shows the baseline mean versus reconstructed energy. The color scale is not clearly defined in the caption; it should specify that it represents the number of events. Also, the orange dashed line indicating the baseline cut is not described in the caption.

show 4 more

The paper says, at §4.1, Table 1

Pith says: minor objection 3 of 6

The table lists the 185.72 keV line as 235U with a note that it includes the 186.72 keV emission of 226Ra. However, the energy resolution at this line is reported as 1.00 ± 0.06 keV after upgrade. The overlap of the two lines may affect the resolution measurement; the authors should comment on how this was handled in the fit (e.g., using a doublet model).

The paper says, at §5.2, Table 3

Pith says: minor objection 4 of 6

The table compares activities with XENONnT, LZ, and PandaX-4T. The footnote indicates that upper limits for LZ and PandaX-4T are at 90% CL, while the R17317 values are at 95% CL (from Table 2). This difference in confidence level should be noted in the main text to avoid misleading comparisons.

The paper says, at §6, Conclusions

Pith says: minor objection 5 of 6

The sentence 'While the concrete implications for the XLZD background budget require dedicated Monte Carlo simulations' is important and should be included in the abstract or introduction to set expectations. Currently, the abstract states the PMTs are 'suitable candidate photosensors' without this caveat.

The paper says, at General

Pith says: minor objection 6 of 6

The paper uses the notation 'counts/(kg day)' for background rate. It should be clarified whether the mass refers to the detector mass (2.2 kg) or the sample mass. From context, it appears to be the detector mass, but this should be stated explicitly.

What would settle it

Verify that the 17% resolution improvement at 2614.5 keV is robust by re-fitting the pre- and post-upgrade peaks with an alternative peak model (e.g., a simple Gaussian plus linear background instead of the Crystal Ball function) and checking that the FWHM difference remains consistent within uncertainties.

from the skeptic's reading

How the review went

4 readings by deepseek-v4-flash · 13 Nov 2026

  1. Reader holds up confidence high
  2. Skeptic agrees 9 s · 14,939 tokens read, 482 written
  3. Editor wrote the verdict above 6 s · 15,737 tokens read, 507 written
  4. Referee minor revision: 3 major objections, 6 minor 12 s · 15,209 tokens read, 1,355 written

Not run on this paper: the simulated authors' defence.

Pith did not check this review's objections against the paper.

The review in full

The review in brief, machine-written

This is a straightforward, well-executed technical report on two things: (1) a readout upgrade to the Gator HPGe spectrometer that adds waveform digitisation, and (2) a radioassay of sixteen Hamamatsu R17317 PMTs with an improved stem design. The paper does what it says on the tin. The upgrade is incremental—replacing an MCA with a digitiser is a standard move—but the characterisation is thorough: they show a 17% improvement in energy resolution at 2614.5 keV, suppression of a noise feature near 100 keV, and a background rate of (86.6 ± 0.5) counts/(kg day). The PMT assay is the more interesting part for the XLZD collaboration: the new stem design reduces 238U and 232Th chain activities by about a factor of three compared to the PMTs used in XENONnT and LZ. The analysis uses two methods (counting and likelihood) that agree, which is reassuring. The paper is honest about its limitations: the noise below 20 keV is not yet understood, and the concrete implications for XLZD require dedicated simulations. The weakest point flagged by the reader—the assumption of secular equilibrium in the likelihood fit—is a standard assumption in gamma-ray spectrometry, and the authors state it explicitly. It is not a flaw unless there is evidence of disequilibrium in these specific PMT stems, and the paper does not claim otherwise. The free parameters (shaping time, filter sigma, baseline cut threshold) are all described; none are hidden. The paper is not a breakthrough, but it is a solid, reproducible piece of work. The code and data are not shipped, but the analysis methods are standard and the results are clearly presented. Who is this for? Members of the XLZD collaboration and anyone doing material screening for rare-event searches. It deserves peer review because the measurements are useful and the analysis is sound. I would accept a referee assignment on this.

The strongest objection

The paper is a well-executed technical report on a readout upgrade and a PMT radioassay. The central claims are: (1) the readout upgrade improves energy resolution by 17% at 2614.5 keV and suppresses electronic noise; (2) the background rate is (86.6 ± 0.5) counts/(kg day); (3) the new PMT stem design reduces activities from 238U and 232Th chains by a factor of three compared to previous PMTs. All three claims are supported by direct measurements, clear analysis methods, and appropriate uncertainty propagation. The weakest assumption identified by the reader—secular equilibrium within the 238U and 232Th chains for the likelihood analysis—is a standard assumption in gamma-ray spectrometry and is explicitly stated by the authors (sentence [b118]). The paper does not overclaim; it acknowledges that concrete implications for XLZD require dedicated simulations. No internal inconsistency, missing support, or circular reasoning is present. The results are useful for the XLZD collaboration and the broader low-background community.

The reader's verdict is ACCEPT with HIGH confidence, and the weakest assumption identified (secular equilibrium) is a standard assumption that the authors explicitly state. My stress-test pass finds no load-bearing concern that would change the verdict. The paper is technically sound, the claims are supported by data, and the limitations are acknowledged. The concrete test I propose is a simple robustness check that would further validate the headline resolution improvement, but it is not required to address a flaw. The agreement with the reader is 'agree' because the reader's weakest assumption is indeed the most notable assumption, but it is not a flaw—it is a standard practice. The verdict should remain UNCHANGED.

The referee's report

The results are significant for the low-background community, particularly for the XLZD experiment. The readout upgrade demonstrates a clear improvement in energy resolution and noise suppression, which enhances the sensitivity of the Gator facility for material screening. The PMT radioassay provides valuable input for the selection of photosensors for XLZD, showing lower activities than those reported for XENONnT and LZ. The paper is well-structured and the methods are standard and appropriate.

Pith's reading of the paper (a machine's summary, not the authors')

The Gator facility is a highly sensitive germanium detector used to measure tiny amounts of radioactivity in materials for rare-event searches like dark matter and neutrinoless double beta decay. This paper describes two main achievements. First, the readout electronics were upgraded to record the full waveform of each detector pulse, rather than just a histogram of pulse heights. This allows better rejection of electronic noise and more accurate energy reconstruction. After the upgrade, the energy resolution improved by 17% at the 2614.5 keV gamma line from 208Tl, and a noise feature near 100 keV was suppressed. The background rate in the 100-2700 keV range is 86.6 counts per kg per day. Second, the paper reports measurements of sixteen 3-inch photomultiplier tubes (PMTs) from Hamamatsu that have a new stem design using cobalt-free metal and small glass beads instead of a ceramic plate. This design reduces the radioactive contamination from uranium and thorium decay chains by about a factor of three compared to previous PMTs. The measured activities for key isotopes like 226Ra, 228Ra, 228Th, and 60Co are lower than those reported for PMTs used in current experiments XENONnT and LZ, making these tubes promising for the next-generation XLZD observatory.

Core claim. The readout upgrade of the Gator spectrometer improves energy resolution by 17% at the 2614.5 keV line of 208Tl (FWHM 2.14 ± 0.22 keV) and suppresses electronic noise in the low-energy region. The background rate in the (100-2700) keV range is (86.6 ± 0.5) counts/(kg day). The radioassay of sixteen Hamamatsu R17317 PMTs with improved stem design shows reduced activities from 238U and 232Th chains by a factor of three compared to previous PMTs, with lower central activities for 226Ra, 228Ra, 228Th, and 60Co than XENONnT and LZ values.

The premise it rests on. The assumption of secular equilibrium within the 238U and 232Th decay chains for the likelihood analysis (Section 3.5, Eq. 3: 's is shared by all lines of the same radionuclide or decay-chain segment, assuming secular equilibrium within the segment'). If secular equilibrium is not valid for the PMT materials (e.g., due to chemical processing that separates parent and daughter isotopes), the activity estimates for chain segments could be biased. The paper does not provide evidence that secular equilibrium holds for these specific PMT stems.

Check this review

Pith stored no record of this review's passes and took no fingerprint of it. Reviews made by code from 13 Nov 2026 on carry both.

How often reviews like this one are right

Reviews like this one: written by deepseek-v4-flash, where what was read was not recorded.

How often objections from reviews like this one are right: not yet measured

Every value above, in the review's public record (JSON)

Reviews by people

This review was written by a machine. No person has reviewed this paper on Pith yet. A scientist listed on Pith can review it for a fee, and their review is published here whatever it concludes.

Request a human review

How Pith checks papers · The full audit of this paper

The paper

Abstract

We report on a readout upgrade of Gator, a low-background high-purity germanium spectrometer operated at the Laboratori Nazionali del Gran Sasso for material radioassay. The upgraded data acquisition records digitised waveforms, enabling noise rejection on an event-level basis, and offline energy reconstruction. Measurements before and after the upgrade show an improved energy resolution of 17% at the 2614.5 keV line of 208-Tl, with (2.14 +- 0.22) keV at FWHM, and suppression of electronic noise in the low-energy region. We assess the resulting sensitivity for a specified sample geometry, and present a new background measurement. The background rate in the (100-2700) keV range is (86.6 +- 0.5) counts/(kg day). We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment.

The paper's own sentences, arranged, kept beside the review and separate from it. Each label is drawn from a closed vocabulary; the number beside it is the probability the classifier gave that label. No sentence in this record was written by a model.

abstract compiled · 2026-11-14

The authors' main claim, as the classifier picked it out 0.92

“We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment.”

How the main claim is stated: asserted 0.99

What the abstract itself states, as probabilities: claims first 0.04 · states limitation 0.16 · human participants 0.02 · controlled comparison 0.63 · randomized assignment 0.01 · states quantitative result 0.99 · compares with prior results 0.87 · mentions code or data availability 0.03

The abstract, sentence by sentence

This table labels the abstract only (arXiv listing abstract v1). The number is the probability the classifier gave the label.

SentenceRoleThe authors' own workStates the main conclusionStrength
We report on a readout upgrade of Gator, a low-background high-purity germanium spectrometer operated at the Laboratori Nazionali del Gran Sasso for material radioassay. contribution 0.99 0.90 0.10 asserted 0.99
The upgraded data acquisition records digitised waveforms, enabling noise rejection on an event-level basis, and offline energy reconstruction. method 0.94 0.84 0.06 asserted 1.00
Measurements before and after the upgrade show an improved energy resolution of 17% at the 2614.5 keV line of 208-Tl, with (2.14 +- 0.22) keV at FWHM, and suppression of electronic noise in the low-energy region. result 1.00 0.96 0.18 asserted 1.00
We assess the resulting sensitivity for a specified sample geometry, and present a new background measurement. contribution 0.98 0.95 0.07 asserted 0.99
The background rate in the (100-2700) keV range is (86.6 +- 0.5) counts/(kg day). result 1.00 0.95 0.09 asserted 1.00
We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment. result 0.73 0.95 0.37 asserted 0.99

What kind of paper the classifier reports this is: method or tool 0.76 · evidence from observation or experiment 1.00

What was examined, and what was not

Sources read

  • arXiv listing abstract v1

Not examined

  • full text
  • figures and tables
  • references and citing sentences
  • relations to other papers

Levels run

  • P abstract

Not run

  • S sections
  • T body sentences
  • C citations
  • R cross-paper relations

How the labels were made

Compiled by classifier-1.13.0 via TypeSafe API, question bank abstract-p1-20260923, 2026-11-14: 1 request, 5,170 input tokens.

Every quoted sentence on this record is the authors'. Every label is from Pith's closed vocabulary, with the probability shown beside it. This record as JSON

Figures (11)
Figure 1 of the paper
Figure 1. Section view of the Gator HPGe facility. The copper cryostat is located inside a copper-lined sample cavity, surrounded by lead and polyethylene shielding within a stainless-steel enclosure. A U-shaped copper cold finger immersed in liquid nitrogen provides the cooling of the crystal. channel analyser (MCA), which stored pulse-height spec￾tra [2,11]. Individual waveforms were thus not retained, limiting the information available for analysis, and for investigating electronic artefacts. Broad noise-related features below approximately 100 keV, as well as in￾creased analysis…
Figure 2 of the paper
Figure 2. Signal routing in the upgraded readout system, showing the devices relevant to signal acquisition and distribution. Only the inputs and outputs involved in the signal routing are represented. The HPGe detector is schematically represented as a reverse-biased diode inside the cryostat (orange box), with the preamplifier input JFET and feedback RC network located close to the crystal. The HV input of the Canberra 2002C is omitted for clarity. Data readout is performed via an A4818 CONET-USB adapter connected to the digitiser through optical fibres. tion for the identification of pile
Figure 3 of the paper
Figure 3. (Left): Energy spectrum in the (0-200) keV energy range before (blue) and after (orange) the baseline cut. The artefact peak is visible just below 100 keV. (Right): Baseline of the spectroscopy amplifier waveform versus reconstructed energy. A population of events (marked in red) corresponding to the peak artefact can be seen, along with a distribution of events at zero baseline. These two populations are removed by the baseline cut. The baseline mean selection is shown by the orange dashed line. The colour scale gives the number of events. 2500 5000 7500 10000 12500 15000 ADC…
Figure 4 of the paper
Figure 4. Energy calibration performed after the upgrade, using the Bedretto granite sample. error-weighted average. If none is detected, the lowest of the individual upper limits is adopted. The likelihood approach uses the observed counts directly, retaining separate Poisson terms for the peak and sideband regions of the sample and background spectra. For each line k, the peak region is defined as ±3σk around the nominal line energy, where σk denotes the corresponding energy resolution. The sideband re￾gion consists of two adjacent windows of width 3σk, one on each side of the peak, summed
Figure 5 of the paper
Figure 5. compares the peak widths obtained from the granite measurements before and after the upgrade. The γ-peaks in the post-upgrade spectrum have smaller resolution values over the measured energy range, in￾cluding the 208Tl line at 2614.5 keV. A higher than ex￾pected resolution is observed at an energy of 186 keV. This is caused by the overlap of two γ-lines from 226Ra (186.21 keV) and 235U (185.72 keV) which cannot be re￾solved individually. The two resolution curves are used in the sensitivity comparison presented in Sect. 4.3. In
Figure 6 of the paper
Figure 6. Background spectra before (orange) and after (blue) the readout upgrade, normalised per unit energy and measurement time. The post-upgrade spectrum includes the baseline selection. The measurement durations were 42.4 and 159.4 days, respectively. detection efficiencies and the measured background in windows around selected γ-ray lines. The study con￾siders the 214Bi line at 609.3 keV, the 228Ac line at 911.2 keV, the 40K line at 1460.8 keV and the 208Tl line at 2614.5 keV. The counting window extends to ±1.5σ around each γ-line, corresponding to a contain￾ment fraction of ∼0.866 for…
Figure 7 of the paper
Figure 7. Projected Minimum Detectable Activity (MDA), in mBq/PMT at 95% CL, for prominent radionuclides, before and after the DAQ upgrade. A measurement time of 42 days is assumed. 5 Radioassay of low-activity-stem PMTs To provide the internal components of the PMT such as the photocathode, the dynode chain and the screening electrode need to be delivered with high voltage. This is applied using pins protruding from the back side of the vacuum tube and being sealed with a stem. In the previous model of the R17317 PMTs this seal is es-
Figure 8 of the paper
Figure 8. Improvements in the stem design at the back of the R17317 PMTs. The original design (left) using a large ceramic plate for the stem is improved using low activity cobalt-free metal and smaller borosilicate glass beads (right). tablished via a ceramic plate (
Figure 9 of the paper
Figure 9. Arrangement of the sixteen R17317 PMTs in the Gator cavity as implemented in the geometry used for the detection efficiency MC simulations. Fifteen tubes are supported in PTFE holders (white) around the cop￾per cryostat (red), while one is placed on the cryostat endcap. background. As examples, the 40K peak and the two 60Co peaks are prominent, as well as lines from the ura￾nium and thorium decay chains. The likelihood analysis uses the sample and background counts separately, as we described in Sect. 3.5. The obtained activities or upper limits are shown in
Figure 10 of the paper
Figure 10. Measured energy spectra of the sixteen PMTs with holders (blue) and the holder background (orange). The durations of the two runs was 102.17 and 57.46 days, respectively. Statistical uncertainties are indicated as transparent bands
Figure 11 of the paper
Figure 11. Simulated detection efficiencies of Gator for the 16 R17317 PMTs as a function of the selected γ-ray en￾ergies. Error bars show the 10% systematic uncertainty

Figures are the authors', reproduced from the paper.

References (25)
  1. [3] G. R. Araujo et al., The upgraded low-background germanium counting facility Gator for high- sensitivityγ-ray spectrometry, JINST17, P08010 (2022). doi:10.1088/1748-0221/17/08/P08010 on Pithdoi:10.1088/1748-0221/17/08/p08010
  2. [1] G. Heusser, Low-radioactivity background tech- niques, Ann. Rev. Nucl. Part. Sci.45(1995), 543-590 doi:10.1146/annurev.ns.45.120195.002551 on Pithdoi:10.1146/annurev.ns.45.120195.002551
  3. [2] L. Baudis et al., Gator: a low-background count- ing facility at the Gran Sasso underground labo- ratory, JINST6, P08010 (2011). doi:10.1088/1748- 0221/6/08/P08010 on Pithdoi:10.1088/1748-
  4. [4] A. Bismark, Tests of the Fundamentals of Quantum Mechanics with Low-Background Experiments and Detector Technologies for Liquid Xenon Time Projec- tion Chambers, PhD Thesis University Zürich (2024). doi:10.5167/uzh-264330 on Pithdoi:10.5167/uzh-264330
  5. [5] G. A. P. Cirrone, G. Cuttone, F. Di Rosa, L. Pan- dola, F. Romano and Q. Zhang, Validation of the Geant4 electromagnetic photon cross-sections for ele- ments and compounds, Nucl. Instrum. Meth. A618, 315-322 (2010). doi:10.1016/j.nima.2010.02.112 on Pithdoi:10.1016/j.nima.2010.02.112
  6. [6] E. Aprile, K. Arisaka, F. Arneodo, A. Askin, L. Baudis, A. Behrens, K. Bokeloh, E. Brown, J. M. R. Cardoso and B. Choi, et al., Material screening and selection for XENON100, Astropart. Phys.35(2011), 43-49 doi:10.1016/j.astropartphys.2011.06.001 on Pithdoi:10.1016/j.astropartphys.2011.06.001
  7. [7] E. Aprileet al.[XENON], Material radioassay and selection for the XENON1T dark matter ex- periment, Eur. Phys. J. C77(2017) no.12, 890 doi:10.1140/epjc/s10052-017-5329-0 on Pithdoi:10.1140/epjc/s10052-017-5329-0
  8. [8] E. Aprileet al.[XENON], Material radiopurity con- trol in the XENONnT experiment, Eur. Phys. J. C82 (2022) no.7, 599 doi:10.1140/epjc/s10052-022-10345- 6 on Pithdoi:10.1140/epjc/s10052-022-10345-
  9. [9] K. H. Ackermannet al.[GERDA], The GERDA experiment for the search of0νββdecay in 76Ge, Eur. Phys. J. C73(2013) no.3, 2330 doi:10.1140/epjc/s10052-013-2330-0 on Pithdoi:10.1140/epjc/s10052-013-2330-0
  10. [10] H. Acharyaet al.[LEGEND], First Results on the Search for Lepton Number Violating Neutrino- less Double-βDecay with the LEGEND-200 Exper- iment, Phys. Rev. Lett.136(2026) no.2, 022701 doi:10.1103/25tk-nctn on Pithdoi:10.1103/25tk-nctn
  11. [11] L. Baudis et al., Search for Pauli Exclusion Princi- ple violations with Gator at LNGS, Eur. Phys. J. C 84, 1137 (2024). doi:10.1140/epjc/s10052-024-13510- 1 on Pithdoi:10.1140/epjc/s10052-024-13510-
  12. [12] L. Baudis, DARWIN/XLZD: A future xenon observatory for dark matter and other rare in- teractions, Nucl. Phys. B1003, 116473 (2024). doi:10.1016/j.nuclphysb.2024.116473 on Pithdoi:10.1016/j.nuclphysb.2024.116473
  13. [13] J. Aalbers et al. (XLZD Collaboration), The XLZD Design Book: towards the next-generation liq- uid xenon observatory for dark matter and neu- trino physics, Eur. Phys. J. C85, 1192 (2025). doi:10.1140/epjc/s10052-025-14810-w on Pithdoi:10.1140/epjc/s10052-025-14810-w
  14. [14] B. Penning et al., Characterisation of the Bedretto underground site for fundamental physics experiments, Eur. Phys. J. C86, 866 (2026). doi:10.1140/epjc/s10052-026-15916-5 on Pithdoi:10.1140/epjc/s10052-026-15916-5
  15. [15] Ma, X. et al., Multi-disciplinary characterizations of the BedrettoLab – a new underground geoscience research facility, Solid Earth,13, 301–322 (2026). doi:10.5194/se-13-301-2022, 2022 on Pithdoi:10.5194/se-13-301-2022
  16. [16] Agostinelli, S. et al. (GEANT4 Collaboration), GEANT4: A Simulation toolkit, Nucl. Instrum. Meth. A502, 250–303 (2003). doi:10.1016/S0168- 9002(03)01368-8 on Pithdoi:10.1016/s0168-
  17. [17] Hans Dembinski and Piti Ongmongkolkul et al., scikit-hep/iminuit, Zenodo, (2020). doi: 10.5281/zen- odo.3949207 on Pithdoi:10.5281/zen-
  18. [18] C. Hurtgen, S. Jerome, and M. Woods, Revisiting Currie – how low can you go?, Appl. Radiat. Isot.53, 11 45 (2000). doi:10.1016/S0969-8043(00)00171-8 on Pithdoi:10.1016/s0969-8043(00)00171-8
  19. [19] D. S. Akerib et al. (LZ Collaboration), The LUX- ZEPLIN (LZ) radioactivity and cleanliness con- trol programs, Eur. Phys. J. C80, 1044 (2020). doi:10.1140/epjc/s10052-020-8420-x on Pithdoi:10.1140/epjc/s10052-020-8420-x
  20. [20] Qian, Z., Si, L. et al. (PandaX-4T Collabo- ration), Low radioactive material screening and background control for the PandaX-4T experi- ment, J. High Energy Phys.2022, 147 (2022). doi:10.1007/JHEP06(2022)147 on Pithdoi:10.1007/jhep06(2022)147
show 5 more
  1. [21] J. Aalberset al.[XLZD], Neutrinoless double beta decay sensitivity of the XLZD rare event observatory,’ J. Phys. G52(2025) no.4, 045102 doi:10.1088/1361- 6471/adb900 on Pithdoi:10.1088/1361-
  2. [22] FMCDAQ software repository.https://github. com/Physik-Institut-UZH/FMCDAQ on Pith
  3. [23] GatorTools software repository.https://github. com/Physik-Institut-UZH/GatorTools on Pith
  4. [24] M. Béet al.Table of Radionuclides (vol. 4), Mono- graphie BIPM-5 (2008), Bureau International des Poids et Mesures on Pith
  5. [25] L. Baudiset al.[QROCODILE], First Sub-MeV Dark Matter Search with the QROCODILE Experi- mentUsingSuperconductingNanowireSingle-Photon Detectors, Phys. Rev. Lett.135(2025) no.8, 081002 doi:10.1103/4hb6-f6jl Appendix A Detection efficiencies Table 4: Detection efficiencies and b... on Pithdoi:10.1103/4hb6-f6jl

Challenges and discussion

Disagree with the review, or know something it missed? Leave a comment, or challenge this review formally: a challenge becomes a permanent entry in this paper's record, and its decision is published here.

No comments yet.

Sign in to comment